Residual Solvents Analysis Based on Agilent DB-Select 624 UI Columns
Contenido
Overviewp. 2
Background on residual solvents analysis and regulatory context
Specpp. 3-7
Chromatographic results for class 1, 2A, and 2B solvents
Qualityp. 8
Retention time and area repeatability on both column types
Procedurepp. 9-10
Single run analysis of combined residual solvent classes
Servicespp. 11-13
GC Assist guided troubleshooting workflow for column performance
p. 14
Summary of column performance for residual solvents analysis
p. 3table, 25 rows
p. 4table, 7 rows
p. 8table, 35 rows
Table 1. Gas chromatography conditions.
Parameter
Value
Agilent 8697 Headspace Sampler
Vial Pressurization Gas
N2
Loop Size
1 mL
Vial Size
20 mL
Vial Shaking
8 (188 shakes/min)
Vial Equilibration Time
40 min
Inject Time
0.5 min
Oven Temperature
85 °C
Loop Temperature
85 °C
Transfer Line
0.53 mm id, deactivated fused silica (p/n 160-2535-5)
Transfer Line Temperature
100 °C
Vial Fill Pressure
15 psi
Loop Fill Mode
Custom
Loop Ramp Rate
40 psi/min
Loop Final Pressure
4 psi
Agilent 8890B GC
Inlet
SSL, 140 °C, split ratio 10:1
Liner
Ultra inert, straight, 1 mm id (p/n 5190-4047)
CFT Device
Unpurged two-way splitter (p/n G3181B)
Column
Column 1: Agilent DB-Select 624 UI, 30 m × 0.32 mm, 1.8 μm (p/n 123-0334UI)
Column 2: Agilent DB-WAX UI, 30 m × 0.32 mm, 0.25 μm (p/n 123-7032UI)
Carrier
He and N2 carrier gas
2 mL/min, constant flow
Tubing
0.32 mm id, deactivated fused silica tubing (p/n 160-2325-5)
Oven
Helium carrier gas: 40 °C (5.5 min), then 15 °C/min to 180 °C, hold 2.5 min
Nitrogen carrier gas: 38 °C (5.5 min), then 15 °C/min to 180 °C, hold 4 min
FID (Both Channels)
Temperature: 250 °C, hydrogen: 30 mL/min, air: 300 mL/min,
Data Rate
10 Hz
Table 2. USP <467> system suitability requirements for residual solvent analysis.
Procedure
Solvent Class
Compound Name
Requirement by USP <467>
A
Class 1
S/N (1,1,1 trichloroethane)
≥ 5
Class 1
S/N (other solvents)
≥ 3
Class 2A
Resolution (acetonitrile/ methylene chloride)
≥ 1
B
Class 1
S/N (benzene)
≥ 5
Class 1
S/N (other solvents)
≥ 3
Class 2A
Resolution (MIBK/ cis-1,2-dichloroethene)
≥ 1
Class 2B
Resolution (1,2-dimethoxyethane and cyclopentyl methyl ether)
≥ 1.5
Table 3. RT and area repeatability (n = 9) obtained on Agilent DB-Select 624 UI and Agilent DB-WAX UI columns using helium and nitrogen carrier gas.
Compound Name
He Carrier Gas — Procedure A DB-Select 624 UI — RT %RSD
He Carrier Gas — Procedure A DB-Select 624 UI — Area %RSD
He Carrier Gas — Procedure B DB-WAX UI — RT %RSD
He Carrier Gas — Procedure B DB-WAX UI — Area %RSD
N2 Carrier Gas — Procedure A DB-Select 624 UI — RT %RSD
N2 Carrier Gas — Procedure A DB-Select 624 UI — Area %RSD
N2 Carrier Gas — Procedure B DB-WAX UI — RT %RSD
N2 Carrier Gas — Procedure B DB-WAX UI — Area %RSD
Class 1
1,1-Dichloroethene
0.021
1.275
0.016
1.931
0.034
2.186
0.034
1.714
1,1,1-Trichloroethane
0.005
1.373
0.008
1.409
0.005
2.201
0.012
1.799
Carbon tetrachloride
0.016
1.955
Coelutes with 1,1,1-trichloroethane
Coelutes with 1,1,1-trichloroethane
0.03
2.825
Coelutes with 1,1,1-trichloroethane
Coelutes with 1,1,1-trichloroethane
Benzene
0.004
1.833
0.013
1.36
0.006
1.828
0.01
2.002
1,2-Dichloroethane
0.003
2.107
0.01
1.768
0.007
2.807
0.009
2.265
Class 2A
Methanol
0.047
2.521
0.04
2.558
0.014
2.629
0.021
2.647
Acetonitrile
0.014
1.406
0.017
1.38
0.003
1.765
0.004
1.797
Methylene chloride
0.007
0.777
0.012
0.765
0.003
0.324
0.004
0.37
trans-1,2-Dichloroethene
0.01
0.845
Coelutes with tetrahydrofuran
Coelutes with tetrahydrofuran
0.005
0.53
Coelutes with tetrahydrofuran
Coelutes with tetrahydrofuran
cis-1,2-Dichloroethene
0.004
0.696
0.014
0.719
0.004
0.399
0.003
0.389
Tetrahydrofuran
0.009
0.937
0.009
0.837
0.004
0.485
0.005
0.504
Cyclohexane
0.006
0.882
0.009
0.879
0.007
0.398
0.004
0.399
Methylcyclohexane
0.003
1.89
0.01
1.868
0.002
0.403
0.003
0.418
1,4-Dioxane
0.006
2.393
0.009
2.002
0.003
2.483
0.007
2.77
Methyl isobutyl ketone
0.002
0.704
0.011
0.712
0.003
0.338
0.005
0.332
Toluene
0.003
0.641
0.01
0.636
0.003
0.398
0.005
0.393
Chlorobenzene
0.001
0.655
0.073
0.644
0.002
0.406
0.02
0.411
Ethylbenzene
0.002
0.643
0.007
0.635
0.002
0.408
0.003
0.404
m-Xylene
0.002
0.645
0.005
0.632
0.003
0.416
0.002
0.409
p-Xylene
Coelutes with m-xylene
Coelutes with m-xylene
0.005
0.692
Coelutes with m-xylene
Coelutes with m-xylene
0.003
0.436
o-Xylene
0.002
0.541
0.088
0.522
0.002
0.386
0.024
0.38
Cumene
0.002
0.788
0.1
0.787
0.003
0.402
0.028
0.42
Class 2B
Tertiary butyl alcohol
0.038
1.025
0.008
1.033
0.078
1.064
0.012
1.051
Hexane
0.007
2.59
0.016
2.575
0.011
1.779
0.015
1.57
Nitromethane
0.026
3.016
0.012
2.598
0.02
2.942
0.017
2.38
Chloroform
0.004
1.728
0.008
1.722
0.007
1.833
0.014
1.604
1,2-Dimethoxyethane
0.008
1.912
0.014
2.256
0.017
2.802
0.018
1.729
Trichloroethylene
0.003
2.141
0.008
2.092
0.003
2.076
0.016
2.084
Cyclopentyl methyl ether
0.003
0.58
0.006
0.581
0.004
0.75
0.012
0.746
Pyridine
0.004
1.167
0.048
1.786
0.007
3.291
0.104
2.264
2-Hexanone
0.002
0.61
0.005
0.643
0.002
0.879
0.009
0.682
Tetralin
0.003
1.025
0.012
1.03
0.003
1.227
0.024
1.194
Engineering Facts
14 facts · every claim linked to a page
Selection
A dual-column, dual-FID configuration on a single inlet lets screening and confirmation run simultaneously, cutting total analysis time versus sequential approaches.
Determines whether you need a CFT splitter and two detectors/columns instead of a single-column setup.
Source
“By implementing a dual‑column, dual-FID configuration on a single inlet, screening and confirmatory analyses can be performed simultaneously”
Selection
Both helium and nitrogen carrier gases met acceptance criteria, so nitrogen can be chosen where helium supply is constrained.
Affects carrier gas supply planning and cost without compromising compliance.
Source
“Both helium and nitrogen were evaluated as carrier gases, and the results met or exceeded the acceptance criteria defined in both USP <467> and ChP 0861.”
Installation
The method requires an unpurged two-way CFT splitter to split sample equally to the two columns.
You must install the correct CFT device and splitter part to achieve parallel dual-column detection.
Source
“Unpurged two-way splitter (p/n G3181B)”
Installation
A 1 mm id ultra inert straight liner is specified for the SSL inlet.
Wrong liner geometry/inertness can degrade peak shape for reactive solvents like pyridine.
Source
“Ultra inert, straight, 1 mm id (p/n 5190-4047)”
Operation
Oven programs differ by carrier gas: helium starts at 40 °C, nitrogen at 38 °C, with different final holds.
You must load the correct method per carrier gas or retention times and resolution will shift.
Source
“Nitrogen carrier gas: 38 °C (5.5 min), then 15 °C/min to 180 °C, hold 4 min”
Operation
Headspace vials are equilibrated 40 min at 85 °C with shaking at setting 8.
Sets the minimum cycle time per sample and the required oven/shaker capability of the sampler.
Source
“Vial Equilibration Time 40 min”
Performance
Some compounds give low FID response and tailing, making integration parameters critical.
Warns that integration settings must be validated, not left at defaults, for reliable quantitation.
Source
“acetonitrile and pyridine showed both low response and peak tailing on the DB-Select 624 UI column, making peak integration sensitive to baseline definition”
Cumplimiento
USP <467> mandates minimum S/N and resolution criteria that must be verified as system suitability.
Defines the pass/fail checks your method must include before sample analysis.
Source
“USP <467> specifies mandatory system suitability requirements for residual solvent analysis”
Cumplimiento
The 2025 USP revision added tertiary butyl alcohol and CPME to class 2B with new resolution requirements.
Existing methods must be re-validated to cover the expanded class 2B list.
Source
“tertiary butyl alcohol and cyclopentyl methyl ether (CPME) have been added to the list of class 2B solvents”
Performance
CPME and pyridine coelute risk on procedure A but are fully resolved on the DB-WAX UI column.
Justifies the second column choice for confirming the newly added class 2B solvents.
Source
“On the DB WAX UI column operated under procedure B, CPME eluted at 3.91 minutes, while pyridine eluted at 9.14 minutes”
Performance
Peak area repeatability over nine injections averaged about 1.3 %RSD with either carrier gas.
Sets realistic precision expectations for method validation and system suitability limits.
Source
“the average percent relative standard deviation (%RSD) of peak areas across all analytes was 1.32%”
Operation
In a 32-compound mixed standard some pairs coelute on both columns, requiring MSD for confirmation.
Tells you when a mass spectrometer is needed instead of dual-FID alone.
Source
“several compounds, such as cis‑1,2‑dichloroethene, tetrahydrofuran, and CPME, coelute with other components on both columns”
Maintenance
GC Assist can run key performance checks and abort or continue the sequence on failure.
Enables automated system suitability gating and GMP-compliant documentation of failures.
Source
“If any parameter is found to be outside the allowable limits, the sequence can be configured to either continue or be aborted.”
Safety
Before column trimming or reinstallation the GC performs a controlled oven cool-down and then an automatic inlet leak test.
Supports safe column maintenance and leak verification without manual procedures.
Source
“the Agilent 8890B GC initiates a controlled oven cool-down sequence to ensure safe handling”