Per- and Polyfluoroalkyl Substances Analysis in Soil
Contenido
Overviewp. 2
Background on PFAS compounds and analytical approach
Specp. 3
LC-MS/MS instrument parameters and settings
Procedurep. 4
Calibration standards and quantitation methodology
p. 5
Analysis results and interpretation
Qualityp. 6
Blank sample quality control results
Procedurep. 7
Spike recovery results for sandy loam matrix
Procedurep. 8
PFAS analysis in reed sedge peat and topsoil
p. 9
Summary of findings and method performance
p. 10
Supplementary data tables
p. 11
Cited literature and sources
p. 3table, 7 rows
p. 8table, 40 rows
p. 10table, 40 rows
p. 11table, 16 rows
Table 1. PFAS-suitable consumables and supplies.
Agilent Consumables and Supplies
Part Number
Carbon S for PFAS cartridge, 250 mg, 6 mL
5610-2247
Polypropylene autosampler screw top vials, 2 mL, and caps
5191-8151 and 5191-8121
Centrifuge tubes and caps, 15 mL
5610-2039
InfinityLab PFC delay column, 4.6 × 30 mm
5062-8100
ZORBAX RRHD Eclipse Plus C18 column, 2.1 × 100 mm, 1.8 µm
959758-902
Vac Elut SPS 24 manifold with collection rack for 10 × 75 mm test tubes
12234003
Collection rack and funnel set for 12 or 15 mL conical tubes, for Vac Elut SPS 24 manifold
12234027
Table 4. Average concentrations of PFAS measured in soil samples.
Acronym
Reed Sedge Peat
Topsoil
10:2 FTCA
<MRL
<MRL
10:2 FTSA
<MRL
<MRL
10:2 FTUCA
<MRL
<MRL
11Cl-PF3OUdS
<MRL
<MRL
3:3 FTCA
<MRL
<MRL
4:2 FTSA
<MRL
<MRL
4-PFecHS
<MRL
<MRL
5:3 FTCA
<MRL
<MRL
6:2 FTCA
<MRL
<MRL
6:2 FTSA
<MRL
<MRL
6:2 FTUCA
<MRL
<MRL
6:2/8:2 diPAP
<MRL
<MRL
6:6 PFPi
<MRL
<MRL
6:8 PFPi
<MRL
<MRL
7:3 FTCA
<MRL
<MRL
8:2 diPAP
<MRL
<MRL
8:2 FTCA
<MRL
<MRL
8:2 FTSA
<MRL
<MRL
8:2 FTUCA
<MRL
<MRL
8:8 PFPi
<MRL
<MRL
9Cl-PF3ONS
<MRL
<MRL
diSAmPAP
<MRL
<MRL
DONA
<MRL
<MRL
EtFOSE
<MRL
<MRL
FBSA
<MRL
<MRL
FDSA
<MRL
<MRL
FHxSA
<MRL
<MRL
HFPO-DA
<MRL
<MRL
MeFOSE
<MRL
<MRL
N-EtFOSA
<MRL
<MRL
N-EtFOSAA
<MRL
<MRL
NFDHA
<MRL
<MRL
N-MeFOSA
<MRL
<MRL
N-MeFOSAA
<MRL
<MRL
PFBA
4.5 ng/g
<MRL
PFBS
<MRL
<MRL
PFDA
<MRL
<MRL
PFDoDA
<MRL
<MRL
PFDoS
<MRL
<MRL
PFDS
<MRL
<MRL
Таблица напечатана в PDF разбитой на три последовательные части с повторяющимся заголовком Acronym / Reed Sedge Peat / Topsoil; здесь объединена в одну.
Table A1. Target spiking solution and target concentration in matrix.
Target Compound
CAS
Ret Time (min)
Target Quant Ion MRM Transition
Isotope Analog
Isotope MRM Transition
PFBA
375-22-4
2.43
213 & 169
13C4-PFBA
217 & 172
PFMPA
377-73-1
3.57
229 & 85
13C5-PFPeA
268 & 223
3:3 FTCA
356-02-5
3.88
241 & 177
13C5-PFPeA
268 & 223
PFPeA
2706-90-3
3.89
263 & 219
13C5-PFPeA
268 & 223
PFBS
375-73-5
3.97
299 & 80
13C3-PFBS
302 & 80
PFMBA
863090-89-5
4.01
279 & 85
13C5-PFPeA
268 & 223
PFEESA
113507-82-7
4.12
315 & 135
13C3-PFBS
302 & 80
NFDHA
151772-58-6
4.25
295 & 85
13C5-PFHxA
318 & 273
4:2 FTSA
757124-72-4
4.29
327 & 307
13C2-4:2 FTSA
329 & 309
PFHxA
307-24-4
4.33
313 & 269
13C5-PFHxA
318 & 273
PFPeS
2706-91-4
4.39
349 & 80
13C3-PFHxS
402 & 80
HFPO-DA
13252-13-6
4.48
285 & 169
13C3-HFPO-DA
287 & 169
FBSA
30334-69-1
4.58
298 & 78
13C3-PFHxS
402 & 80
PFHpA
375-85-9
4.90
363 & 319
13C4-PFHpA
367 & 322
PFHxS
355-46-4
4.95
399 & 80
13C3-PFHxS
402 & 80
DONA
919005-14-4
4.98
377 & 251
13C4-PFHpA
367 & 322
5:3 FTCA
914637-49-3
5.04
341 & 237
13C5-PFHxA
318 & 273
6:2 FTUCA
70887-88-6
5.06
357 & 293
13C2-6:2 FTUCA
359 & 294
6:2 FTCA
53826-12-3
5.11
377 & 293
13C2-6:2 FTCA
379 & 294
4-PFecHS
646-83-3
5.52
461 & 381
13C8-PFOS
507 & 80
6:2 FTSA
27619-97-2
5.57
427 & 407
13C2-6:2 FTSA
429 & 409
PFOA
335-67-1
5.60
413 & 369
13C8-PFOA
421 & 376
PFHpS
375-92-8
5.63
449 & 80
13C8-PFOS
507 & 80
FHxSA
41997-13-1
6.10
398 & 78
13C8-PFOS
507 & 80
PFNA
375-95-1
6.15
463 & 419
13C9-PFNA
472 & 427
PFOS
1763-23-1
6.17
499 & 80
13C8-PFOS
507 & 80
8:2 FTUCA
70887-84-2
6.36
457 & 393
13C2-8:2 FTUCA
459 & 394
7:3 FTCA
812-70-4
6.37
441 & 337
13C5-PFHxA
318 & 273
8:2 FTCA
27854-31-5
6.40
477 & 393
13C2-8:2 FTCA
479 & 394
9Cl-PF3ONS
756426-58-1
6.55
531 & 351
13C8-PFOS
507 & 80
8:2 FTSA
39108-34-4
7.02
527 & 507
13C2-8:2 FTSA
529 & 509
PFDA
335-76-2
7.07
513 & 469
13C6-PFDA
519 & 474
PFNS
68259-12-1
7.09
549 & 80
13C8-PFOS
507 & 80
N-MeFOSAA
2355-31-9
7.73
570 & 419
d3-N-MeFOSAA
573 & 419
PFDS
335-77-3
8.10
599 & 80
13C8-PFOS
507 & 80
PFUnDA
2058-94-8
8.11
563 & 519
13C7-PFUnDA
570 & 525
N-EtFOSAA
2991-50-6
8.13
584 & 419
d5-N-EtFOSAA
589 & 419
PFOSA
754-91-6
8.32
498 & 78
13C8-PFOSA
506 & 78
10:2 FTUCA
70887-94-4
8.36
557 & 493
13C2-10:2 FTUCA
559 & 494
10:2 FTCA
53826-13-4
8.40
577 & 493
13C2-10:2 FTCA
579 & 494
Appendix A
Table
Target Compound
CAS
Ret Time (min)
Target Quant Ion MRM Transition
Isotope Analog
Isotope MRM Transition
6:6 PFPi
40143-77-9
8.90
701 & 401
13C2-PFDoDA
615 & 570
PFDoS
79780-39-5
9.09
699 & 80
13C8-PFOS
507 & 80
PFTrDA
72629-94-8
9.12
663 & 619
13C2-PFDoDA
615 & 570
PFTDA
376-06-7
9.49
713 & 669
13C2-PFTDA
715 & 670
N-MeFOSA
31506-32-8
9.50
512 & 219
d3-N-MeFOSA
515 & 169
FDSA
N/A
9.52
598 & 78
13C8-PFOSA
506 & 78
6:8 PFPi
610800-34-5
9.54
801 & 401
(13C2)2-6:2 diPAP
993 & 97
MeFOSE
24448-09-7
9.54
616 & 59
d7-MeFOSE
623.1 & 59
N-EtFOSA
4151-50-2
9.88
526 & 219
d5-N-EtFOSA
531 & 169
EtFOSE
1691-99-2
9.89
630 & 59
d9-EtFOSE
639.1 & 59
6:2/8:2 diPAP
943913-15-3
10.02
889 & 443
(13C2)2-6:2 diPAP
793 & 97
8:8 PFPi
40143-79-1
10.11
901 & 501
(13C2)2-6:2 diPAP
793 & 445
PFHxDA
67905-19-5
10.18
813 & 269
13C2-PFHxDA
815 & 770
8:2 diPAP
678-41-1
10.55
989 & 543
(13C2)2-8:2 diPAP
993 & 97
PFODA
16517-11-6
10.81
913 & 369
13C2-PFHxDA
815 & 770
diSAmPAP
2965-52-8
11.10
1,203 & 526
(13C2)2-8:2 diPAP
993 & 97
Engineering Facts
11 facts · every claim linked to a page
Selection
Choose Carbon S for PFAS cartridges when both low PFAS background and high recovery of GCB-sensitive analytes are required.
It justifies selecting Carbon S over conventional GCB for PFAS soil cleanup.
Source
“Carbon S sorbent provides a better balance between analyte recovery and pigment removal efficiency than traditional GCB sorbent.”
Selection
The cartridge is manufactured specifically for PFAS work so common PFAS contaminants stay below typical reporting limits.
Low-residue consumables are essential to reach ng/g detection limits without false positives.
Source
“Bond Elut Carbon S for PFAS has been manufactured for PFAS applications, ensuring that common PFAS contaminants are below typical reporting limits.”
Installation
Use a PFC delay column in the LC flow path to keep PFAS background under control.
The delay column is a required hardware item for low-level PFAS analysis.
Source
“InfinityLab PFC delay column, 4.6 × 30 mm”
Operation
Condition the Carbon S cartridge with two 5 mL methanol rinses and discard the rinsate before loading sample extract.
Proper conditioning prevents carryover and ensures consistent cleanup performance.
Source
“Rinse the Carbon S cartridge with two 5 mL volumes of methanol and discard.”
Operation
Load the extract in two portions and elute under gravity or slight vacuum to avoid forcing matrix through the sorbent.
Controlled flow protects recovery and pigment removal efficiency.
Source
“Decant the extract into Carbon S cartridges in two portions, elute under gravity or with a slight vacuum.”
Operation
Neutralize the collected eluate with 50 µL acetic acid before combining with internal standard and water.
Neutralization is required before LC/MS/MS injection for consistent analyte response.
Source
“Neutralize with 50 µL of acetic acid.”
Performance
Expect average recovery near 99% with RSD around 13.5% at low spike levels for the 59 PFAS tested.
This sets realistic acceptance criteria for method validation.
Source
“The average recovery for all compounds was 99.3% with an RSD of 13.5%.”
Performance
Carbon S cleanup improves early-eluting PFBA peak shape and retention time consistency in peat extracts.
Matrix cleanup directly affects chromatographic data quality for early eluters.
Source
“the use of Carbon S improved the peak shape integrity and retention consistency of PFBA compared to extracts without the use of Carbon S cleanup.”
Cumplimiento
The extraction procedure follows ASTM D7968-17a except that syringe filtration is replaced by Carbon S passthrough cleanup.
Knowing the deviation from the validated standard is needed for method documentation and compliance.
Source
“The sample preparation closely followed the extraction procedure in ASTM D7968‑17a except for replacing the syringe filtration step with a passthrough cleanup using the Carbon S for PFAS cartridge.”
Maintenance
Run extraction blanks and cartridge rinsate blanks with each extraction set to confirm the system and cartridges are PFAS-free.
Blank monitoring is the primary ongoing contamination control measure.
Source
“Cartridge blanks (rinsate collected from the methanol rinse) and matrix blanks were also analyzed to ensure the system and cartridges were free from PFAS contamination before sample analysis.”
Operation
Use a tube rotator to fully invert sample tubes during the 1-hour extraction for reproducible contact.
Consistent agitation is needed to meet recovery and precision targets.
Source
“A multipurpose rotator model 150 (Scientific Industries, Springfield, MA) tube rotator was used to fully invert the sample tubes during extraction.”