CC-308B Technical Data Sheet

Contenido

  1. Overview p. 1
    Product name and cover page for CC-308B
  2. Spec p. 2
    Technical data per DIN 12876 with specification table
  3. p. 2 table, 4 rows

Technical data according to DIN 12876

ParameterValue
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:S393466

1.2/20

Engineering Facts

14 facts · every claim linked to a page
Selection

The operating temperature range depends on the cooling method: 28–300 °C without cooling, 20–300 °C with water cooling, and -20–300 °C with a refrigerator.

Determines whether the bath can reach the required process temperatures with the available cooling utility.

Source
“Operating temperature range 28...300 °C with water cooling 20...300 °C with refrigerator -20...300 °C”
Performance

Heating power is voltage-dependent, dropping from 3 kW at 230–240 V to 2.2 kW at 200 V.

Affects heat-up time and achievable control performance on sites with low supply voltage.

Source
“Heating power at 200V 2,2 kW”
Installation

The unit requires a 200–240 V single- or two-phase supply with a maximum current of 14 A and a 16 A fuse.

Needed to size the branch circuit, cable, and protective device before installation.

Source
“Power supply requirement 200-240V 1~/2~ 50/60Hz max. current 14 A Fuse 16 A”
Installation

External branch circuit protection according to UL 489 is required.

Mandates a specific upstream breaker type for compliant electrical installation.

Source
“External branch circuit protection according to UL 489 required.”
Installation

Voltage and frequency tolerances are ±10% only if they do not act in opposite directions.

Prevents selecting a supply point where combined voltage and frequency deviations exceed the allowed envelope.

Source
“Voltage + / - 10%, as long as the frequency tolerance does not run in the opposite direction.”
Installation

Single- and two-phase devices are delivered with a power cable and country-specific plug that must be specified at ordering.

Avoids delays and rework because the correct plug must be requested with the order.

Source
“Single / two-phase devices (100V to 240V) --> with power cable and country-specific plug (please specify when ordering)”
Operation

The maximum permissible kinematic viscosity of the bath fluid is 50 mm²/s.

Limits the choice of thermofluid, especially at low temperatures where viscosity rises.

Source
“max. permissible kin. viscosity 50 mm²/s”
Operation

The minimum filling capacity is 5.8 l and the bath volume is 7.6 l, reduced to 5.2 l with a displacement rack.

Defines the required fluid charge and usable volume for samples or displacement inserts.

Source
“min. filling capacity 5,8 l Bath volume 7,6 l Bath capacity with displacement rack 5,2 l”
Installation

The pump connections are M16x1 male threads.

Determines the hose fittings and adapters needed for the external circuit.

Source
“Pump connection M16x1 male”
Performance

The pressure pump delivers up to 25 l/min at 0.7 bar head, while the suction pump delivers up to 18.5 l/min at 0.4 bar.

Used to check whether the pump can overcome the pressure drop of the external loop.

Source
“max. delivery 25 l/min Delivery pressure (head) 0,7 bar”
Installation

The unit is rated IP20 and must be operated between 5 °C and 40 °C ambient temperature.

Restricts installation location regarding dust, water ingress, and room climate.

Source
“Degree of Protection IP20 min. ambient temperature 5 °C max. ambient temperature 40 °C”
Safety

The unit has adjustable overtemperature protection according to DIN 12876 and safety classification III/FL.

Confirms built-in safety cut-out that must be set correctly for the fluid and process.

Source
“With adjustable overtemperature protection according to DIN 12876.”
Cumplimiento

The equipment is classified as pressure equipment category Art. 4.3 PED.

Relevant for CE compliance and for assessing whether the unit falls under the Pressure Equipment Directive.

Source
“Pressure equipment category Art. 4.3 PED”
Cumplimiento

The disturbance classification is EN55011 Class A, Group 1.

Class A equipment may require additional EMC measures in residential or light-industrial environments.

Source
“Classification (disturbance) to EN55011: Class A, Group 1”