Quantitative Analysis of 210 Veterinary Drugs in Organ Meat Using the Agilent 6470A Triple Quadrupole LC/MS
Technical Reference Summary
Covered Models & Document Scope
Nota de aplicación que cubre la Agilent Comprehensive Veterinary Drug dMRM Solution (G5368AA) para el cribado de 210 residuos de medicamentos veterinarios en carne de órganos (riñón e hígado de pollo) utilizando el Agilent 6470 Triple Quadrupole LC/MS con el sistema 1290 Infinity II LC; no se describen variantes de modelo de instrumento ni códigos de pedido.
Technical Data
| Parameter | Value | Condición |
|---|---|---|
| Number of target residues | 210 p.1 ↗ | |
| Number of chemical classes covered | >28 p.1 ↗ | |
| Sample amount | 2.0 ±0.1 g p.2 ↗ | homogenized chicken kidney and liver |
| Retention time precision (%RSD) | <0.6 % p.3 ↗ | duplicate injections of technical preparations, each QC level |
| Response precision (%RSD) | <20 % p.3 ↗ | duplicate injections of technical preparations, each QC level |
| Accuracy | 70 to 120 % p.3 ↗ | kidney and liver QC levels |
| Target recovery | 60 to 120 % p.3 ↗ | |
| QC RSD | <20 % p.5 ↗ | |
| Calibration linearity R2 | >0.99 p.5 ↗ | LOD range |
| LOD range | 0.1–10 µg/kg p.5 ↗ | kidney and liver, per chemical class |
| MRL range | 0.1–2,000 µg/kg p.5 ↗ | kidney and liver, per chemical class |
| Total targets with established MRL | 91 p.5 ↗ | |
| Targets detected in kidney | 206 p.5 ↗ | |
| Targets detected in liver | 195 p.5 ↗ | |
| Targets meeting recovery criteria in kidney | 189 p.5 ↗ | |
| Targets meeting recovery criteria in liver | 206 p.5 ↗ | |
| Targets with negligible interference | about 92 % p.4 ↗ | MRL-based screening |
Operating Limits
- Retention time precision (%RSD): <0.6% p.3 ↗
- Response precision (%RSD): <20% p.3 ↗
- Accuracy: 70% to 120% p.3 ↗
- Target recovery: 60 to 120% p.3 ↗
- QC RSD: <20% p.5 ↗
- Calibration linearity R2: >0.99 p.5 ↗