Induztr
Resumen técnico Visor de documentos Modelos relacionados

Agilent IDP-4 Dry Scroll Vacuum Pump

Contenido

  1. Overview pp. 1-3
    Introduces quiet, oil-free, environmentally friendly pump benefits
  2. Overview pp. 4-5
    Describes hermetic design, quiet operation, reliability, oil-free construction
  3. Overview pp. 6-7
    Presents the new compact IDP-4 and smart features
  4. Config. p. 8
    Guidance on selecting the appropriate IDP dry scroll pump
  5. Ordering pp. 9-10
    Lists ordering codes, part numbers, and available configurations
  6. Services p. 11
    Highlights partnership, service, and support offerings
  7. p. 10 table, 33 rows
  8. p. 9 table, 38 rows
  9. p. 8 table, 7 rows

Ordering information — Agilent IDP Dry Scroll Pumps

Model / DescriptionPart number
IDP-10X3810-64000
IDP-10 with inlet isolation valveX3810-64010
Service part numbers
IDP-7/10 tip seal replacement kitX3807-67000
Accesory part numbers
Exhaust silencer kitX3807-68003
Purge kitX3807-68004
Gas ballast kitX3807-68008
Inlet trapSCRINTRPNW25
Vibration isolation kitSH110VIBISOKIT
IDP-35X3835-64011
IDP-35 with GaugeX3835-64010
IDP-35 with Inlet ValveX3835-64001
IDP-35 with Inlet Valve and GaugeX3835-64000
Accesory part numbers
Inlet HEPA filter, NW40SCRINTRPNW40
HEPA filter, NW25SCRINTRPNW25
IDP-45X3845-64011
IDP-45 with GaugeX3845-64010
IDP-45 with Inlet ValveX3845-64001
IDP-45 with Inlet Valve and GaugeX3845-64000
Accesory part numbers
Inlet HEPA filter, NW40SCRINTRPNW40
HEPA filter, NW25SCRINTRPNW25
IDP-15X3815-64000
IDP-15 with inlet isolation valveX3815-64010
Service part numbers
IDP-15 tip seal replacement kitX3815-67000
Accesory part numbers
Exhaust silencer kitEXSLRSH110
Purge kitX3807-68004
Gas ballast kitX3807-68008
Inlet trapSCRINTRPNW25

Таблица на странице напечатана как несколько последовательных блоков (IDP-10, IDP-35, IDP-45, IDP-15) с одинаковой структурой «Description / Part number»; здесь они объединены в одну таблицу с сохранением порядка строк.

Ordering information — Agilent IDP Dry Scroll Pumps

ModelDescriptionPart number
IDP-3IDP-3, 1φ, 220 V, 50/60 HzIDP3A01
IDP-3IDP-3, 1φ, 115 V, 60 HzIDP3B01
IDP-3IDP-3, 1φ, 100 V, 50/60 HzIDP3C01
IDP-3IDP-3, 24 VDCIDP3D01
IDP-3With inlet isolation valve
IDP-3IDP-3, 1φ, 220 V, 50/60 HzIDP3A21
IDP-3IDP-3, 1φ, 115 V, 60 HzIDP3B21
IDP-3IDP-3, 1φ, 100 V, 50/60 HzIDP3C21
IDP-3IDP-3, 24 VDCIDP3D21
IDP-3Service part numbers
IDP-3Tip seal replacement kitIDP3TS
IDP-3Accessory part numbers
IDP-3Exhaust silencer kitEXSLRIDP3
IDP-3Inlet trapSCRINTRPNW16
IDP-3Breather vent and adapter kitIDP3GBKIT
IDP-3Replacement filter elementREPLSLRFILTER2
IDP-3Vibration isolation kitIDP3VIBISOKIT
IDP-4IDP-4 dry scroll pumpX3802-64000
IDP-4IDP-4 dry scroll pump, with VPIX3802-64010
IDP-4IDP-4 dry scroll pump, for helium useX3802-64400
IDP-4Service part numbers
IDP-4Tip seal replacement kitIDP3TS
IDP-4Accesory part numbers
IDP-4Exhaust silencer kitEXSLRIDP3
IDP-4Inlet valve kitVPI16IDP24DC
IDP-4Breather vent and adapter kitIDP3GBKIT
IDP-4Vibration isolation kitIDP3VIBISOKIT
IDP-4Replacement filter elementREPLSLRFILTER2
IDP-7IDP-7X3807-64000
IDP-7IDP-7 with inlet isolation valveX3807-64010
IDP-7Service part numbers
IDP-7IDP-7/10 tip seal replacement kitX3807-67000
IDP-7Accesory part numbers
IDP-7Exhaust silencer kitX3807-68003
IDP-7Purge kitX3807-68004
IDP-7Gas ballast kitX3807-68008
IDP-7Inlet trapSCRINTRPNW25
IDP-7Vibration isolation kitSH110VIBISOKIT

Модель (IDP-3, IDP-4, IDP-7) добавлена как отдельная колонка для группировки; в исходной таблице она указана заголовком раздела. Написание «Accesory» сохранено как в тексте.

How to select your Agilent IDP dry scroll vacuum pump — Agilent IDP Dry Scroll Pumps

IDP modelPumping SpeedBase PressureMotor Rating/operating voltageCommon ApplicationsAdvantages
IDP-360 Hz/24 VDC 60 L/min 3.6 m³/h 50 Hz 50 L/min 3.0 m³/h60 Hz/24 VDC 3.3 x 10-¹ mbar 2.5 x 10-¹ Torr 50 Hz 3.3 x 10-¹ mbar 2.5 x 10-¹ Torr0.16 HP (120 W)/ 100-120; 200-240 VAC- Leak detection - GC/MS - Helium recirculation- Small and light primary pump - 24 V drive
IDP-455 Hz 65 L/min 4 m3/h at full speed setting55 Hz 2.4 x 10-¹ mbar 1.8 x 10-¹ Torr at full speed setting0.16 HP (120 W)/100-127; 200-240 VAC- Leak detection - GC/MS - Helium recirculation- Lightest, most compact primary pump - Compact size for integrating the pump into your instrument - Universal pumping performance at all global frequencies and input voltages - Worldwide voltages compatibility
IDP-760 Hz 152 L/min 9.1 m³/h 50 Hz 120 L/min 7.2 m³/h60 Hz 2.6 x 10-² mbar 2.0 x 10-² Torr 50 Hz 4.0 x 10-² mbar 3.0 x 10-² Torr0.38 HP (300 W)/ 100-120; 200-240 VAC- Vacuum ovens - Helium recirculation - Backing high vacuum pump - General laboratory vacuum- More pumping speed in a compact, tabletop pump - Handles water vapor
IDP-1050 Hz or 60 Hz at full rotational speed (factory setting) 170 L/min 10.2 m³/h50 Hz or 60 Hz at full rotational speed (factory setting) 2.0 x 10-² mbar 1.5 x 10-² Torr0.5 HP (350 W)/ 100-127; 200-240 VAC- Microscopy instruments - Surface analysis instruments - Mass Spectrometry - Helium recirculation - Thin film deposition- Universal pumping performance at all input frequencies - Remote speed control or on/off capability
IDP-1560 Hz 256 L/min 15.4 m³/h 50 Hz 214 L/min 12.8 m³/h60 Hz 1.3 x 10-² mbar 1.0 x 10-² Torr 50 Hz 1.3 x 10-² mbar 1.0 x 10-² Torr0.75 HP (560 W)/ 100-115; 220-230 VAC- General laboratory vacuum - Thin film deposition - LED, bulb manufacturing With inlet isolation valve - GC/MS - Beam and particle physics - Backing a turbo pump- Quietest pump available - Lowest vibration at inlet available - Good speed to evacuate a large chamber - Safegaurds your system and high vacuum pump in case of power loss - Isolates the pump inlet during turbulent stop and start
IDP-3560 Hz 583 L/min 35 m³/h60 Hz 1.07 × 10-² mbar 8 × 10-³ Torr1.5 HP (1.1 KW)/ 100-127; 200-240 VAC- Helium recirculation - Leak detection - LC/MS - Vacuum ovens - Dry roughing for backing turbo pumps - Beam and particle physics- Hermetically sealed (magnetic torque coupling) - Precise vacuum control - Smart features- visibility and control - Communications/Remote access and control (analog and digital) - Inverter for universal voltage, flexible and easy
IDP-4560 Hz 750 L/min 45 m³/h60 Hz 1.07 × 10-² mbar 8 × 10-³ Torr1.5 HP (1.1 KW)/ 100-127; 200-240 VAC

Значения приведены дословно, как напечатано на странице; для IDP-45 колонки Common Applications и Advantages на странице не заполнены.

Engineering Facts

14 facts · every claim linked to a page
Maintenance

Tip seal or pumping module replacement is a DIY task needed only every two to three years and takes about half an hour, so plan for minimal downtime and no service contract.

Determines maintenance schedule, spare parts stocking, and total cost of ownership when selecting a pump.

Source
“This can be performed approximately every two to three years and requires about half an hour.”
Installation

The pump uses standard IEC power cords that are not hard-wired, simplifying electrical integration and replacement.

Affects cabinet wiring, connector selection, and serviceability during installation.

Source
“Agilent IDP dry scroll pumps also use standard IEC power cords that are not hard-wired to the motor.”
Safety

The optional integral isolation valve closes the inlet in about 20 ms on power loss and reopens in 10–30 s after power restore, protecting the vacuum system from backstreaming and sudden venting.

Critical for backing turbo pumps and contamination-sensitive systems; dictates whether the valve option must be ordered.

Source
“Upon power loss, the solenoid valve vents a small chamber beneath the valve, which closes the inlet valve (approximately 20 milliseconds).”
Installation

Integral vibration isolation feet or wheels separate motor and scroll vibration from the mounting surface, reducing transmitted vibration.

Matters for sensitive instruments (microscopy, surface analysis) and for floor/cabinet mounting decisions.

Source
“Integral vibration isolation feet or wheels, depending on the model, reduce vibrations by separating motor and scroll vibrations from the mounting brackets and support surface.”
Selection

The IDP-4 is the smallest and lightest pump in the family and is recommended for helium applications with proven long-term He tightness.

Guides selection for space-constrained OEM instruments and helium recirculation where footprint and He loss matter.

Source
“The Agilent IDP-4 dry scroll pump delivers clean, oil-free vacuum in the most compact footprint of the IDP family”
Operation

The IDP-4 offers adjustable pumping speed via inverter drive, allowing speed reduction once base pressure is reached to lower noise, energy use, and wear.

Enables optimization of power consumption and noise in continuous lab operation.

Source
“Adjustable pumping speed to match application requirements”
Operation

IDP-35 and IDP-45 have a tool-free adjustable gas ballast knob with Hi, Low, or No Flow settings to handle vapor loads.

Necessary when pumping condensable vapors (ovens, wet processes) to avoid condensation inside the pump.

Source
“The IDP-35 and IDP-45 models come standard with a simple (no tools), user-adjustable gas ballast knob.”
Installation

IDP-35 and IDP-45 use a Variable Frequency Drive giving consistent performance from 105 VAC to 240 VAC at 50 or 60 Hz.

Simplifies global deployment and OEM qualification without voltage-specific variants.

Source
“Get consistent performance worldwide from 105 VAC to 240 VAC (50 Hz or 60 Hz).”
Operation

IDP-35 and IDP-45 can be fitted with an integrated Pirani gauge head covering 1x10-4 to 1000 mbar for system monitoring.

Affects whether a separate gauge is needed and how pressure monitoring is integrated.

Source
“The IDP-35 and IDP-45 pumps can be fitted with an integrated Pirani gauge head to measure pressures in the range of 1x10-4 to 1000 mbar”
Installation

IDP-35 and IDP-45 support RS-232, RS-485, or analog interfaces for PC or PLC integration and remote control.

Determines control architecture and cabling for automated vacuum systems.

Source
“The IDP-35 and IDP-45 dry scroll pumps can be used with RS-232 or RS-485 or with an analog interface.”
Selection

The IDP-15 is described as the quietest pump available with the lowest vibration at inlet, and is offered with an inlet isolation valve.

Key selection criterion for noise-sensitive labs and vibration-sensitive instruments.

Source
“Quietest pump available”
Selection

The IDP-7 is noted for handling water vapor, making it suitable for vacuum ovens and general lab vacuum.

Important when selecting a pump for wet or condensable-vapor processes.

Source
“Handles water vapor”
Performance

IDP-35 and IDP-45 are hermetically sealed with magnetic torque coupling, ensuring complete isolation of pumped gases from the environment.

Required for rare gas and helium recirculation where leakage or contamination is unacceptable.

Source
“Hermetically sealed (magnetic torque coupling)”
Cumplimiento

The IDP-4 undergoes rigorous helium permeation testing aligned with worldwide standards.

Provides confidence for helium-critical applications and regulatory documentation.

Source
“Rigorous He permeation testing aligned with worldwide standards”