Induztr
Resumen técnico Visor de documentos Modelos relacionados

Residual Solvents Analysis Based on Agilent DB-Select 624 UI Columns

Contenido

  1. Overview p. 2
    Background on residual solvents analysis and regulatory context
  2. Spec pp. 3-7
    Chromatographic results for class 1, 2A, and 2B solvents
  3. Quality p. 8
    Retention time and area repeatability on both column types
  4. Procedure pp. 9-10
    Single run analysis of combined residual solvent classes
  5. Services pp. 11-13
    GC Assist guided troubleshooting workflow for column performance
  6. p. 14
    Summary of column performance for residual solvents analysis
  7. p. 3 table, 25 rows
  8. p. 4 table, 7 rows
  9. p. 8 table, 35 rows

Table 1. Gas chromatography conditions.

ParameterValue
Agilent 8697 Headspace Sampler
Vial Pressurization GasN2
Loop Size1 mL
Vial Size20 mL
Vial Shaking8 (188 shakes/min)
Vial Equilibration Time40 min
Inject Time0.5 min
Oven Temperature85 °C
Loop Temperature85 °C
Transfer Line0.53 mm id, deactivated fused silica (p/n 160-2535-5)
Transfer Line Temperature100 °C
Vial Fill Pressure15 psi
Loop Fill ModeCustom
Loop Ramp Rate40 psi/min
Loop Final Pressure4 psi
Agilent 8890B GC
InletSSL, 140 °C, split ratio 10:1
LinerUltra inert, straight, 1 mm id (p/n 5190-4047)
CFT DeviceUnpurged two-way splitter (p/n G3181B)
ColumnColumn 1: Agilent DB-Select 624 UI, 30 m × 0.32 mm, 1.8 μm (p/n 123-0334UI) Column 2: Agilent DB-WAX UI, 30 m × 0.32 mm, 0.25 μm (p/n 123-7032UI)
CarrierHe and N2 carrier gas 2 mL/min, constant flow
Tubing0.32 mm id, deactivated fused silica tubing (p/n 160-2325-5)
OvenHelium carrier gas: 40 °C (5.5 min), then 15 °C/min to 180 °C, hold 2.5 min Nitrogen carrier gas: 38 °C (5.5 min), then 15 °C/min to 180 °C, hold 4 min
FID (Both Channels)Temperature: 250 °C, hydrogen: 30 mL/min, air: 300 mL/min,
Data Rate10 Hz

Table 2. USP <467> system suitability requirements for residual solvent analysis.

ProcedureSolvent ClassCompound NameRequirement by USP <467>
AClass 1S/N (1,1,1 trichloroethane)≥ 5
Class 1S/N (other solvents)≥ 3
Class 2AResolution (acetonitrile/ methylene chloride)≥ 1
BClass 1S/N (benzene)≥ 5
Class 1S/N (other solvents)≥ 3
Class 2AResolution (MIBK/ cis-1,2-dichloroethene)≥ 1
Class 2BResolution (1,2-dimethoxyethane and cyclopentyl methyl ether)≥ 1.5

Table 3. RT and area repeatability (n = 9) obtained on Agilent DB-Select 624 UI and Agilent DB-WAX UI columns using helium and nitrogen carrier gas.

Compound NameHe Carrier Gas — Procedure A DB-Select 624 UI — RT %RSDHe Carrier Gas — Procedure A DB-Select 624 UI — Area %RSDHe Carrier Gas — Procedure B DB-WAX UI — RT %RSDHe Carrier Gas — Procedure B DB-WAX UI — Area %RSDN2 Carrier Gas — Procedure A DB-Select 624 UI — RT %RSDN2 Carrier Gas — Procedure A DB-Select 624 UI — Area %RSDN2 Carrier Gas — Procedure B DB-WAX UI — RT %RSDN2 Carrier Gas — Procedure B DB-WAX UI — Area %RSD
Class 1
1,1-Dichloroethene0.0211.2750.0161.9310.0342.1860.0341.714
1,1,1-Trichloroethane0.0051.3730.0081.4090.0052.2010.0121.799
Carbon tetrachloride0.0161.955Coelutes with 1,1,1-trichloroethaneCoelutes with 1,1,1-trichloroethane0.032.825Coelutes with 1,1,1-trichloroethaneCoelutes with 1,1,1-trichloroethane
Benzene0.0041.8330.0131.360.0061.8280.012.002
1,2-Dichloroethane0.0032.1070.011.7680.0072.8070.0092.265
Class 2A
Methanol0.0472.5210.042.5580.0142.6290.0212.647
Acetonitrile0.0141.4060.0171.380.0031.7650.0041.797
Methylene chloride0.0070.7770.0120.7650.0030.3240.0040.37
trans-1,2-Dichloroethene0.010.845Coelutes with tetrahydrofuranCoelutes with tetrahydrofuran0.0050.53Coelutes with tetrahydrofuranCoelutes with tetrahydrofuran
cis-1,2-Dichloroethene0.0040.6960.0140.7190.0040.3990.0030.389
Tetrahydrofuran0.0090.9370.0090.8370.0040.4850.0050.504
Cyclohexane0.0060.8820.0090.8790.0070.3980.0040.399
Methylcyclohexane0.0031.890.011.8680.0020.4030.0030.418
1,4-Dioxane0.0062.3930.0092.0020.0032.4830.0072.77
Methyl isobutyl ketone0.0020.7040.0110.7120.0030.3380.0050.332
Toluene0.0030.6410.010.6360.0030.3980.0050.393
Chlorobenzene0.0010.6550.0730.6440.0020.4060.020.411
Ethylbenzene0.0020.6430.0070.6350.0020.4080.0030.404
m-Xylene0.0020.6450.0050.6320.0030.4160.0020.409
p-XyleneCoelutes with m-xyleneCoelutes with m-xylene0.0050.692Coelutes with m-xyleneCoelutes with m-xylene0.0030.436
o-Xylene0.0020.5410.0880.5220.0020.3860.0240.38
Cumene0.0020.7880.10.7870.0030.4020.0280.42
Class 2B
Tertiary butyl alcohol0.0381.0250.0081.0330.0781.0640.0121.051
Hexane0.0072.590.0162.5750.0111.7790.0151.57
Nitromethane0.0263.0160.0122.5980.022.9420.0172.38
Chloroform0.0041.7280.0081.7220.0071.8330.0141.604
1,2-Dimethoxyethane0.0081.9120.0142.2560.0172.8020.0181.729
Trichloroethylene0.0032.1410.0082.0920.0032.0760.0162.084
Cyclopentyl methyl ether0.0030.580.0060.5810.0040.750.0120.746
Pyridine0.0041.1670.0481.7860.0073.2910.1042.264
2-Hexanone0.0020.610.0050.6430.0020.8790.0090.682
Tetralin0.0031.0250.0121.030.0031.2270.0241.194

Engineering Facts

14 facts · every claim linked to a page
Selection

A dual-column, dual-FID configuration on a single inlet lets screening and confirmation run simultaneously, cutting total analysis time versus sequential approaches.

Determines whether you need a CFT splitter and two detectors/columns instead of a single-column setup.

Source
“By implementing a dual‑column, dual-FID configuration on a single inlet, screening and confirmatory analyses can be performed simultaneously”
Selection

Both helium and nitrogen carrier gases met acceptance criteria, so nitrogen can be chosen where helium supply is constrained.

Affects carrier gas supply planning and cost without compromising compliance.

Source
“Both helium and nitrogen were evaluated as carrier gases, and the results met or exceeded the acceptance criteria defined in both USP <467> and ChP 0861.”
Installation

The method requires an unpurged two-way CFT splitter to split sample equally to the two columns.

You must install the correct CFT device and splitter part to achieve parallel dual-column detection.

Source
“Unpurged two-way splitter (p/n G3181B)”
Installation

A 1 mm id ultra inert straight liner is specified for the SSL inlet.

Wrong liner geometry/inertness can degrade peak shape for reactive solvents like pyridine.

Source
“Ultra inert, straight, 1 mm id (p/n 5190-4047)”
Operation

Oven programs differ by carrier gas: helium starts at 40 °C, nitrogen at 38 °C, with different final holds.

You must load the correct method per carrier gas or retention times and resolution will shift.

Source
“Nitrogen carrier gas: 38 °C (5.5 min), then 15 °C/min to 180 °C, hold 4 min”
Operation

Headspace vials are equilibrated 40 min at 85 °C with shaking at setting 8.

Sets the minimum cycle time per sample and the required oven/shaker capability of the sampler.

Source
“Vial Equilibration Time 40 min”
Performance

Some compounds give low FID response and tailing, making integration parameters critical.

Warns that integration settings must be validated, not left at defaults, for reliable quantitation.

Source
“acetonitrile and pyridine showed both low response and peak tailing on the DB-Select 624 UI column, making peak integration sensitive to baseline definition”
Cumplimiento

USP <467> mandates minimum S/N and resolution criteria that must be verified as system suitability.

Defines the pass/fail checks your method must include before sample analysis.

Source
“USP <467> specifies mandatory system suitability requirements for residual solvent analysis”
Cumplimiento

The 2025 USP revision added tertiary butyl alcohol and CPME to class 2B with new resolution requirements.

Existing methods must be re-validated to cover the expanded class 2B list.

Source
“tertiary butyl alcohol and cyclopentyl methyl ether (CPME) have been added to the list of class 2B solvents”
Performance

CPME and pyridine coelute risk on procedure A but are fully resolved on the DB-WAX UI column.

Justifies the second column choice for confirming the newly added class 2B solvents.

Source
“On the DB WAX UI column operated under procedure B, CPME eluted at 3.91 minutes, while pyridine eluted at 9.14 minutes”
Performance

Peak area repeatability over nine injections averaged about 1.3 %RSD with either carrier gas.

Sets realistic precision expectations for method validation and system suitability limits.

Source
“the average percent relative standard deviation (%RSD) of peak areas across all analytes was 1.32%”
Operation

In a 32-compound mixed standard some pairs coelute on both columns, requiring MSD for confirmation.

Tells you when a mass spectrometer is needed instead of dual-FID alone.

Source
“several compounds, such as cis‑1,2‑dichloroethene, tetrahydrofuran, and CPME, coelute with other components on both columns”
Maintenance

GC Assist can run key performance checks and abort or continue the sequence on failure.

Enables automated system suitability gating and GMP-compliant documentation of failures.

Source
“If any parameter is found to be outside the allowable limits, the sequence can be configured to either continue or be aborted.”
Safety

Before column trimming or reinstallation the GC performs a controlled oven cool-down and then an automatic inlet leak test.

Supports safe column maintenance and leak verification without manual procedures.

Source
“the Agilent 8890B GC initiates a controlled oven cool-down sequence to ensure safe handling”