CC-E Immersion Circulator

Contenido

  1. Overview p. 1
    Product introduction and key features of CC-E circulator
  2. Spec p. 2
    Specifications according to DIN 12876 standard
  3. p. 1 table, 34 rows

Technical data according to DIN 12876

ParameterValue
Operating temperature range25...200 °C
with water cooling20...200 °C
with refrigerator-30...200 °C
Temperature stability at 70°C0,01 K
temperature set point / display5,7" colour Touchscreen
Absolute accuracysetup for calibration
Internal temperature sensorPt100
Sensor external connectionPt100
Interface digitalEthernet, USB (Host u. Device), RS232
Safety classificationIII / FL
Heating power at 240V2,1 kW
Heating power at 230V2 kW
Heating power at 220V1,8 kW
Heating power at 208V1,6 kW
Heating power at 200V1,5 kW
Pressure pumpyes
max. delivery27 l/min
max. delivery pressure0,7 bar
Suction pumpyes
max. delivery (suction)22 l/min
max. delivery pressure (suction)0,4 bar
Pump connenction (optional)M16x1 male
Immersion depth150 mm
Overall dimensions WxDxH **132x159x315 mm
Net weight4 kg
Power supply requirement200-240V 1~/2~ 50/60Hz
Power input2,1 kW
max. current10 A
min. Fuse10A
max. Fuse16A
Pressure equipment categoryN/A
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C

** Overall dimensions WxDxH

Engineering Facts

13 facts · every claim linked to a page
Selection

The operating temperature range depends on the cooling method: 25–200 °C without cooling, 20–200 °C with water cooling, and -30–200 °C with a refrigerator.

Determines whether the unit can reach the required process temperature with the available cooling utility.

Source
“Operating temperature range 25...200 °C with water cooling 20...200 °C with refrigerator -30...200 °C”
Performance

Heating power scales with supply voltage from 1.5 kW at 200 V to 2.1 kW at 240 V, so available mains voltage affects achievable heating rate.

Needed to verify the electrical supply can deliver the rated heating power and to predict performance at the site voltage.

Source
“Heating power at 240V 2,1 kW Heating power at 230V 2 kW Heating power at 220V 1,8 kW Heating power at 208V 1,6 kW Heating power at 200V 1,5 kW”
Installation

The unit requires a 200–240 V single- or two-phase 50/60 Hz supply with a minimum 10 A fuse and maximum 16 A fuse.

Defines the branch circuit, breaker sizing, and supply configuration required for safe installation.

Source
“Power supply requirement 200-240V 1~/2~ 50/60Hz Power input 2,1 kW max. current 10 A min. Fuse 10A max. Fuse 16A”
Installation

Voltage and frequency tolerances are ±10% but must not combine in opposite directions, so a -10% voltage with +3% frequency is not allowed.

Prevents installation on a supply whose combined voltage/frequency deviation would exceed the motor and electronics rating.

Source
“Voltage + / - 10%, as long as the frequency tolerance does not run in the opposite direction.”
Cumplimiento

External branch circuit protection according to UL 489 is required, so the installation must include a UL 489 listed breaker.

Mandatory for compliance and for passing electrical inspection in UL jurisdictions.

Source
“External branch circuit protection according to UL 489 required.”
Installation

Single- and two-phase devices are delivered with a power cable and country-specific plug that must be specified at ordering.

Ensures the correct plug is supplied and avoids rework or adapters at commissioning.

Source
“Single / two-phase devices (100V to 240V) --> with power cable and country-specific plug (please specify when ordering)”
Operation

The pressure pump delivers up to 27 l/min at 0.7 bar and the suction pump up to 22 l/min at 0.4 bar, defining the external circuit flow capability.

Used to check whether the pump can overcome the pressure drop of the external application loop.

Source
“max. delivery 27 l/min max. delivery pressure 0,7 bar Suction pump yes max. delivery (suction) 22 l/min max. delivery pressure (suction) 0,4 bar”
Installation

The immersion depth is 150 mm, so the bath or vessel must accommodate the immersion thermostat to at least this depth.

Determines minimum bath depth and mounting position for correct temperature sensing and safe operation.

Source
“Immersion depth 150 mm”
Safety

The unit has IP20 protection and must be operated between 5 °C and 40 °C ambient, so it cannot be used in wet or hot environments.

Restricts installation location and rules out washdown or outdoor placement.

Source
“Degree of Protection IP20 min. ambient temperature 5 °C max. ambient temperature 40 °C”
Operation

Polyglycol must not be used as heat transfer fluid when a Huber pump adapter is fitted.

Prevents selection of an incompatible fluid that could damage the pump adapter or seals.

Source
“When using Huber pump adapter: Polyglycol is not permissible to be used as a heat transfer fluid”
Safety

The unit provides adjustable overtemperature protection according to DIN 12876, which must be set for the application.

Critical safety function that must be configured during commissioning to protect the bath and process.

Source
“With adjustable overtemperature protection according to DIN 12876.”
Selection

The safety classification is III/FL, which constrains the permissible heat transfer fluid and application setup.

Determines which fluids and operating conditions are allowed under DIN 12876.

Source
“Safety classification III / FL”
Installation

Space requirements must be respected per the operating conditions published online, so clearance around the 132x159x315 mm unit must be planned.

Prevents overheating and allows access for service and ventilation.

Source
“Please respect space requirements. See operating conditions at www.huber-online.com”