CC-104A Immersion Circulator

Contenido

  1. Overview p. 1
    Introduces the CC-104A immersion circulator and its main features
  2. Spec p. 2
    Lists technical data according to DIN 12876 standard
  3. p. 2 table, 3 rows
  4. p. 1 table, 35 rows

Technical data according to DIN 12876

ParameterValue
max. ambient temperature40 °C
from Serial-No.:398007
1.0/20

Technical data according to DIN 12876

ParameterValue
Operating temperature range25...100 °C
with water cooling20...100 °C
with refrigerator15...100 °C
Temperature stability at 70°C0,02 K
temperature set point / display5,7" colour Touchscreen
Absolute accuracysetup for calibration
Internal temperature sensorPt100
Sensor external connectionPt100
Interface digitalEthernet, USB (Host u. Device), RS232
Safety classificationIII / FL
Heating power at 240V2,1 kW
Heating power at 230V2 kW
Heating power at 220V1,8 kW
Heating power at 208V1,6 kW
Heating power at 200V1,5 kW
Pressure pumpyes
max. delivery27 l/min
max. delivery pressure0,7 bar
Suction pumpyes
max. delivery (suction)22 l/min
max. delivery pressure (suction)0,4 bar
Pump connectionM16x1 male
Bath volume3 l
Filling capacity2 l
Bath opening diameter25 mm
Bath depth150 mm
Overall dimensions WxDxH **147x235x330 mm
Net weight6 kg
Power supply requirement200-240V 1~/2~ 50/60Hz
max. current10 A
min. Fuse10A
max. Fuse16A
Pressure equipment categoryArt. 4.3 PED
Degree of ProtectionIP20
min. ambient temperature5 °C

** Overall dimensions WxDxH

Engineering Facts

13 facts · every claim linked to a page
Selection

The operating temperature range depends on the cooling method: 25–100 °C without cooling, 20–100 °C with water cooling, and 15–100 °C with a refrigerator.

Determines whether an external water supply or chiller is required to reach the desired setpoint.

Source
“Operating temperature range 25...100 °C with water cooling 20...100 °C with refrigerator 15...100 °C”
Installation

The unit requires a 200–240 V single- or two-phase 50/60 Hz supply with a minimum 10 A fuse and maximum 16 A fuse.

Electrical planning must provide a branch circuit rated between 10 A and 16 A at the correct voltage and phase configuration.

Source
“Power supply requirement 200-240V 1~/2~ 50/60Hz max. current 10 A min. Fuse 10A max. Fuse 16A”
Installation

External branch circuit protection according to UL 489 is mandatory.

The installer must provide a UL 489-listed breaker; otherwise the installation does not comply.

Source
“External branch circuit protection according to UL 489 required.”
Installation

Voltage and frequency tolerances are ±10 %, but a −10 % voltage combined with a +3 % frequency is not allowed.

Site power quality must be checked; some voltage/frequency combinations are outside the permitted operating envelope.

Source
“Voltage + / - 10%, as long as the frequency tolerance does not run in the opposite direction.”
Installation

Single- and two-phase devices are delivered with a power cable and country-specific plug, which must be specified at ordering.

The correct plug type must be stated on the order to avoid a non-connectable unit on site.

Source
“Single / two-phase devices (100V to 240V) --> with power cable and country-specific plug (please specify when ordering)”
Safety

The unit has safety classification III / FL according to DIN 12876 and adjustable overtemperature protection.

Confirms the safety class for flammable liquids and the presence of adjustable overtemperature cut-out for safe operation.

Source
“Safety classification III / FL”
Operation

The pressure pump delivers up to 27 l/min at 0.7 bar and the suction pump up to 22 l/min at 0.4 bar.

External circuit pressure drop must be checked against these pump curves to ensure adequate circulation.

Source
“max. delivery 27 l/min max. delivery pressure 0,7 bar”
Operation

The bath holds 3 l total but the filling capacity is 2 l, and the bath opening is only 25 mm in diameter with 150 mm depth.

Limits the size of immersed samples and the volume of external loop that can be connected.

Source
“Bath volume 3 l Filling capacity 2 l Bath opening diameter 25 mm Bath depth 150 mm”
Installation

The pump connection is M16x1 male, and the ambient temperature must stay between 5 °C and 40 °C.

Hose adapters must match M16x1, and the installation location must respect the 5–40 °C ambient window.

Source
“Pump connection M16x1 male”
Cumplimiento

The unit is classified as Class A, Group 1 for electromagnetic disturbance to EN55011.

Class A equipment is intended for industrial environments; residential use may require additional EMC measures.

Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Selection

The transparent polycarbonate bath limits the maximum temperature to 100 °C.

The bath material caps the usable upper temperature regardless of the controller range.

Source
“Heating Circulator consisting of transparent baths made of polycarbonate with temperature range up to max. 100°C.”
Operation

The pressure equipment category is Art. 4.3 PED and the protection degree is IP20.

IP20 means no water protection, so the unit must be kept dry; PED category affects pressure-system compliance.

Source
“Pressure equipment category Art. 4.3 PED Degree of Protection IP20”
Maintenance

A 4-year warranty is available only with registration.

Registration must be completed to secure the extended warranty coverage.

Source
“4-year warranty - registration required.”