The cooling water circuit needs a differential pressure of at least 3 bar and must not exceed 6 bar, so the building water supply must be checked.
Insufficient or excessive cooling water pressure will prevent correct operation or damage the unit.
Source
“min. cooling water differential pressure 3 bar”
Installation
The unit is designed for cooling water intake between 5 °C and 20 °C, and pipes must be insulated if the water can fall below the dew point.
Cooling water outside this range reduces cooling power or causes condensation, affecting reliability.
Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Performance
Cooling power drops from 3.3 kW at 0 °C to 1.6 kW at -20 °C and 0.6 kW at -40 °C, so low-temperature duty must be sized against the actual load.
Prevents oversizing expectations and ensures the process heat load is removable at the target temperature.
Source
“at -20°C 1,6 kW”
Performance
Cooling water consumption is 192 l/h at 0 °C flow and 138 l/h at -20 °C flow, which must be available continuously.
Sizes the cooling water supply and drainage and confirms utility availability.
Source
“Consumption at water 15°C, flow 0°C 192 l/h”
Operation
The pressure pump delivers up to 25 l/min at 0.7 bar and the suction pump up to 18.5 l/min at 0.4 bar, with M16x1 male connections.
Determines whether the external circuit flow resistance and hose connections are compatible.
Source
“max. delivery 25 l/min”
Operation
The maximum permissible kinematic viscosity of the bath fluid is 50 mm²/s, which limits fluid selection at low temperatures.
Using a too-viscous fluid at low temperature will overload the pump and degrade control.
Source
“max. permissible kin. viscosity 50 mm²/s”
Safety
The refrigerant R-452A is classified A1 with GHS H280 and has a GWP of 2141 with 2.1 t CO2 equivalent, so refrigerant handling and reporting obligations apply.
Affects site safety classification, leak detection and environmental reporting requirements.
Source
“R-452A (A1, H280)”
Safety
The unit has adjustable overtemperature protection according to DIN 12876 and safety classification III / FL.
Confirms the safety concept for flammable or sensitive bath fluids and for unattended operation.
Source
“With adjustable overtemperature protection according to DIN 12876.”
Installation
The unit has IP20 protection and must be operated between 5 °C and 40 °C ambient temperature.
Restricts installation location regarding dust, water ingress and room climate.
Source
“max. ambient temperature 40 °C”
Installation
The bath opening is 270 x 150 mm with 200 mm depth and the overall dimensions are 455x515x1014 mm at 105 kg net weight.
Needed for space planning, floor loading and access for installation and maintenance.
Source
“Overall dimensions WxDxH ** 455x515x1014 mm”
Cumplimiento
Voltage tolerance is ±5% only with a simultaneous frequency tolerance of ±2%, and combinations such as -5% voltage with +2% frequency are not allowed.
Prevents operation on unstable site power that could damage the compressor or electronics.
Source
“-5% voltage and +2% frequency -> not allowed!”
Installation
Three-phase devices below 63 A are supplied with a cable but without a plug, so the customer must provide the site plug.
Affects electrical installation scope and commissioning planning.
Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Cumplimiento
The equipment is classified Class A, Group 1 for electromagnetic disturbance to EN55011, which is intended for industrial environments.
May restrict use in residential or light-industrial areas without additional EMC measures.
Source
“Classification (disturbance) to EN55011: Class A, Group 1”