Quantitative Screening of Multiresidue Veterinary Drugs in Milk and Egg Using the Agilent 6495C Triple Quadrupole LC/MS
Technical Reference Summary
Covered Models & Document Scope
Nota de aplicación que cubre la Agilent Comprehensive Veterinary Drug dMRM Solution para el cribado/cuantificación de 210 residuos de fármacos veterinarios en matrices de leche y huevo mediante el Agilent 6495C Triple Quadrupole LC/MS; verificada con dos espectrómetros de masas.
Technical Data
| Parameter | Value | Condición |
|---|---|---|
| Number of target residues | 210 p.1 ↗ | milk and egg matrices |
| LOD | ≤1 μg/kg p.1 ↗ | milk, >93% of veterinary drugs |
| Calibration curve range | LOQ to 100 μg/kg p.1 ↗ | all targets |
| Coefficient of correlation (R2) | ≥0.99 p.1 ↗ | all targets |
| Target peak area response %RSD | <15 % p.1 ↗ | |
| Retention time %RSD | <0.5 % p.1 ↗ | |
| Method accuracy | 87 to 117 % p.1 ↗ | matrix-matched calibration |
| Average recovery | 60 to 120 % p.1 ↗ | 95% of targets |
| Repeatability %RSD | ≤15 % p.1 ↗ | |
| Injection-to-injection robustness | 400 injections p.1 ↗ | peak area and RT reproducibility |
| Injection loop volume | 20 µL p.3 ↗ | LC system |
| Mobile phase A ammonium formate concentration | 4.5 mM p.3 ↗ | |
| Mobile phase A ammonium fluoride concentration | 0.5 mM p.3 ↗ | |
| Mobile phase A formic acid concentration | 0.1 % p.3 ↗ | |
| Mobile phase B composition | 50/50 ACN/MeOH p.3 ↗ | |
| Mobile phase B ammonium formate concentration | 0.5 mM p.3 ↗ | |
| Dynamic MRM transition dwell time range | 7 to 370 ms p.3 ↗ | |
| EDTA solution concentration for milk | 1 M p.2 ↗ | milk |
| EDTA volume added for milk | 200 µL p.2 ↗ | milk |
| Calibration standard levels | 0.025, 0.05, 0.10, 0.25, 0.5, 1.0, 2.5 μg/kg p.3 ↗ | |
| Neat standard concentration | 2.5 μg/L p.3 ↗ | 50/50 |
Operating Limits
- Sample storage temperature: –20 °C p.2 ↗
- LOD for >93% of veterinary drugs in milk: ≤1 μg/kg p.1 ↗
- Calibration curve range: LOQ to 100 μg/kg p.1 ↗
- Coefficient of correlation (R2): ≥0.99 p.1 ↗
- Target peak area response %RSD: <15% p.1 ↗
- Retention time %RSD: <0.5% p.1 ↗