Comparison of Agilent Femto Pulse

Contenido

  1. Overview p. 2
    Introduces the purpose and context of the comparison study
  2. Procedure p. 3
    Describes experimental methods, materials, and conditions used
  3. pp. 4-7
    Presents and discusses comparison results including smear size data
  4. p. 8
    Summarizes findings and key takeaways from the comparison
  5. p. 3 table, 3 rows

Table 1. Sequencing metrics and statistical analysis of the sequencing runs. N50: the length at which half of the nucleotides in an assembly belongs to reads equal or larger than that length. Number of reads above Q#: The number of reads that remain after removing those that fall below a failure threshold specified by the Q number (7, 10, 12, or 15), and is indicative of the accuracy of a sequencing run.

SampleMean read length (bp)Mean read qualityMedian read length (bp)Median read qualityNo. of readsRead length N50Total bases (in millions)No. of reads above Q7No. of reads above Q10No. of reads above Q12No. of reads above Q15
20 kb sheared9,578.911.08,58711.2115,22812,1521,103115,22791,41624,42014
15 kb size selected16,658.811.015,57811.258,41317,54797358,40947,90311,6825
40 kb size selected36,259.810.836,92811.08,23542,9962988,2356,1471,3180

MinION Sequencing Read Data Summary (After Trimming and Filtering)

Engineering Facts

12 facts · every claim linked to a page
Selection

Size selection should be chosen to remove fragments below a threshold so only long fragments load onto the sequencer, avoiding preferential sequencing of small fragments.

Determines whether a size-selection device and high-pass kit are needed in the workflow.

Source
“One strategy to improve long-read sequencing results is size selection of the sample, which can help eliminate smaller-sized fragments prior to sequencing”
Selection

High-pass filtering kits let the user collect all sample above a specified threshold, so the cutoff must be matched to the fragment length of interest.

Guides selection of the correct cassette/kit definition for the desired cutoff.

Source
“High-pass filtering kits for the Blue Pippin are beneficial for long-read sequencing, as they allow the user to collect all the sample above a specified threshold”
Operation

A 25-gauge needle shears 50 µL at 150 ng/µL to about 40 kb by aspirating 40 times.

Defines the manual shearing procedure and input concentration for reproducible fragment sizing.

Source
“A 25-gauge needle was utilized to shear a sample of 50 µL at 150 ng/µL to an average size of approximately 40 kb by aspirating 40 times.”
Operation

The g-TUBE shearing protocol uses 150 µL of gDNA at 100 ng/µL centrifuged at 7,600 rpm in 60-second intervals until the sample passes the filter.

Specifies the centrifugation speed, volume, and concentration needed for the 20 kb shear.

Source
“150 µL of gDNA at 100 ng/µL was placed into the g-TUBE and centrifuged with an Eppendorf MiniSpin at 7,600 rpm in 60 second intervals”
Selection

The 40 kb sheared samples used the 40 kb cutoff method with a 0.75% DF Marker U1 high-pass 30–40 kb v3 cassette.

Identifies the exact cassette part number and definition required for a 40 kb cutoff.

Source
“40 kb sheared samples were selected using the 40 kb cutoff method (Sage Science, p/n PAC-30KB; 0.75% DF Marker U1 high-pass 30–40 kb v3 cassette definition).”
Selection

The 20 kb sheared samples used the 15 kb cutoff method with a 0.75% DF marker S1 high-pass 15–20 kb cassette.

Identifies the exact cassette part number and definition required for a 15 kb cutoff.

Source
“20 kb sheared samples were size selected with the 15 kb cutoff method (Sage Science, p/n PAC-20KB; 0.75% DF marker S1 high-pass 15–20 kb cassette definition).”
Operation

Libraries were prepared with 500 ng DNA input using the Ligation Sequencing kit SQK-LSK109.

Sets the required DNA input amount for library preparation.

Source
“using 500 ng DNA input”
Operation

Two samples can share one flow cell if the manufacturer's flow cell washing protocol is followed between runs.

Enables reuse of a flow cell and requires a validated wash step between runs.

Source
“were sequenced on one flow cell, following the manufacturer’s protocol for washing the flow cell between runs”
Performance

The gDNA 165 kb kit provides fast automated sizing of HMW DNA through 165 kb with high accuracy.

Confirms the upper sizing range and accuracy relevant to selecting the QC method.

Source
“The optimized pulsed-field method of the gDNA 165 kb kit for the Femto Pulse is ideal for fast and automated analysis of HMW DNA through 165 kb, providing high sizing accuracy.”
Operation

Setting the cutoff too low removes little sample, while setting it too high removes almost all of it, so smear analysis is needed to pick the cutoff.

Prevents excessive sample loss or ineffective selection when configuring the cutoff.

Source
“setting the size selection too small, such as 6 kb, is unlikely to eliminate much of the sample, while setting it too big, such as 20 kb, would eliminate almost all of the sample”
Performance

The 40 kb cutoff is less sharp than the 15 kb cutoff, leaving a reproducible portion between 30 and 40 kb across replicates.

Sets realistic expectations for cutoff sharpness at larger fragment sizes.

Source
“indicating that the size-selection cutoff may not be as sharp at this larger size, compared to the 15 kb size-selected sample”
Performance

Small residual fragments are preferentially sequenced and lower the mean read length below the Femto Pulse average size.

Explains why measured mean read length can differ from the pre-library smear size.

Source
“The mean read length of the sheared gDNA is smaller than the Femto Pulse smear size due to the presence of smaller fragments in the sample, which are preferentially sequenced”