A Fast and Robust GC/MS/MS Analysis of 203 Pesticides in 10 Minutes in Spinach
Technical Reference Summary
Covered Models & Document Scope
Nota de aplicación que cubre dos métodos rápidos de GC/MS/MS para el análisis de 203 pesticidas en espinaca: (a) configuración de retroflujo en columna intermedia de 15 m × 15 m (0.25 mm × 0.25 µm) y (b) configuración de retroflujo en columna intermedia minibore de 10 m × 10 m (0.18 mm × 0.18 µm), en GC Agilent 8890 con triple cuadrupolo GC/MS 7000E o 7010C.
Technical Data
| Parameter | Value | Condición |
|---|---|---|
| Analysis time | 10 min p.1 ↗ | 15 m × 15 m configuration |
| Analysis time | 10 min p.1 ↗ | 10 m × 10 m configuration |
| Number of pesticides analyzed | 203 p.1 ↗ | |
| Calibration range | 0.1 to 1,000 ppb p.1 ↗ | 7010C and 7000 series |
| Calibration correlation coefficient | R2 > 0.99 p.10 ↗ | linear or quadratic fit |
| Number of consecutive injections (robustness) | 700 injections p.1 ↗ | spinach extract spiked at 20 ppb |
| Continuous running time (robustness) | 175 hours p.1 ↗ | 700 injections |
| Liner and septum replacement interval | every 100 injections p.2 ↗ | |
| GC oven voltage | 220 V p.4 ↗ | Agilent 8890 GC |
| Column 1 length | 15 m p.4 ↗ | 15 m × 15 m configuration |
| Column 1 diameter | 0.25 mm p.4 ↗ | 15 m × 15 m configuration |
| Column 1 film thickness | 0.25 µm p.4 ↗ | 15 m × 15 m configuration |
| Injection volume | 1.0 µL p.4 ↗ | |
| Inlet temperature | 60 °C for 0.1 min, then 310 °C p.4 ↗ | |
| Initial oven temperature | 60 °C p.4 ↗ | 15 m × 15 m configuration |
| Initial oven temperature | 60 °C p.4 ↗ | 10 m × 10 m configuration |
| Ramp rate 1 | 80 °C/min p.4 ↗ | 15 m × 15 m configuration |
| Ramp rate 1 | 80 °C/min p.4 ↗ | 10 m × 10 m configuration |
| Final temperature 2 | 310 °C p.4 ↗ | 15 m × 15 m configuration |
| Final temperature 2 | 310 °C p.4 ↗ | 10 m × 10 m configuration |
| Total run time | 10 min p.4 ↗ | 15 m × 15 m configuration |
Operating Limits
- Minimum calibration concentration: 0.1 ppb p.1 ↗
- Maximum calibration concentration: 1,000 ppb p.1 ↗
- Maximum number of consecutive injections demonstrated: 700 injections p.1 ↗
- Maximum continuous running time demonstrated: 175 hours p.1 ↗
- Liner and septum replacement interval: every 100 injections p.2 ↗