Induztr
Resumen técnico Visor de documentos Modelos relacionados

Data Sheet Model BD S 115

Technical Reference Summary

Covered Models & Document Scope

Modelo BD-S 115 (incubadora estándar Solid.Line con convección natural), volumen interior 118 L. Dos variantes: BDS115-230V (Art.-No. 9090-0022, 230 V 1~ 50/60 Hz, enchufe CEE 7/7) y BDS115UL-120V (Art.-No. 9090-0023, 120 V 1~ 60 Hz, enchufe NEMA 5-15).

Technical Data

Parameter Value Condición
Temperature range +5 °C above ambient temperature to 70 °C °C p.2 ↗ BDS115-230V / BDS115UL-120V
Temperature uniformity at 37°C 0.5 ± K p.2 ↗ BDS115-230V / BDS115UL-120V
Temperature fluctuation at 37°C 0.3 ± K p.2 ↗ BDS115-230V / BDS115UL-120V
Heating up time to 37°C 250 min p.2 ↗ BDS115-230V / BDS115UL-120V
Recovery time after door was opened for 30 s at 37°C 15 min p.2 ↗ BDS115-230V / BDS115UL-120V
Rated voltage 230 V p.2 ↗ BDS115-230V
Rated voltage 120 V p.2 ↗ BDS115UL-120V
Power frequency 50/60 Hz p.2 ↗ BDS115-230V
Power frequency 60 Hz p.2 ↗ BDS115UL-120V
Nominal power 0,35 kW p.2 ↗ BDS115-230V / BDS115UL-120V
Unit fuse 6,3 A p.2 ↗ BDS115-230V / BDS115UL-120V
Phase (nominal voltage) 1~ p.2 ↗ BDS115-230V / BDS115UL-120V
Interior volume 118 L p.2 ↗ BDS115-230V / BDS115UL-120V
Net weight of the unit (empty) 47 kg p.2 ↗ BDS115-230V / BDS115UL-120V
Load per rack 20 kg p.2 ↗ BDS115-230V / BDS115UL-120V
Permitted load 75 kg p.2 ↗ BDS115-230V / BDS115UL-120V
Wall clearance back 160 mm p.2 ↗ BDS115-230V / BDS115UL-120V
Wall clearance sidewise 100 mm p.2 ↗ BDS115-230V / BDS115UL-120V
Width net 765 mm p.2 ↗ BDS115-230V / BDS115UL-120V
Height net 735 mm p.2 ↗ BDS115-230V / BDS115UL-120V
Depth net 616 mm p.2 ↗ BDS115-230V / BDS115UL-120V
Interior width 550 mm p.2 ↗ BDS115-230V / BDS115UL-120V
Interior height 550 mm p.2 ↗ BDS115-230V / BDS115UL-120V
Interior depth 390 mm p.3 ↗ BDS115-230V / BDS115UL-120V
Energy consumption at 37°C 25 Wh/h p.3 ↗ BDS115-230V / BDS115UL-120V

Operating Limits

  • Temperature range: +5 °C above ambient temperature to 70 °C p.2 ↗
  • Wall clearance back: 160 mm p.2 ↗
  • Wall clearance sidewise: 100 mm p.2 ↗
  • Permitted load: 75 kg p.2 ↗
  • Load per rack: 20 kg p.2 ↗
  • Number of shelves (std./max.): 1/5 p.3 ↗

Type Code Decoder

Position Token Meaning
1-3 p.1 ↗ BDS Product line: Solid.Line standard incubator
4-6 p.1 ↗ 115 Model / size designation (interior volume 118 L)
suffix p.1 ↗ UL UL variant for 120 V / 60 Hz supply with NEMA 5-15 plug (absent = standard 230 V / 50/60 Hz with CEE 7/7 plug)
voltage field p.1 ↗ 230V Rated voltage 230 V, 1~, 50/60 Hz
voltage field p.1 ↗ 120V Rated voltage 120 V, 1~, 60 Hz

Ordering Data

Ordering Code Part No. Variant / Code Aplica a Note
BDS115-230V p.1 ↗ 9090-0022 230 V 1~ 50/60 Hz, CEE 7/7 plug, Standard Standard model version
BDS115UL-120V p.1 ↗ 9090-0023 120 V 1~ 60 Hz, NEMA 5-15 plug, Standard UL model version