Captiva EMR–GPF Pass-Through Cleanup for the Analysis of Pesticides in Celery
Technical Reference Summary
Covered Models & Document Scope
Nota de aplicación sobre el análisis de pesticidas en apio mediante limpieza de paso directo con Agilent Captiva EMR–GPF (cartucho de 3 mL, número de referencia 5610-2090) en comparación con Bond Elut QuEChERS Universal dSPE con GCB; abarca las condiciones del método de LC/MS/MS y los resultados de recuperación/RSD en niveles de QC de 4 ng/g y 50 ng/g.
Technical Data
| Parameter | Value | Condición |
|---|---|---|
| SPE cartridge volume | 3 mL p.2 ↗ | Captiva EMR–GPF, part number 5610-2090 |
| SPE cartridge sorbent | Carbon S p.2 ↗ | Captiva EMR–GPF |
| Sample loading volume | 800 µL p.3 ↗ | Captiva EMR–GPF, mixed with water |
| Average recovery, all compounds | >90 % p.3 ↗ | Captiva EMR–GPF, 4 ng/g QC level |
| Average recovery, all compounds | >90 % p.3 ↗ | Captiva EMR–GPF, 50 ng/g QC level |
| Average RSD, all compounds | ≤20 % p.3 ↗ | Captiva EMR–GPF, both QC levels |
| Extraction kit | Bond Elut QuEChERS AOAC, part number 5982-5755CH p.2 ↗ | with ceramic homogenizers |
| dSPE kit (comparison) | Bond Elut QuEChERS Universal dSPE with GCB, 15 mL, part number 5982-0029 p.2 ↗ | traditional cleanup |
| Positive pressure manifold | PPM-48, part number 5191-4101 p.2 ↗ | Captiva EMR–GPF passthrough |
| LC system | 1290 Infinity II with high-speed pump G7120A, multisampler G7167B, multicolumn thermostat G7116B p.2 ↗ | |
| MS system | triple quadrupole LC/MS G6470A with Jet Stream Electrospray ion source p.2 ↗ | |
| Software | MassHunter Workstation p.2 ↗ | data acquisition and analysis |
Operating Limits
- Sample loading volume on Captiva EMR–GPF cartridge: 800 µL p.3 ↗
- Target recovery pass rate (average recovery, all compounds): >90% p.3 ↗
- Average RSD, all compounds: ≤20% p.3 ↗
Ordering Data
| Ordering Code | Part No. | Variant / Code | Aplica a | Note |
|---|---|---|---|---|
| Agilent Bond Elut QuEChERS AOAC extraction kit with ceramic homogenizers p.2 ↗ | 5982-5755CH | Extraction kit used in method | ||
| Agilent Bond Elut QuEChERS Universal dispersive SPE kit with GCB, 15 mL p.2 ↗ | 5982-0029 | Traditional dSPE cleanup comparison | ||
| Agilent Captiva EMR–GPF cartridge, 3 mL p.2 ↗ | 5610-2090 | Passthrough cleanup cartridge | ||
| Agilent positive pressure manifold 48 processor (PPM-48) p.2 ↗ | 5191-4101 | Manifold for passthrough cleanup |