Technical data according to DIN 12876, including temperature and performance figures
p. 1table, 35 rows
Technical data according to DIN 12876
Parameter
Value
Operating temperature range
28...200 °C
with water cooling
20...200 °C
with refrigerator
-20...200 °C
Temperature stability at 70°C
0,02 K
temperature set point / display
5,7" colour Touchscreen
Absolute accuracy
setup for calibration
Internal temperature sensor
Pt100
Sensor external connection
Pt100
Interface digital
Ethernet, USB (Host u. Device), RS232
Safety classification
III / FL
Heating power at 240V
2,1 kW
Heating power at 230V
2 kW
Heating power at 220V
1,8 kW
Heating power at 208V
1,6 kW
Heating power at 200V
1,5 kW
Pressure pump
yes
max. delivery
27 l/min
max. delivery pressure
0,7 bar
Suction pump
yes
max. delivery (suction)
22 l/min
max. delivery pressure (suction)
0,4 bar
Pump connenction (optional)
M16x1 male
Immersion depth
128 mm
max. permissible kin. viscosity
50 mm²/s
Overall dimensions WxDxH **
345x200x326 mm
Net weight
7 kg
Power supply requirement
200-240V 1~/2~ 50/60Hz
max. current
10 A
alternative max. current
10 A
min. Fuse
10A
max. Fuse
16A
Pressure equipment category
Art. 4.3 PED
Degree of Protection
IP20
min. ambient temperature
5 °C
max. ambient temperature
40 °C
** Overall dimensions WxDxH
Engineering Facts
14 facts · every claim linked to a page
Selection
The operating temperature range depends on the cooling method: 28–200 °C without cooling, 20–200 °C with water cooling, and -20–200 °C with a refrigerator.
Determines whether the unit can reach the required process temperature with the available cooling utility.
Source
“Operating temperature range 28...200 °C with water cooling 20...200 °C with refrigerator -20...200 °C”
Installation
The unit requires a 200–240 V single- or two-phase supply at 50/60 Hz with a maximum current of 10 A and a fuse rated between 10 A and 16 A.
Needed to verify site power availability and correct branch-circuit protection before installation.
Source
“Power supply requirement 200-240V 1~/2~ 50/60Hz max. current 10 A alternative max. current 10 A min. Fuse 10A max. Fuse 16A”
Installation
Voltage and frequency tolerances are ±10% only if they do not run in opposite directions, so a -10% voltage with +3% frequency is not allowed.
Prevents installation on a supply where combined voltage and frequency deviations exceed the permitted envelope.
Source
“Voltage + / - 10%, as long as the frequency tolerance does not run in the opposite direction.”
Safety
External branch circuit protection according to UL 489 is required.
Specifies the type of upstream protection that must be provided for compliant installation.
Source
“External branch circuit protection according to UL 489 required.”
Installation
Single- and two-phase devices are supplied with a power cable and country-specific plug, which must be specified at ordering.
Ensures the correct plug is ordered for the destination country and avoids rework at commissioning.
Source
“Single / two-phase devices (100V to 240V) --> with power cable and country-specific plug (please specify when ordering)”
Operation
The ambient temperature must stay between 5 °C and 40 °C, and the unit has IP20 protection, so it must be kept in a dry indoor location.
Defines the permissible site environment and rules out wet or dusty locations.
Source
“Degree of Protection IP20 min. ambient temperature 5 °C max. ambient temperature 40 °C”
Performance
The pressure pump delivers up to 27 l/min at 0.7 bar and the suction pump up to 22 l/min at 0.4 bar, which limits the connected external circuit.
Used to check that the external loop pressure drop and flow demand are within pump capability.