Induztr
Resumen técnico Visor de documentos Modelos relacionados

Fast and Robust GC/MS/MS Analysis of Pesticides

Contenido

  1. Overview pp. 2-3
    Overview of GC/MS/MS pesticide analysis goals and system options
  2. Config. p. 4
    Instrument conditions for 8890 GC with 7000E/7010 GC/TQ
  3. Procedure p. 5
    Sample preparation procedure for pesticide residue analysis
  4. pp. 6-11
    Chromatograms, calibration, sensitivity and method robustness results
  5. Overview p. 12
    Summary of method performance and suitability
  6. pp. 13-14
    Compound list and MRM transition parameters
  7. p. 15
    Agilent contact and website information
  8. p. 4 table, 40 rows
  9. p. 13 table, 38 rows
  10. p. 14 table, 40 rows
  11. p. 15 table, 23 rows

Tabelle 1: Bedingungen für die Systemkombinationen Agilent 8890 GC und 7000 GC/TQ bzw. Agilent 8890 GC und 7010C GC/TQ für 10-Minuten-Analysen von Pestiziden.

ParameterMit 15 x 15 mMit 10 x 10 m
GC
Agilent 8890 GC (220-V-Ofen) mit schnellem Ofen, automatischem Probengeber und Probenteller
Einlass
Multimode-Einlass (MMI)
Modus
Cold Splitlos
Purge-Fluss zum Splitauslass60 ml/min bei 0,75 min
Septumspülungs-fluss3 ml/min
Modus Septums­pülungsflussGeschaltet
Injektionsvolumen1,0 µl
InjektionstypStandard
L1 Airgap0,2 µl
GassparflussEinschaltung mit 30 ml/min nach 3 min
Einlasstemperatur60 °C für 0,1 min, dann bis 280 °C mit 600 °C/min
Einlasstemperatur Nachlauf310 ºC
Gesamtfluss Nachlauf25 ml/min
TrägergasHelium
Einlass-LinerAgilent Ultra Inert 2 mm Liner mit Verwirbelungszone
Einlass-Liner Bestellnummer5190-2297
Ofen
Anfängliche Ofentemperatur60 °C60 °C
Anfängliche Ofen-Haltezeit1 min0,5 min
Anstiegsrate 180 °C/min80 °C/min
Endtemperatur 1170 °C170 °C
Endtemperatur-Haltezeit 10 min0 min
Anstiegsrate 235 °C/min20 °C/min
Endtemperatur 2310 °C310 °C
Endtemperatur-Haltezeit 23,625 min1,125 min
Gesamtlaufzeit10 min10 min
Nachlaufzeit1,5 min1,5 min
Äquilibrierungszeit0,25 min0,25 min
Einsatz für Hochge-schwindig-keitsofen (Kissen)
Säule 1
TypAgilent J&W HP-5ms Ultra InertAgilent J&W HP-5ms Ultra Inert
Agilent Bestellnummer19091S- 431UI-KEYSpezielle Säule2
Länge15 m10 m
Durchmesser0,25 mm0,25 mm
Filmdicke0,25 µm0,25 µm
SteuerungsmodusKonstanter FlussKonstanter Fluss
Fluss1,016 ml/min1,3 ml/min

2 Spezielle Säule

Die in dieser Arbeit verwendeten Verbindungen und deren beobachtete Retentionszeiten mit einer Zwei-Säulen-Konfiguration und 10-Minuten-Trennungen.

NameRetentionszeit (min) 15 x 15 mRetentionszeit (min) 10 x 10 mNameRetentionszeit (min) 15 x 15 mRetentionszeit (min) 10 x 10 m
Allidochlor3,7732,542BHC-gamma (Lindan, gamma-HCH)5,2014,174
Dichlorbenzonitril, 2,6-3,9722,720Pyrimethanil5,2224,246
Biphenyl4,0552,812Tefluthrin5,2234,310
Mevinphos, E-4,1102,901Fonofos5,2254,223
3,4-Dichloranilin4,1932,954Pentachlornitrobenzol5,2274,210
Pebulat4,2233,006Pentachlorbenzonitril5,2474,228
Etridiazol4,2463,016Disulfoton5,2734,312
N-(2,4-dimethylphenyl)formamid4,3053,091Isazofos5,2854,361
cis-1,2,3,6-Tetrahydrophthalimid4,3123,090Terbacil5,2854,323
Methacrifos4,3213,129Triallat5,3224,379
Chloroneb4,3753,171BHC-delta5,3304,351
2-Phenylphenol4,4443,228Chlorthalonil5,3504,392
Pentachlorbenzol4,4953,276Propanil5,4634,570
Propachlor4,7023,546Endosulfanether5,4664,523
Tecnazen4,7123,547Transfluthrin5,4764,658
Diphenylamin4,7343,582Dimethachlor5,4774,596
Cycloat4,7573,626Pentachloranilin5,4824,552
Chlorpropham4,7693,656Acetochlor5,5024,641
2,3,5,6-Tetrachloranilin4,7933,633Vinclozolin5,5034,654
Trifluralin4,7983,724Parathion-methyl5,5264,668
Benfluralin4,8113,740Chlorpyrifos-methyl5,5264,668
Ethalfluralin4,8123,670Tolclofos-methyl5,5594,710
Sulfotep4,8693,789Alachlor5,5644,725
Diallat I4,9283,846Propisochlor5,5794,765
Phorat4,9323,852Metalaxyl5,5834,763
BHC-beta5,0104,115Ronnel5,6144,791
BHC-alpha (Benzolhexachlorid)5,0113,918Prodiamin5,6224,871
Hexachlorbenzol5,0693,987Heptachlor5,6304,763
Atrazin5,0724,048Pirimiphos-methyl5,6504,892
Dichloran5,0723,998Fenitrothion5,6764,891
Pentachloranisol5,0834,013Malathion5,6964,962
Clomazon5,1224,092Linuron5,7084,927
Profluralin5,1234,156Dichlofluanid5,7454,980
Terbuthylazin5,1554,163Pentachlorthioanisol5,7674,972
Terbufos5,1734,178Aldrin5,7685,061
Propyzamid5,1754,188Fenthion5,7795,057
Diazinon5,1914,244Metolachlor5,7835,046
Fluchloralin5,1994,261Chlorpyrifos5,7905,075

Anhang 1

Retentionszeiten (min) für 15 x 15 m und 10 x 10 m

NameRetentionszeit (min) 15 x 15 mRetentionszeit (min) 10 x 10 mNameRetentionszeit (min) 15 x 15 mRetentionszeit (min) 10 x 10 m
Parathion5,7935,081Chlorfenson6,2755,784
Triadimefon5,8115,100Nonachlor, trans-6,2795,787
DCPA (Dacthal, Chlorthal-dimethyl)5,8295,124Dieldrin6,2795,955
Anthrachinon5,8315,053Fludioxonil6,2945,876
Dichlorbenzophenon, 4,4'-5,8405,110Prothiofos6,3005,844
Pirimiphos-ethyl5,8695,241Oxadiazon6,3035,920
MGK-2645,8815,315Pretilachlor6,3035,895
Isopropalin5,8985,267Iodofenphos6,3045,828
Fenson5,9025,194Profenofos6,3125,877
Diphenamid5,9085,235Oxyfluorfen6,3145,960
Bromophos5,9185,237DDE-p,p'6,3425,906
Cyprodinil5,9415,314Bupirimat6,3616,014
Pendimethalin5,9755,356Myclobutanil6,3645,970
Chlozolinat5,9765,378Chlorfenapyr6,3656,122
Allethrin5,9795,393Flusilazol6,3705,995
Triflumizol5,9795,473Fluazifop-P-butyl6,3886,090
Fipronil5,9935,431DDD-o,p'6,4045,990
Penconazol5,9985,375Tricyclazol6,4125,932
Metazachlor5,9995,358Endrin6,4236,153
Chlorfenvinphos6,0165,436Ethylan6,4536,121
Heptachlorexoepoxid6,0165,402Nitrofen6,4776,101
Isodrin6,0185,319Chlorbenzilat6,5066,189
Captan6,0205,472Ethion6,5716,315
Tolylfluanid6,0265,413DDD-p,p'6,5826,280
Bromfenvinfos-methyl6,0365,436DDT-o,p'6,5826,318
Quinalphos6,0475,463Chlorthiophos6,5876,338
Triadimenol6,0535,476Endosulfan II (beta-Isomer)6,6036,235
Procymidon6,0905,515Triazophos6,6446,428
Folpet6,1275,513Sulprofos6,6596,420
Paclobutrazol6,1375,653Nonachlor, cis-6,6676,341
Chlorbensid6,1375,549Carfentrazon-ethyl6,6686,509
Bromophos-ethyl6,1395,609Methoxychlor-olefin6,7026,519
DDE-o,p'6,1765,631Endrinaldehyd6,7096,402
Tetrachlorvinphos6,1815,680Carbophenothion6,7266,513
Chlordan-trans6,1875,610Norflurazon6,7546,576
Chlordan-cis6,1965,744Edifenphos6,7866,566
Fenamiphos6,2275,797Lenacil6,7876,588
Flutolanil6,2335,801DDT-p,p'6,8056,615
Bromfenvinfos6,2525,800Iprodion6,8266,947
Flutriafol6,2555,764Methoxychlor, o,p'-6,8466,703

Table

NameRetentionszeit (min) 15 x 15 mRetentionszeit (min) 10 x 10 mNameRetentionszeit (min) 15 x 15 mRetentionszeit (min) 10 x 10 m
Piperonylbutoxid6,8546,788Acrinathrin7,4157,607
Propargit6,8566,760Leptophos7,4177,413
Resmethrin6,8576,756Pyrazophos7,5567,660
Hexazinon6,8616,708Fenarimol7,6317,641
Tebuconazol6,8866,739Mirex7,6367,533
Captafol6,8906,805Pyraclofos7,6457,728
Nitralin6,9136,862Azinphos-ethyl7,6757,700
Bifenthrin7,0447,057Permethrin, (1R)-cis-7,7857,901
Pyridaphenthion7,0487,004Permethrin, (1R)-trans-7,8427,962
Tetramethrin I7,0526,999Pyridaben7,9167,980
Fenpropathrin7,1067,121Coumaphos7,9648,028
Bromopropylat7,1097,061Fluquinconazol7,9648,023
EPN7,1127,061Prochloraz7,9888,058
Tebufenpyrad7,1307,152Cyfluthrin I8,1578,184
Methoxychlor, p,p'-7,1317,111Cypermethrin I8,2508,339
Phosmet7,1357,054Flucythrinat I8,3598,444
Endrinketon7,1897,033Acequinocyl8,4098,534
Phenothrin I7,2307,243Ethofenprox8,4318,485
Azinphos-methyl7,3307,405Fluridon8,7088,662
Tetradifon7,3307,305Fenvalerat I8,8818,799
Cyhalothrin (lambda)7,3347,438Fluvalinat-tau I8,9708,894
Pyriproxyfen7,3587,406Deltamethrin9,4449,166
Phosalon7,3897,387

Engineering Facts

14 facts · every claim linked to a page
Selection

The 15x15 m configuration requires no hardware changes, while the 10x10 m minibore configuration requires a new column set and a GC oven insert (cushion).

Determines upfront hardware investment and whether an existing GC can be converted without additional parts.

Source
“Für diese Konfiguration waren gegenüber der herkömmlichen 15 x 15 m-Konfiguration ein neuer Säulensatz und ein GC-Ofen-Einsatz (ein Kissen) erforderlich.”
Performance

Mid-column backflush preserves chromatographic resolution and allows 203 pesticides to be analyzed in 10 minutes with three MRM transitions per compound.

Confirms the method meets multi-residue screening requirements without sacrificing confirmatory transitions.

Source
“ermöglichten eine 10-Minuten-Analyse von 203 Pestiziden mit drei MRM-Übergängen pro Verbindung”
Maintenance

Mid-column backflush reduces maintenance frequency by minimizing column trimming and ion source cleaning.

Lower maintenance translates to higher uptime and lower cost of ownership for high-throughput labs.

Source
“Backflush in der Mitte der Säulenanordnung fördert nachweislich die Robustheit der Methode und verlängert den wartungsfreien Betrieb, da das Kürzen von Säulen und die Reinigung der Ionenquelle auf ein Minimum begrenzt werden.”
Operation

The MMI inlet uses a cold splitless mode with a 60 ml/min purge flow at 0.75 min and a 3 ml/min septum purge.

These inlet parameters must be programmed correctly to avoid analyte loss and ensure reproducible injections.

Source
“Purge-Fluss zum Splitauslass 60 ml/min bei 0,75 min”
Installation

The inlet liner specified is an Agilent Ultra Inert 2 mm liner with a swirl zone, part number 5190-2297, and no glass wool.

Using the correct inert liner is critical for labile pesticides and prevents breakdown of sensitive analytes.

Source
“Agilent Ultra Inert 2 mm Liner mit Verwirbelungszone”
Operation

The oven program for the 15x15 m configuration uses 80 °C/min to 170 °C then 35 °C/min to 310 °C with a 3.625 min hold, totaling 10 min.

The ramp rates and hold times must be entered exactly to reproduce retention times and resolution.

Source
“Anstiegsrate 2 35 °C/min”
Installation

Column 1 and column 2 are both Agilent J&W HP-5ms Ultra Inert with 0.25 mm diameter and 0.25 µm film for the 15x15 m setup.

Column phase and dimensions must match the method to maintain retention time locking and elution order.

Source
“Agilent J&W HP-5ms Ultra Inert”
Operation

The MSD uses an inert extractor ion source with 3-mm lens or HES, a solvent delay of 3 min, and quad temperatures of 150 °C.

Solvent delay protects the filament and detector; incorrect settings shorten source lifetime.

Source
“Solvent Delay 3 min”
Operation

The dMRM method uses helium quench gas at 2.25 ml/min and nitrogen collision gas at 1.5 ml/min.

Collision and quench gas flows directly affect MRM sensitivity and must be set per the method.

Source
“Quenchgas He 2,25 ml/min”
Selection

The 7010C with High Efficiency Source was shown to be superior in sensitivity to the 7000 GC/TQ, enabling quantification down to 0.1 ppb.

Guides instrument selection when low-ppb quantification of difficult analytes like deltamethrin is required.

Source
“Das 7010C GC/TQ-System mit High Efficiency Source (HES) war hinsichtlich der Empfindlichkeit dem 7000 GC/TQ nachweislich überlegen.”
Maintenance

During 700 injections over 175 hours, the only maintenance performed was septum and liner replacement every 100 injections.

Sets realistic preventive maintenance intervals and confirms no source cleaning or column trimming is needed.

Source
“Die einzige Wartungsprozedur, die während der Robustheitsprüfung durchgeführt wurde, war ein Wechsel von Septum und Liner alle 100 Injektionen.”
Operation

Best practices include leak-free GC/TQ connections using self-tightening column nuts with screw fixation and gold-plated flexible CFT metal ferrules.

Leak-free connections are essential for retention time stability and to avoid oxygen damage to the column.

Source
“Leckagefreies GC/TQ-System dank selbstsichernder Säulenmuttern mit Schraubfixierung und vergoldeten flexiblen CFT-Metall-Ferrulen.”
Operation

The sample preparation uses 10 g homogenized fresh spinach extracted with 10 ml ACN containing 1% acetic acid via QuEChERS EN.

Exact extraction solvent and ratio must be followed to achieve consistent recoveries and matrix removal.

Source
“10 ml ACN mit 1 % Essigsäure wurde hinzugefügt.”
Installation

The Captiva EMR–HCF1 cartridge (part number 5610-2088) is used for cleanup of high-chlorophyll spinach matrix.

Selecting the correct EMR cartridge is critical for removing chlorophyll and protecting the GC inlet and column.

Source
“Captiva EMR–HCF1 Kartusche (Bestellnummer 5610-2088)”