Induztr
Resumen técnico Visor de documentos Modelos relacionados

Data Sheet Model M 56

Technical Reference Summary

Covered Models & Document Scope

Cámara de secado y calefacción Modelo M 56 con convección forzada y funciones de programa avanzadas; versión estándar M056-230V, Art.-No. 9010-0371, 230 V 1~ 50/60 Hz, enchufe CEE 7/7, volumen interior de 59 L.

Technical Data

Parameter Value Condición
Interior volume 59 L p.1 ↗ M 56, 50/60 Hz
Rated voltage 230 V p.2 ↗ M 56, 50/60 Hz
Power frequency 50/60 Hz p.2 ↗ M 56
Phase (nominal voltage) 1~ p.2 ↗ M 56
Nominal power 2 kW p.2 ↗ M 56
Unit fuse 10,0 A p.2 ↗ M 56
Temperature range +23 °C above ambient temperature to 300 °C p.2 ↗ M 56
Temperature uniformity at 150°C 1.4 ± K p.2 ↗ M 56, at 150 °C
Temperature fluctuation at 150°C 0.2 ± K p.2 ↗ M 56, at 150 °C
Heating up time to 150°C 9 min p.2 ↗ M 56
Recovery time after door was opened for 30 s at 150°C 4 min p.2 ↗ M 56, at 150 °C
Air change rate at 100°C 175 x/h p.2 ↗ M 56, at 100 °C
Net weight of the unit (empty) 43 kg p.2 ↗ M 56
Load per rack 30 kg p.2 ↗ M 56
Permitted load 70 kg p.2 ↗ M 56
Wall clearance back 160 mm p.2 ↗ M 56
Wall clearance sidewise 100 mm p.2 ↗ M 56

Operating Limits

  • Temperature range: +23 °C above ambient temperature to 300 °C p.2 ↗
  • Permitted load: 70 kg p.2 ↗
  • Load per rack: 30 kg p.2 ↗
  • Wall clearance back: 160 mm p.2 ↗
  • Wall clearance sidewise: 100 mm p.2 ↗

Type Code Decoder

Position Token Meaning
1-4 p.1 ↗ M056 Model series and size designation (M series, size 56)
field 2 p.1 ↗ 230V Rated voltage variant 230 V

Ordering Data

Ordering Code Part No. Variant / Code Aplica a Note
M056-230V p.1 ↗ 9010-0371 Standard, 230 V 1~ 50/60 Hz, CEE 7/7 plug Standard model version