Induztr
Resumen técnico Visor de documentos Modelos relacionados

FRG-720 Vacuum Gauge User Manual

Contenido

  1. Overview pp. 2-3
    Legal notices and table of contents
  2. Overview pp. 4-5
    Instructions for use and basic operating concept
  3. Overview pp. 6-7
    Manual conventions and terminology definitions
  4. Warning p. 8
    Safety warning symbols used in the manual
  5. Warning pp. 9-13
    Safety, improper use, disposal and service guidance
  6. Overview pp. 14-16
    Gauge construction and functional description
  7. Spec pp. 17-20
    Measurement ranges, accuracy and operating conditions
  8. Spec pp. 21-22
    Dimensional drawing and pressure definitions
  9. Spec pp. 23-25
    Correction factors for gases and notes
  10. Procedure pp. 26-33
    Mounting, electronics unit removal and optional accessories
  11. Procedure pp. 34-35
    Wiring and use with other controllers
  12. Procedure pp. 36-42
    Serial communication setup and commands
  13. Procedure pp. 43-47
    Startup procedure and display description
  14. Spec pp. 48-52
    Interface parameters and command reference
  15. Spec pp. 53-55
    Profibus configuration and data exchange
  16. Procedure pp. 56-58
    Setting and configuring switching functions
  17. Procedure pp. 59-60
    Removing the gauge from the system
  18. Procedure pp. 61-63
    Maintenance, repair and safety warnings
  19. Procedure p. 64
    Procedure for zero point calibration
  20. Procedure pp. 65-66
    Fault diagnosis and remedies
  21. Procedure p. 67
    Sensor replacement instructions
  22. Ordering pp. 68-71
    Ordering information for accessories and spare parts
  23. pp. 72-73
    Customer care contacts and book index
  24. p. 15 table, 9 rows
  25. p. 68 table, 5 rows
  26. p. 14 table, 36 rows
  27. p. 18 table, 31 rows
  28. p. 20 table, 13 rows
  29. p. 41 table, 14 rows
  30. p. 47 table, 11 rows
  31. p. 22 table, 3 rows

Table 1 Part Numbers

Part NumberDescription
Without Display
FRG720KF25FRG-720 Pirani/Bayard-Alpert combination full range gauge, single filament, NW 25KF flange.
FRG720CF35FRG-720 Pirani/Bayard-Alpert combination full range gauge, single filament, DN40 CFF 2.75 flange.
With Display
FRG720KF25DFRG-720 Pirani/Bayard-Alpert combination full range gauge, single filament, LCD display, NW 25KF flange.
FRG720CF35DFRG-720 Pirani/Bayard-Alpert combination full range gauge, single filament, LCD display, DN 40 CFF 2.75 flange.
With Profibus interface and 2 switching functions
FRG720KF25SPFRG-720 Pirani/Bayard-Alpert combination full range gauge, single filament, with setpoints and profibus, NW 25KF flange.
FRG720CF35SPFRG-720 Pirani/Bayard-Alpert combination full range gauge, single filament, DN40 CFF 2.75 flange, Profibus.

Table 14 FRG-720 Accessories and Spare Parts

PART NUMBERDESCRIPTION
FRG720BKEBakeout extension, FRG-720 Pirani/Bayard-Alpert combination full range gauge, 100 mm.
FRG720KF25RSReplacement sensor, FRG-720 Pirani/Bayard-Alpert combination full range gauge, NW 25KF.
FRG720CF35RSReplacement sensor, FRG-720 Pirani/Bayard-Alpert combination full range gauge, 2.75 in ConFlat Flange.
FRG720BFLBaffle, FRG-720 Pirani/Bayard-Alpert combination full range gauge.
FRG720PWRPower supply, FRG-720 full range gauge, 24 V DC, RS232C.

Содержание

РазделСтраница
Technical Information
Product Description15
Technical Specifications17
FRG-720 Outline21
Measurement Signal vs. Pressure22
Gas Type Dependence23
Installation26
Vacuum Connection26
Removing and Installing the Electronics Unit28
Installing the Optional Extension29
Using the Optional Baffle31
Electrical Connection34
Use With Agilent Vacuum Gauge Controller34
Use With Other Controllers35
Operation43
Measuring Principle, Measuring Behavior43
Putting the Gauge Into Operation46
Degas46
Display47
RS232C Interface48
Description of the Functions48
Profibus Interface53
Description of the Functions53
Operating Parameters54
Switching functions56
Setting the Switching Functions57
Deinstallation59
Maintenance, Repair61
Cleaning the Gauge61
Adjusting the Gauge62
Adjustment at Atmospheric Pressure62
Zero Point Adjustment64
Troubleshooting65
What to Do in Case of Problems65
Replacing the Sensor67
Accessories and Spare Parts68

Table 2 Technical Specifications (continued)

FRG-720 SpecificationsValue
Power Supply
WARNING The gauge must only be connected to power supplies, instruments or control devices that conform to the requirements of a grounded extra-low voltage (SELV). The connection to the gauge has to be fused (Agilent controllers fulfill these requirements).
Operating voltage at the gauge+24 VDC (+20 … +28 VDC) 1) ripple ≤2 Vpp
Power consumption
Standard≤0.5 A
Degas≤0.8 A
Emission start (<200 ms)≤1.4 A
Power consumption
Standard gauge≤16 W
Profibus gauge≤18 W
Fuse necessary1.25 AT
Sensor cable connection
Electrical connectorD-Sub,15-pin, male
Standard gauge
Cable for standard gauge
Analog values only Without degas function4 conductors plus shielding
Analog values with degas function5 conductors plus shielding
Analog values with degas function and RS232C interface7 conductors plus shielding
Profibus gaugedepending on the functions used, max 15 conductors plus shielding
Cable for Profibus gauge
Max. cable length (supply voltage 24 V 1))
Analog and fieldbus operation≤35 m, conductor cross-section 0.25 mm² ≤50 m, conductor cross-section 0.34 mm² ≤100 m, conductor cross-section 1.0 mm²
RS232C operation≤30 m
Gauge identification42 kΩ resistor between Pin 10 (sensor cable) and GND
Switching functions
Standard gaugenone
Profibus gauge2 (setpoints A and B)
Adjustment range1×10-9 mbar … 100 mbar
Relay contact ratingSetpoints adjustable via potentiometers (setpoints A and B), one floating, normally open relay contact per setpoint. Adjusting the setpoints via field bus is described in the corresponding bus sections.
Voltage≤60 V
Current≤0.5 ADC

1) Measured at sensor cable connector (consider the voltage drop as function of the sensor cable length).

Table 2 Technical Specifications (continued)

FRG-720 SpecificationsValue
Weight
FRG720KF25, FRG720KF25D≈285 g
FRG720CF35, FRG720CF35D≈550 g
FRG720KF25SP≈430 g
FRG720CF35SP≈695 g
Ambiance
Admissible temperatures
Storage–20 … +70 °C
Operation0 … +50 °C
Bakeout+150 °C (without electronics unit or with bakeout extension
Relative humidity (year's mean / during 60 days)≤65 / 85% (no condensation)
Useindoors only altitude up to 2000 m NN
Type of protectionIP 30

CAUTION Vacuum component. Inappropriate storage leads to an increase of the desorption rate and/or may result in mechanical damage of the product. Cover the vacuum ports of the product with protective lids or grease free aluminum foil. Do not exceed the admissible storage temperature range.

Table 3 Technical data

Value
Mains connection
Mains connection
Mains voltage90 … 250 VAC 50 … 60 Hz
Mains cable1.8 meter (Schuko DIN and U.S. connectors)
Output (operating voltage of gauge)
Voltage21 … 27 VDC, set to 24 VDC
CurrentMax. 1.5 A
Gauge connection
ConnectorD-Sub, 15-pin, female
24 V cable5 m, black
Connection of the instrument or control device
RS232C connection
CableD-Sub, 9-pin, female
5 m, black, 3 conductors, shielded

Display

Display elementMeaning
Pressure reading, pressure unit
(non) Pirani operation
EEmission 25 μA
E.Emission 5 mA
DDegas
A1000 mbar adjustment (Pirani)
Okno error (green background illumination)
5Pirani sensor warning (red background illumination)
9Pirani sensor error (red background illumination)
8BA sensor error (red background illumination)
Internal data connection failure (red background illumination)

Function display: E, E., D, A, Ok, 5, 9, 8; Pressure display; Error display

Measurement Signal vs. Pressure

U [V]p [mbar]p [Pa]p [Torr]
Upc
[V][mbar][Pa][Torr]
-0.125

p = 10^((U-7.75)/0.75+c), где p — давление, U — измерительный сигнал, c — константа (зависит от единицы давления)

Engineering Facts

13 facts · every claim linked to a page
Selection

The gauge covers a continuous measurement range of 5×10⁻¹⁰ to 1000 mbar, so it can serve both high-vacuum and rough-vacuum needs with one sensor.

Defines whether a single gauge can replace separate Pirani and BA heads across the whole process range.

Source
“5×10-10 … 1000 mbar continuous”
Operation

The hot cathode is automatically switched on only below 2.4×10⁻² mbar and off above 3.2×10⁻² mbar, protecting the filament from burnout at high pressure.

Confirms the gauge is safe to leave connected during venting and that no external interlock is needed for filament protection.

Source
“The hot cathode is switched on by the Pirani measurement system only below the switching threshold of 2.4×10-2 mbar (to prevent filament burn-out).”
Installation

The gauge must be powered from a grounded SELV supply and the connection must be fused, otherwise the installation is non-compliant.

Determines the power supply and fusing requirements for the electrical cabinet design.

Source
“The gauge must only be connected to power supplies, instruments or control devices that conform to the requirements of a grounded extra- low voltage (SELV).”
Installation

For KF-flanged versions a conductive metallic clamping ring is required to achieve the protective earth connection to the chamber.

Affects the choice of clamp hardware; a non-conductive ring would leave the gauge ungrounded and hazardous.

Source
“For gauges with KF connections, use a conductive metallic clamping ring.”
Installation

The sensor can be baked to 150 °C, but above 50 °C the electronics unit must be removed or a bakeout extension installed.

Dictates whether a bakeout extension must be ordered and how the gauge is mounted on a bakeable chamber.

Source
“At temperatures exceeding 50 °C, the electronics unit has to be removed or an extension has to be installed.”
Installation

The gauge may be mounted in any orientation, but horizontal to upright is preferred to keep condensates and particles out of the measuring chamber.

Guides physical placement on the chamber to avoid contamination-related measurement drift.

Source
“The gauge may be mounted in any orientation. To keep condensates and particles from getting into the measuring chamber preferably choose a horizontal to upright position.”
Cumplimiento

The gauge must not be operated with an Agilent Vacuum Gauge Controller and Profibus at the same time because data transmission errors may occur.

Prevents a wiring/architecture mistake when integrating the gauge into a controller plus fieldbus network.

Source
“The gauge must not be operated with a Agilent Vacuum Gauge Controller and Profibus at the same time.”
Installation

Maximum RS232C cable length is 30 m, while analog/fieldbus operation allows up to 100 m with 1.0 mm² conductors.

Sets the cable gauge and routing limits between the gauge and the control cabinet.

Source
“≤100 m, conductor cross-section 1.0 mm²”
Operation

A stabilization time of about 10 minutes must be allowed after power-on before readings meet the specified accuracy.

Defines the warm-up period to respect before using the gauge for process control or calibration checks.

Source
“Allow for a stabilizing time of approx. 10 min.”
Maintenance

Degas heats the electron collector grid to about 700 °C and terminates automatically after 3 minutes; it should only be run below 7.2×10⁻⁶ mbar.

Specifies the safe pressure window and duration for in-situ cleaning to avoid damage and unstable readings.

Source
“The degas process should be run at pressures below 7.2×10-6 mbar”
Performance

For gases other than air/N2 the indicated pressure must be corrected with a calibration factor, e.g. 0.8 for Ar and 5.9 for He.

Required to interpret readings correctly when the process gas is not nitrogen or air.

Source
“peff = C × indicated pressure”
Safety

Clamps must not be opened while the vacuum system is pressurized above 1 bar, as released parts and escaping process gases can cause injury.

Defines the venting/interlock procedure for maintenance work on the gauge port.

Source
“Do not open any clamps while the vacuum system is pressurized.”
Maintenance

Gauge failures due to contamination or wear, and expendable parts such as the filament, are excluded from warranty.

Affects spare-parts planning and the decision to add a baffle in contaminating processes.

Source
“Gauge failures due to contamination or wear and tear as well as expendable parts (e.g. filament) are not covered by the warranty.”

Modelos relacionados