CC-905w Immersion Circulator Datasheet

Contenido

  1. Overview p. 1
    Introduces the CC-905w immersion circulator and its main features
  2. Spec pp. 1-2
    Operating limits, performance data and dimensions per DIN 12876
  3. p. 2 table, 21 rows
  4. p. 1 table, 33 rows

Technical data according to DIN 12876

ParameterValue
Pump connectionM16x1 male
max. permissible kin. viscosity50 mm²/s
Cooling water connectionG1/2 male
Consumption at water 15°C, flow 0°C170 l/h
min. cooling water differential pressure3 bar
max. cooling water pressure6 bar
Bath volume26 l
min. filling capacity19 l
Bath capacity with displacement rack15 l
Width bath opening WxD / bath depth260 x 260 / 200 mm
Height of bath opening937 mm
Overall dimensions WxDxH **605x706x1136 mm
Net weight175 kg
Power supply requirement (3 phase)400V 3~N 50Hz
max. current (3 Phase)13 A
Fuse (3 phase)3x16A
Pressure equipment categoryArt. 4.3 PED
Degree of ProtectionIP20
max. ambient temperature40 °C
min. ambient temperature5 °C
from Serial-No.:S566077

** Please respect space requirements. See operating conditions at www.huber-online.com

Technical data according to DIN 12876

ХарактеристикаЗначение
Operating temperature range-90...200 °C
Temperature stability at -10°C0,02 K
temperature set point / display5,7" colour Touchscreen
Internal temperature sensorPt100
Sensor external connectionPt100
Interface digitalEthernet, USB (Host u. Device), RS232
Safety classificationIII / FL
Heating power3 kW
Cooling power at 200°C2,5 kW
Cooling power at 100°C2 kW
Cooling power at 20°C2 kW
Cooling power at 0°C2 kW
Cooling power at -20°C1,9 kW
Cooling power at -40°C1,7 kW
Cooling power at -60°C1 kW
Cooling power at -80°C0,34 kW
Cooling power at -90°C0,13 kW
Refrigeration machinewater-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)R-452A (A1, H280)
Global Warming Potential (GWP)2141
Refrigerant quantity0,9 kg
CO2 equivalent1,9 t
Refrigerant 2nd stage (ASHRAE, GHS)R-508B (A1, H280)
Global Warming Potential (GWP)13396
Refrigerant quantity 2nd stage0,43 kg
CO2 equivalent 2nd stage5,8 t
UN-numberUN 2857
Pressure pumpyes
max. delivery25 l/min
max. delivery pressure0,7 bar
Suction pumpyes
max. delivery (suction)18,5 l/min
max. delivery pressure (suction)0,4 bar

Order-No.: CC-905w; 2027.0002.01

Engineering Facts

14 facts · every claim linked to a page
Installation

The unit requires a three-phase 400 V 3~N 50 Hz supply with a 3x16 A fuse, so the electrical installation must be sized accordingly.

Determines the site power supply, cable and breaker rating before ordering.

Source
“Power supply requirement (3 phase) 400V 3~N 50Hz”
Installation

The cooling water circuit needs a minimum differential pressure of 3 bar and must not exceed 6 bar, so the building water supply must be checked.

Insufficient or excessive cooling water pressure will prevent proper operation or damage the unit.

Source
“min. cooling water differential pressure 3 bar”
Installation

Cooling water intake temperature must be between 5 °C and 20 °C, and pipes must be insulated if water temperature can fall below the condensation point.

Defines the acceptable cooling water supply conditions and condensation protection measures.

Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Performance

Cooling power drops significantly at low temperatures, from 2 kW at 20 °C to only 0.13 kW at -90 °C, so process heat loads must be checked against the actual operating temperature.

Prevents oversizing expectations and ensures the unit can remove the process heat at the required low temperature.

Source
“at -90°C 0,13 kW”
Installation

The unit is water-cooled and uses refrigerants R-452A and R-508B with CO2 equivalents of 1.9 t and 5.8 t, which matters for refrigerant regulations and leak monitoring.

Affects compliance with F-gas regulations and the need for leak detection or reporting.

Source
“CO2 equivalent 2nd stage 5,8 t”
Operation

The pressure pump delivers up to 25 l/min at 0.7 bar and the suction pump up to 18.5 l/min at 0.4 bar, so external circuit pressure losses must be evaluated.

Determines whether the pump can overcome the pressure drop of the external application loop.

Source
“max. delivery 25 l/min”
Operation

The maximum permissible kinematic viscosity of the bath fluid is 50 mm²/s, so the heat transfer fluid must be selected within this limit.

Ensures the pump can circulate the fluid and that heat transfer remains adequate.

Source
“max. permissible kin. viscosity 50 mm²/s”
Installation

The bath requires a minimum filling capacity of 19 l and holds 26 l, with only 15 l when a displacement rack is used, so the fluid volume and expansion must be planned.

Prevents insufficient filling and overflow during temperature changes.

Source
“min. filling capacity 19 l”
Installation

The unit has IP20 protection and must be operated between 5 °C and 40 °C ambient temperature, so it cannot be placed in a wet or hot environment.

Restricts installation location regarding moisture and ambient temperature.

Source
“Degree of Protection IP20”
Installation

The overall dimensions are 605x706x1136 mm and the net weight is 175 kg, so floor space and load-bearing capacity must be verified.

Needed for site planning, access routes and floor loading.

Source
“Overall dimensions WxDxH ** 605x706x1136 mm”
Cumplimiento

Voltage tolerance is ±5% with simultaneous frequency tolerance of ±2%, but the combination of -5% voltage and +2% frequency is not allowed.

Defines the acceptable power quality window for reliable operation.

Source
“-5% voltage and +2% frequency -> not allowed!”
Installation

Three-phase devices with current consumption less than 63 A are delivered with cable but without plug, so a suitable plug must be provided on site.

Affects electrical installation scope and commissioning.

Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Safety

The unit has adjustable overtemperature protection according to DIN 12876 and safety classification III/FL, which is relevant for unattended operation and flammable fluids.

Confirms built-in safety functions required for protected operation.

Source
“With adjustable overtemperature protection according to DIN 12876.”
Operation

The pressure equipment category is Art. 4.3 PED, which determines the applicable pressure vessel compliance requirements.

Relevant for compliance documentation and acceptance at site.

Source
“Pressure equipment category Art. 4.3 PED”