Compound Group / Compound Name / Precursor Ion / Product Ion / RT (min) / Fragmentor / Collision Energy / Cell Accelerator Voltage
Compound Group
Compound Name
Precursor Ion
Product Ion
RT (min)
Fragmentor
Collision Energy
Cell Accelerator Voltage
Sulfonate
PFHxS
398.9
80
6.4
100
49
2
Sulfonate
PFHpS
448.9
98.7
6.9
100
44
2
Sulfonate
PFHpS
448.9
80
6.9
148
50
2
Sulfonate
PFOS
498.9
99
7.3
100
50
2
Sulfonate
PFOS
498.9
80
7.3
100
50
2
Sulfonate
PFNS
548.9
99
7.7
148
52
2
Sulfonate
PFNS
548.9
80
7.7
148
56
2
Sulfonate
PFDS
598.9
99
8.1
148
56
2
Sulfonate
PFDS
598.9
80
8.1
148
60
2
Sulfonate
PFDoS
698.9
99
8.8
156
62
2
Sulfonate
PFDoS
698.9
80
8.8
156
67
2
Surrogate
13C4-PFBA
217
172
3.5
60
8
2
Surrogate
13C5-PFPeA
268
223
4.5
60
8
2
Surrogate
13C2-4:2FTS
329
309
5.0
125
20
2
Surrogate
13C2-4:2FTS
329
81
5.0
150
32
2
Surrogate
13C5-PFHxA
318
273
5.2
70
8
2
Surrogate
13C5-PFHxA
318
120
5.2
72
24
2
Surrogate
13C3-PFBS
302
99
5.3
130
32
2
Surrogate
13C3-PFBS
302
80
5.3
130
44
2
Surrogate
13C3-HFPO-DA
287
185
5.4
64
20
5
Surrogate
13C3-HFPO-DA
287
169
5.4
64
4
5
Surrogate
13C4-PFHpA
367
322
5.7
72
4
2
Surrogate
13C4-PFHpA
367
169
5.7
72
16
2
Surrogate
13C2-6:2FTS
429
409
6.0
125
24
2
Surrogate
13C2-6:2FTS
429
81
6.0
150
40
2
Surrogate
13C8-PFOA
421
376
6.2
69
4
2
Surrogate
13C8-PFOA
421
172
6.2
72
20
2
Surrogate
13C3-PFHxS
402
99
6.4
156
44
2
Surrogate
13C3-PFHxS
402
80
6.4
100
45
2
Surrogate
13C9-PFNA
472
427
6.6
66
4
2
Surrogate
13C9-PFNA
472
223
6.6
72
16
2
Surrogate
13C2-8:2FTS
529
509
6.8
170
28
2
Surrogate
13C2-8:2FTS
529
81
6.8
200
52
4
Surrogate
13C6-PFDA
519
474
7.0
72
8
2
Surrogate
D3-NMeFOSAA
573.2
419
7.1
150
20
2
Surrogate
D5-NEtFOSAA
589.2
419
7.2
100
20
2
Surrogate
13C8-PFOS
507
99
7.3
148
52
2
Surrogate
13C8-PFOS
507
80
7.3
100
50
2
Surrogate
13C7-PFUnA
570
525
7.5
100
8
2
Surrogate
13C2-PFDoA
615
570
7.9
90
12
2
Engineering Facts
13 facts · every claim linked to a page
Selection
The method requires weak anion exchange (WAX) SPE and carbon cleanup for all matrices, so the cartridge chemistry must be WAX and a carbon cleanup step must be included.
Determines the SPE sorbent chemistry and the need for a carbon cleanup consumable when selecting the extraction workflow.
Source
“all the methods within EPA Method 1633 specify SPE using weak anion exchange (WAX) and carbon cleanup.”
Selection
The 150 mg, 6 mL Bond Elut PFAS WAX cartridge is the format used for aqueous matrices in this method, so this bed mass and volume should be selected for 500 mL aqueous samples.
Specifies the exact cartridge size to choose for method-compliant aqueous PFAS extraction.
Source
“using a 150 mg, 6 mL Agilent Bond Elut PFAS WAX cartridge for SPE and Carbon S for cleanup.”
Selection
Bond Elut PFAS WAX cartridges are PFAS-specific and undergo PFAS-specific QC testing, so they should be chosen over general-purpose WAX cartridges to control background contamination.
Justifies selecting PFAS-certified cartridges rather than generic WAX to avoid false positives and poor recovery.
Source
“The cartridges undergo PFAS-specific QC testing to ensure optimal cleanliness and recovery.”
Installation
A PFAS-specific delay column must be installed between the pump and the multisampler to separate background contaminants without significantly increasing backpressure.
Defines the required LC plumbing position of the delay column to keep PFAS background low without overpressuring the system.
Source
“The PFAS-specific delay column was placed between the pump and multisampler to separate background contaminants from compounds originating in the sample vial without significantly increasing backpressure.”
Operation
Samples must be loaded onto the SPE cartridge at a slow flow rate of 5 mL/min, so the loading step must be flow-controlled rather than run at full vacuum.
Sets the maximum sample loading flow to preserve analyte recovery during extraction.
Source
“Passthrough cartridge at 5 mL/min.”
Operation
The SPE cartridge must be conditioned with 15 mL of 1% methanolic ammonium hydroxide followed by 5 mL of 0.3 M formic acid before loading the sample.
Defines the conditioning sequence and volumes needed to activate the WAX sorbent correctly.
Source
“15 mL of 1% methanolic ammonium hydroxide”
Operation
After elution, 25 µL of concentrated acetic acid must be added to each eluate and vortexed before Carbon S cleanup.
Specifies the acidification step required before carbon cleanup to achieve the correct recovery.
Source
“Add 25 µL of concentrated acetic acid to each sample eluate and vortex.”
Operation
Carbon S cleanup uses 10 mg per sample with hand-shaking for less than 5 minutes, vortexing for 30 seconds, and centrifugation for 10 minutes at 2,800 rpm.
Gives the exact carbon dose and mixing/centrifugation conditions needed for reproducible matrix removal.
Source
“Add 10 mg of Carbon S to each sample.”
Operation
The final extract must be filtered through a nylon syringe filter into a collection tube containing the nonextracted internal standard (NIS).
Ensures the NIS is added at the correct point and that particulates are removed before LC/TQ analysis.
Source
“Filter the entire extract into a NIS collection tube and vortex.”
Performance
The chromatographic separation between TDCA and PFOS must be at least 1 minute; the conditions used achieved 5.7 and 7.1 minutes respectively.
Sets a mandatory chromatographic resolution criterion that the LC method and column must satisfy.
Source
“EPA Method 1633 specifies that the chromatographic separation between the bile salts (TDCA) and PFOS must be at least 1 minute.”
Performance
Initial MDLs with Bond Elut PFAS WAX SPE were equivalent to or better than the method-documented MDLs for most analytes, supporting method-compliant sensitivity.
Confirms the cartridge can meet required detection limits, which is a key selection criterion for trace PFAS work.
Source
“Initial MDLs achieved following extraction of low-level spikes and lab reagent blanks (LRB) with the Bond Elut PFAS WAX SPE are equivalent or better than those documented in the method.”
Performance
In wastewater effluent, the FTS class of extracted surrogates had consistently high recoveries, showing that matched labeled internal standards are needed to correct matrix effects.
Warns that matrix effects can bias FTS results unless appropriate isotope-labeled internal standards are used.
Source
“However, the FTS class of PFAS extracted surrogates had consistently high recoveries in matrix, demonstrating the importance of matched and labeled internal standards to accurately correct for matrix effects.”
Cumplimiento
The method was followed closely with only permitted changes in supplies and instrumentation parameters, so any deviation in consumables or instrument settings must be within the method's allowed changes.
Reminds the engineer that consumable and instrument substitutions must remain within EPA Method 1633 allowances for compliance.
Source
“The method was followed closely; only permitted changes in supplies and instrumentation parameters were made.”