Technical data per DIN 12876 with specification table
p. 2table, 4 rows
Technical data according to DIN 12876
Parameter
Value
Degree of Protection
IP20
min. ambient temperature
5 °C
max. ambient temperature
40 °C
from Serial-No.:
S393466
1.2/20
Engineering Facts
14 facts · every claim linked to a page
Selection
The operating temperature range depends on the cooling method: 28–300 °C without cooling, 20–300 °C with water cooling, and -20–300 °C with a refrigerator.
Determines whether the bath can reach the required process temperatures with the available cooling utility.
Source
“Operating temperature range 28...300 °C with water cooling 20...300 °C with refrigerator -20...300 °C”
Performance
Heating power is voltage-dependent, dropping from 3 kW at 230–240 V to 2.2 kW at 200 V.
Affects heat-up time and achievable control performance on sites with low supply voltage.
Source
“Heating power at 200V 2,2 kW”
Installation
The unit requires a 200–240 V single- or two-phase supply with a maximum current of 14 A and a 16 A fuse.
Needed to size the branch circuit, cable, and protective device before installation.
Source
“Power supply requirement 200-240V 1~/2~ 50/60Hz max. current 14 A Fuse 16 A”
Installation
External branch circuit protection according to UL 489 is required.
Mandates a specific upstream breaker type for compliant electrical installation.
Source
“External branch circuit protection according to UL 489 required.”
Installation
Voltage and frequency tolerances are ±10% only if they do not act in opposite directions.
Prevents selecting a supply point where combined voltage and frequency deviations exceed the allowed envelope.
Source
“Voltage + / - 10%, as long as the frequency tolerance does not run in the opposite direction.”
Installation
Single- and two-phase devices are delivered with a power cable and country-specific plug that must be specified at ordering.
Avoids delays and rework because the correct plug must be requested with the order.
Source
“Single / two-phase devices (100V to 240V) --> with power cable and country-specific plug (please specify when ordering)”
Operation
The maximum permissible kinematic viscosity of the bath fluid is 50 mm²/s.
Limits the choice of thermofluid, especially at low temperatures where viscosity rises.
Source
“max. permissible kin. viscosity 50 mm²/s”
Operation
The minimum filling capacity is 5.8 l and the bath volume is 7.6 l, reduced to 5.2 l with a displacement rack.
Defines the required fluid charge and usable volume for samples or displacement inserts.
Source
“min. filling capacity 5,8 l Bath volume 7,6 l Bath capacity with displacement rack 5,2 l”
Installation
The pump connections are M16x1 male threads.
Determines the hose fittings and adapters needed for the external circuit.
Source
“Pump connection M16x1 male”
Performance
The pressure pump delivers up to 25 l/min at 0.7 bar head, while the suction pump delivers up to 18.5 l/min at 0.4 bar.
Used to check whether the pump can overcome the pressure drop of the external loop.