Citómetro de flujo Agilent NovoCyte. Sistemas estándar: 1000, 2000R, 2060R, 3000, 3005. Sistemas con láser amarillo: 2100YB, 3000VYB, 3000RYB. Sistemas con láser configurable con opciones de láser de 405 nm, 488 nm, 561 nm, 640 nm.
Technical Data
| Parameter | Value | Condición |
|---|---|---|
| Laser configuration | Spatially separated beams with 10 × 80 μm elliptical spots p.3 ↗ | |
| Optical alignment procedure | Fixed, no operator alignment required p.3 ↗ | |
| Flow cell | 170 × 290 μm rectangular quartz flow cell p.3 ↗ | |
| Scatter resolution | 0.2 μm p.3 ↗ | |
| Cell size | 0.2 to 50 μm p.3 ↗ | |
| Fluorescence threshold sensitivity | FITC <75 MESF, PE <50 MESF, APC <20 MESF p.3 ↗ | |
| Fluorescence resolution | <3% CV for CEN p.3 ↗ | |
| Filters | User exchangeable p.3 ↗ | |
| Sample acquisition rate | 35,000 events/second p.3 ↗ | |
| Volumetric absolute count precision | Syringe pump: CV <5% p.3 ↗ | |
| Sample flow rate | 5 to 120 μL/min p.3 ↗ | |
| Sheath flow rate | 6.5 mL/min p.3 ↗ | |
| Sample aspiration volume | 10 μL to 5 mL p.3 ↗ | |
| Fluid container capacity | 3 L sheath, 3 L waste, 500 mL cleaning, 500 mL decontamination p.3 ↗ | |
| Carryover | <0.1 % p.3 ↗ | |
| Fluidics maintenance | Automated startup, cleaning, decontamination, and shutdown p.3 ↗ | |
| Parameters | Height and area for FSC, SSC and all fluorescence channels, width, and time p.3 ↗ | |
| Dynamic range | 24 bit, 7.2 decades logarithmic scale, no need for PMT voltage adjustment p.3 ↗ | |
| Compensation | Automatic compensation, manual compensation, visual compensation tools available for pre/post/live acquisitions p.3 ↗ | |
| Output data files | FCS 3 p.3 ↗ | |
| Instrument weight | 86 lb (39 kg) p.3 ↗ | |
| Power requirements | 100/115/230 VAC, 50 to 60 Hz p.3 ↗ |
Operating Limits
- Cell size range: 0.2 to 50 μm p.3 ↗
- Sample flow rate range: 5 to 120 μL/min p.3 ↗
- Sample aspiration volume range: 10 μL to 5 mL p.3 ↗
- Power requirements: 100/115/230 VAC, 50 to 60 Hz p.3 ↗