Induztr
Resumen técnico Visor de documentos Modelos relacionados

Quantitative Screening of Multiresidue Veterinary Drugs in Milk and Egg Using the Agilent 6495C Triple Quadrupole LC/MS

Technical Reference Summary

Covered Models & Document Scope

Nota de aplicación que cubre la Agilent Comprehensive Veterinary Drug dMRM Solution para el cribado/cuantificación de 210 residuos de fármacos veterinarios en matrices de leche y huevo mediante el Agilent 6495C Triple Quadrupole LC/MS; verificada con dos espectrómetros de masas.

Technical Data

Parameter Value Condición
Number of target residues 210 p.1 ↗ milk and egg matrices
LOD ≤1 μg/kg p.1 ↗ milk, >93% of veterinary drugs
Calibration curve range LOQ to 100 μg/kg p.1 ↗ all targets
Coefficient of correlation (R2) ≥0.99 p.1 ↗ all targets
Target peak area response %RSD <15 % p.1 ↗
Retention time %RSD <0.5 % p.1 ↗
Method accuracy 87 to 117 % p.1 ↗ matrix-matched calibration
Average recovery 60 to 120 % p.1 ↗ 95% of targets
Repeatability %RSD ≤15 % p.1 ↗
Injection-to-injection robustness 400 injections p.1 ↗ peak area and RT reproducibility
Injection loop volume 20 µL p.3 ↗ LC system
Mobile phase A ammonium formate concentration 4.5 mM p.3 ↗
Mobile phase A ammonium fluoride concentration 0.5 mM p.3 ↗
Mobile phase A formic acid concentration 0.1 % p.3 ↗
Mobile phase B composition 50/50 ACN/MeOH p.3 ↗
Mobile phase B ammonium formate concentration 0.5 mM p.3 ↗
Dynamic MRM transition dwell time range 7 to 370 ms p.3 ↗
EDTA solution concentration for milk 1 M p.2 ↗ milk
EDTA volume added for milk 200 µL p.2 ↗ milk
Calibration standard levels 0.025, 0.05, 0.10, 0.25, 0.5, 1.0, 2.5 μg/kg p.3 ↗
Neat standard concentration 2.5 μg/L p.3 ↗ 50/50

Operating Limits

  • Sample storage temperature: –20 °C p.2 ↗
  • LOD for >93% of veterinary drugs in milk: ≤1 μg/kg p.1 ↗
  • Calibration curve range: LOQ to 100 μg/kg p.1 ↗
  • Coefficient of correlation (R2): ≥0.99 p.1 ↗
  • Target peak area response %RSD: <15% p.1 ↗
  • Retention time %RSD: <0.5% p.1 ↗