Induztr
Resumen técnico Visor de documentos Modelos relacionados

Pesticide Analysis of Peppers Using Captiva EMR–GPF Cleanup and LC/MS/MS or GC/MS/MS

Technical Reference Summary

Covered Models & Document Scope

Nota de aplicación que describe un método de determinación de residuos múltiples de pesticidas (240 pesticidas) en pimientos mixtos utilizando limpieza de paso directo con Agilent Captiva EMR–GPF seguida de LC/MS/MS y GC/MS/MS; abarca el cartucho Captiva EMR–GPF de 3 mL (número de pieza 5610-2090) y lo compara con Captiva EMR–HCF1, EMR–HCF2, EMR–GPD, EMR–LPD, U-dSPE con GCB y limpieza dSPE de proveedor 1/proveedor 2.

Technical Data

Parameter Value Condición
LC/MS/MS injection volume 2 μL p.3 ↗ LC/MS/MS conditions
LC/MS/MS stop time 10 min p.3 ↗ LC/MS/MS conditions
LC/MS/MS gas temperature 120 °C p.3 ↗ LC/MS/MS conditions
LC/MS/MS gas flow 20 L/min p.3 ↗ LC/MS/MS conditions
LC/MS/MS nebulizer pressure 40 psi p.3 ↗ LC/MS/MS conditions
LC/MS/MS sheath gas heater temperature 225 °C p.3 ↗ LC/MS/MS conditions
LC/MS/MS low-pressure RF 60 V p.3 ↗ positive polarity
LC/MS/MS low-pressure RF 60 V p.3 ↗ negative polarity
LC/MS/MS polarity Positive and negative p.3 ↗ LC/MS/MS conditions
GC/MS/MS carrier gas Helium p.3 ↗ GC/MS/MS conditions
GC/MS/MS column 1 flow 1.0 mL/min p.3 ↗ GC/MS/MS conditions
GC/MS/MS column 2 flow 1.4 mL/min p.3 ↗ GC/MS/MS conditions
GC/MS/MS injection volume 1 µL p.3 ↗ cold splitless
GC/MS/MS inlet liner Agilent inlet liner, Ultra Inert, splitless p.3 ↗ GC/MS/MS conditions
Captiva EMR–GPF cartridge volume 3 mL p.3 ↗ Captiva EMR–GPF
Captiva EMR–GPF part number 5610-2090 p.3 ↗ Captiva EMR–GPF cartridge, 3 mL
Ceramic homogenizer pack size 100 pk p.3 ↗ 50 mL tubes, part number 5982-9313
Spiking level for recovery study 10 ng/g p.13 ↗ 230 targets
Recovery acceptance criterion 60 to 120 % p.16 ↗ accuracy
RSD acceptance criterion <20 % p.16 ↗ reproducibility
Matrix effect acceptance criterion (LC/MS/MS) 70 to 100 % p.16 ↗ LC/TQ only
Calibration curve acceptance criterion R2 >0.99 p.16 ↗ calibration curve
Number of pesticides targeted 240 p.1 ↗ multiresidue method
Number of LC-amenable pesticides 129 p.2 ↗ LC/MS/MS
Number of pesticides in quantitative analysis 230 p.10 ↗ bell pepper, LC/MS/MS and GC/MS/MS

Operating Limits

  • Recovery acceptance range: 60 to 120% p.16 ↗
  • RSD acceptance criterion: <20% p.16 ↗
  • Matrix effect acceptance range (LC/MS/MS): 70 to 100% p.16 ↗
  • Calibration curve acceptance criterion: R2 >0.99 p.16 ↗
  • Lowest spiking level used in recovery study: 10 ng/g p.13 ↗

Ordering Data

Ordering Code Part No. Variant / Code Aplica a Note
Agilent Captiva EMR–GPF cartridge, 3 mL p.3 ↗ 5610-2090 Passthrough cleanup cartridge used in the method
Agilent ceramic homogenizers, 50 mL tubes, 100/pk p.3 ↗ 5982-9313 Sample preparation consumable