Liquid Handling Brochure Varispenser 2 2x Top Performance
Technical Reference Summary
Covered Models & Document Scope
Dispensadores de botella Eppendorf Varispenser 2 y Varispenser 2x, rangos de volumen 0,2–2 mL, 0,5–5 mL, 1–10 mL, 2,5–25 mL, 5–50 mL, 10–100 mL; Varispenser 2 n.os de pedido 4966 000 010 / 029 / 037 / 045 / 053 / 061; Varispenser 2x n.os de pedido 4967 000 014 / 022 / 030 / 049 / 057 / 065; rosca de botella GL 45 con adaptadores para GL 25, GL 28, GL 32, GL 38, S²40
Technical Data
| Parameter | Value | Condición |
|---|---|---|
| Volume range | 0.2–2 mL p.6 ↗ | Varispenser 2, order no. 4966 000 010 |
| Volume step | 0.05 mL p.6 ↗ | Varispenser 2, 0.2–2 mL |
| Systematic error (relative) | ± 5 % p.6 ↗ | Varispenser 2, 0.2 mL |
| Systematic error (absolute) | ± 10 µL p.6 ↗ | Varispenser 2, 0.2 mL |
| Random error (relative) | ± 1 % p.6 ↗ | Varispenser 2, 0.2 mL |
| Random error (absolute) | ± 2 µL p.6 ↗ | Varispenser 2, 0.2 mL |
| Volume range | 0.5–5 mL p.6 ↗ | Varispenser 2, order no. 4966 000 029 |
| Volume step | 0.1 mL p.6 ↗ | Varispenser 2, 0.5–5 mL |
| Volume range | 1–10 mL p.6 ↗ | Varispenser 2, order no. 4966 000 037 |
| Volume step | 0.2 mL p.6 ↗ | Varispenser 2, 1–10 mL |
| Volume range | 2.5–25 mL p.6 ↗ | Varispenser 2, order no. 4966 000 045 |
| Volume step | 0.5 mL p.6 ↗ | Varispenser 2, 2.5–25 mL |
| Volume range | 5–50 mL p.6 ↗ | Varispenser 2, order no. 4966 000 053 |
| Volume step | 1 mL p.6 ↗ | Varispenser 2, 5–50 mL |
| Volume range | 10–100 mL p.6 ↗ | Varispenser 2, order no. 4966 000 061 |
| Volume step | 1 mL p.6 ↗ | Varispenser 2, 10–100 mL |
| Volume range | 0.2–2 mL p.6 ↗ | Varispenser 2x, order no. 4967 000 014 |
| Volume range | 0.5–5 mL p.6 ↗ | Varispenser 2x, order no. 4967 000 022 |
| Volume range | 1–10 mL p.6 ↗ | Varispenser 2x, order no. 4967 000 030 |
| Volume range | 2.5–25 mL p.6 ↗ | Varispenser 2x, order no. 4967 000 049 |
| Volume range | 5–50 mL p.6 ↗ | Varispenser 2x, order no. 4967 000 057 |
| Volume range | 10–100 mL p.6 ↗ | Varispenser 2x, order no. 4967 000 065 |
| Bottle thread | GL 45 p.6 ↗ | all sizes |
| Thread adapter included | GL 25, GL 28 p.6 ↗ | 0.2–2 mL and 0.5–5 mL |
| Thread adapter included | GL 32, GL 38, S²40 p.6 ↗ | 2.5–25 mL and 5–50 mL |
Operating Limits
- Maximum volume: 100 mL p.2 ↗
- Maximum bottle height with telescopic filling tube: 35 cm p.3 ↗
- Autoclavable: yes (completely autoclavable) p.2 ↗
Type Code Decoder
| Position | Token | Meaning |
|---|---|---|
| order no. suffix p.6 ↗ | 000 010 | 0.2–2 mL volume range |
| order no. suffix p.6 ↗ | 000 029 | 0.5–5 mL volume range |
| order no. suffix p.6 ↗ | 000 037 | 1–10 mL volume range |
| order no. suffix p.6 ↗ | 000 045 | 2.5–25 mL volume range |
| order no. suffix p.6 ↗ | 000 053 | 5–50 mL volume range |
| order no. suffix p.6 ↗ | 000 061 | 10–100 mL volume range |
| order no. suffix p.6 ↗ | 000 014 | 0.2–2 mL volume range (2x) |
| order no. suffix p.6 ↗ | 000 022 | 0.5–5 mL volume range (2x) |
| order no. suffix p.6 ↗ | 000 030 | 1–10 mL volume range (2x) |
| order no. suffix p.6 ↗ | 000 049 | 2.5–25 mL volume range (2x) |
Ordering Data
| Ordering Code | Part No. | Variant / Code | Aplica a | Note |
|---|---|---|---|---|
| Varispenser 2, 0.2–2 mL p.6 ↗ | 4966 000 010 | GL 45; adapters GL 25, GL 28 | ||
| Varispenser 2, 0.5–5 mL p.6 ↗ | 4966 000 029 | GL 45; adapters GL 25, GL 28 | ||
| Varispenser 2, 1–10 mL p.6 ↗ | 4966 000 037 | GL 45; adapters GL 25, GL 28 | ||
| Varispenser 2, 2.5–25 mL p.6 ↗ | 4966 000 045 | GL 45; adapters GL 32, GL 38 | ||
| Varispenser 2, 5–50 mL p.6 ↗ | 4966 000 053 | GL 45; adapters GL 32, GL 38, S²40 | ||
| Varispenser 2, 10–100 mL p.6 ↗ | 4966 000 061 | GL 45; adapters GL 32, GL 38 | ||
| Varispenser 2x, 0.2–2 mL p.6 ↗ | 4967 000 014 | GL 45; adapters GL 25, GL 28 | ||
| Varispenser 2x, 0.5–5 mL p.6 ↗ | 4967 000 022 | GL 45; adapters GL 25, GL 28 | ||
| Varispenser 2x, 1–10 mL p.6 ↗ | 4967 000 030 | GL 45; adapters GL 25, GL 28 | ||
| Varispenser 2x, 2.5–25 mL p.6 ↗ | 4967 000 049 | GL 45; adapters GL 32, GL 38 |