Induztr
Resumen técnico Visor de documentos Modelos relacionados

Data Sheet Model FED 260

Technical Reference Summary

Covered Models & Document Scope

Cámara de secado y calefacción modelo FED 260 (FED Avantgarde.Line), volumen interior 259 L. Dos variantes estándar: FED260-230V (Art.-No. 9010-0299, 230 V 1~ 50/60 Hz, enchufe CEE 7/7) y FED260UL-240V (Art.-No. 9010-0300, 240 V 1~ 60 Hz, enchufe NEMA 6-20).

Technical Data

Parameter Value Condición
Temperature range +10 °C above ambient temperature to 300 °C p.2 ↗ FED260-230V / FED260UL-240V
Temperature uniformity at 150°C 1.6 ± K p.2 ↗ FED260-230V / FED260UL-240V
Temperature fluctuation at 150°C 0.4 ± K p.2 ↗ FED260-230V / FED260UL-240V
Heating up time to 150°C 19 min p.2 ↗ FED260-230V / FED260UL-240V
Recovery time after door was opened for 30 s at 150°C 5 min p.2 ↗ FED260-230V / FED260UL-240V
Air change rate at 100°C 9 x/h p.2 ↗ FED260-230V / FED260UL-240V
Rated voltage 230 V p.2 ↗ FED260-230V
Rated voltage 240 V p.2 ↗ FED260UL-240V
Power frequency 50/60 Hz p.2 ↗ FED260-230V
Power frequency 60 Hz p.2 ↗ FED260UL-240V
Nominal power 2,3 kW p.2 ↗ FED260-230V
Nominal power 2,5 kW p.2 ↗ FED260UL-240V
Unit fuse 12,5 A p.2 ↗ FED260-230V
Unit fuse 16,0 A p.2 ↗ FED260UL-240V
Phase (nominal voltage) 1~ p.2 ↗ FED260-230V / FED260UL-240V
Interior volume 259 L p.2 ↗ FED260-230V / FED260UL-240V
Net weight of the unit (empty) 85 kg p.2 ↗ FED260-230V / FED260UL-240V
Load per rack 40 kg p.2 ↗ FED260-230V / FED260UL-240V
Permitted load 270 kg p.2 ↗ FED260-230V / FED260UL-240V
Wall clearance back 160 mm p.2 ↗ FED260-230V / FED260UL-240V
Wall clearance sidewise 100 mm p.2 ↗ FED260-230V / FED260UL-240V
Width net 810 mm p.2 ↗ FED260-230V / FED260UL-240V
Height net 940 mm p.2 ↗ FED260-230V / FED260UL-240V
Depth net 760 mm p.2 ↗ FED260-230V / FED260UL-240V
Sound-pressure level 43 dB(A) p.3 ↗ FED260-230V / FED260UL-240V

Operating Limits

  • Temperature range: +10 °C above ambient temperature to 300 °C p.2 ↗
  • Permitted load: 270 kg p.2 ↗
  • Load per rack: 40 kg p.2 ↗
  • Wall clearance back: 160 mm p.2 ↗
  • Wall clearance sidewise: 100 mm p.2 ↗

Type Code Decoder

Position Token Meaning
1-6 p.1 ↗ FED260 Model series FED (drying/heating chamber, Avantgarde.Line) and size 260
suffix p.1 ↗ -230V Mains variant 230 V 1~ 50/60 Hz, CEE 7/7 plug, unit fuse 12,5 A
suffix p.1 ↗ UL-240V UL variant 240 V 1~ 60 Hz, NEMA 6-20 plug, unit fuse 16,0 A

Ordering Data

Ordering Code Part No. Variant / Code Aplica a Note
FED260-230V p.1 ↗ 9010-0299 230 V 1~ 50/60 Hz, CEE 7/7, standard Standard model version
FED260UL-240V p.1 ↗ 9010-0300 240 V 1~ 60 Hz, NEMA 6-20, standard Standard model version

Modelos relacionados