Forced Degradation Studies of Synthetic Oligonucleotide
Technical Reference Summary
Covered Models & Document Scope
Nota de aplicación sobre estudios de degradación forzada de Fomivirsen utilizando un sistema bio LC Agilent 1290 Infinity II acoplado a un LC/Q-TOF Agilent 6545XT AdvanceBio; abarca condiciones de estrés térmico (80 °C, 4/8/24 h) y estrés oxidativo (0.30% y 3.00% H2O2, 2/4 h).
Technical Data
| Parameter | Value | Condición |
|---|---|---|
| Solvent A composition | 15 mM TEA and 100 mM HFIP in water p.2 ↗ | |
| Sample concentration | 1 mg/mL p.2 ↗ | |
| Heat stress incubation temperature | 80 °C p.2 ↗ | |
| Heat stress incubation time points | 4, 8, and 24 hours p.2 ↗ | |
| Oxidative stress H2O2 concentration | 0.30 and 3.00 % p.2 ↗ | |
| Oxidative stress duration | 2 and 4 hours p.2 ↗ | |
| Source | Dual AJS p.3 ↗ | |
| Gas Flow | 12 L/min p.3 ↗ | |
| Nebulizer | 35 psi p.3 ↗ | |
| Sheath Gas Flow | 12 L/min p.3 ↗ | |
| Skimmer | 65 V p.3 ↗ | |
| Tune Mode | Extended Dynamic Range (2 GHz) p.3 ↗ | |
| Acquisition Mode | MS Scan p.3 ↗ | |
| Mass Range | m/z 300 to 3,200 p.3 ↗ | |
| MS Scan Range | m/z 400 to 3,000 p.3 ↗ | |
| Software version | 1.5 p.3 ↗ | Agilent MassHunter BioConfirm Software |
| Match Tolerance | Mass ±10 ppm p.3 ↗ |
Operating Limits
- Sample storage temperature: –80 °C p.2 ↗
- Heat stress incubation temperature: 80 °C p.2 ↗
- Oxidative stress H2O2 concentration: 0.30 and 3.00% p.2 ↗