A New AdvanceBio HIC Column for Drug to Antibody Ratio (DAR) Analysis of Antibody Drug Conjugates (ADC)
Technical Reference Summary
Covered Models & Document Scope
Nota de aplicación para el análisis de DAR de conjugados anticuerpo-fármaco (ADC) con enlaces de cisteína mediante HIC utilizando el sistema de LC bio-inerte Agilent 1260 Infinity II (bomba G5654A, multisampler G5668A con opción de enfriador de muestras 100, termostato multicolumna G7116A con opción de intercambiador de calor bio-inerte 019, detector de matriz de diodos WR G7115A con opción de celda de flujo bio-inerte 028) con columnas AdvanceBio HIC de 4,6 × 100 mm (n/p 685975-908) y 4,6 × 30 mm (n/p 681975-908); abarca las variantes DAR 0, DAR 2, DAR 4, DAR 6 y DAR 8.
Technical Data
| Parameter | Value | Condición |
|---|---|---|
| Column | AdvanceBio HIC 4.6 × 100 mm mm p.2 ↗ | p/n 685975-908 |
| Mobile phase A | 50 mM Sodium phosphate, pH 7.0 p.2 ↗ | |
| Mobile phase B | 2 M Ammonium sulfate in 50 mM sodium phosphate, pH 7.0 p.2 ↗ | |
| Mobile phase C | Propan-2-ol p.2 ↗ | |
| Mobile phase D | HPLC grade water p.2 ↗ | |
| Flow rate | 0.5 mL/min p.2 ↗ | |
| Column temperature | 25 °C p.2 ↗ | |
| Injection volume | 5 µL p.2 ↗ | |
| Total run time | 31 minutes p.2 ↗ | |
| Gradient time 0 | 50 / 45 / 5 %A / %B / %C p.2 ↗ | typical gradient profile |
| Gradient time 15 | 75 / 0 / 25 %A / %B / %C p.2 ↗ | typical gradient profile |
| Gradient time 20 | 75 / 0 / 25 %A / %B / %C p.2 ↗ | typical gradient profile |
| Gradient time 21 | 50 / 45 / 5 %A / %B / %C p.2 ↗ | typical gradient profile |
| Gradient time 31 | 50 / 45 / 5 %A / %B / %C p.2 ↗ | typical gradient profile |
| Software | Agilent OpenLab 2.2 CDS p.2 ↗ |
Operating Limits
- column temperature: 25 °C p.2 ↗
- flow rate: 0.5 mL/min p.2 ↗
- injection volume: 5 µL p.2 ↗
- total run time: 31 minutes p.2 ↗
Ordering Data
| Ordering Code | Part No. | Variant / Code | Aplica a | Note |
|---|---|---|---|---|
| AdvanceBio HIC 4.6 × 100 mm p.2 ↗ | 685975-908 | HIC column | ||
| AdvanceBio HIC 4.6 × 30 mm p.2 ↗ | 681975-908 | HIC column |