Installation
The chiller requires a three-phase 400 V supply with a 3x16 A fuse, so the site must provide a suitable three-phase feed and protection.
Determines electrical supply and upstream protection design before installation.
Source
“Power supply (3 Phase) 400V 3~ 50Hz”
Installation
Three-phase devices drawing less than 63 A are delivered with a cable but without a plug, so the installer must provide and fit the site plug.
Affects electrical installation scope and what must be ordered or prepared on site.
Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Installation
The unit must be operated within an ambient temperature window of 5 to 32 °C, so the plant room must be kept within this range.
Ambient conditions affect cooling performance and reliable operation.
Source
“min. ambient temperature 5 °C”
Performance
Cooling capacity is rated at 20 °C ambient and drops as ambient rises, so selection must account for actual room temperature.
Prevents undersizing when the chiller is installed in a warmer environment.
Source
“If the ambient temperature rises, the cooling capacity may drop.”
Installation
The chiller needs 160 to 175 litres of filling capacity, so the loop volume and expansion provisions must be planned accordingly.
Determines system volume, fluid quantity and expansion tank sizing.
Source
“min. filling capacity 160 l”
Operation
The recommended thermofluid is a 50:50 water-monoethylene glycol mixture, so the fluid must be selected to match this recommendation.
Fluid choice affects freeze protection, heat transfer and pump performance.
Source
“Recommended thermofluid: Water - Monoethylene Glycol 50:50”
Safety
The refrigerant is R-449A classified A1 with GHS H280, so refrigerant handling and ventilation precautions must follow this classification.
Needed for safety assessment, leak response and refrigerant documentation.
Source
“R-449A (A1, H280)”
Cumplimiento
The unit is EN55011 Class A Group 1, so it is intended for industrial environments and may require EMC measures in residential areas.
Affects EMC compliance and suitability of the installation location.
Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Installation
Voltage and frequency tolerances are ±5% and ±2% but not simultaneously in the unfavourable combination, so the supply quality must be checked.
Ensures the electrical supply stays within the allowed operating envelope.
Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Installation
The pump connection is G3/4 male, so mating hydraulic fittings must be G3/4 female.
Determines the connection hardware needed for the hydraulic circuit.
Source
“Pump connection G3/4 male”
Operation
Flow rate is adjusted via a manual bypass valve on the back of the chiller, so rear access must be kept clear.
Requires installation clearance at the rear for commissioning and adjustment.
Source
“The flow rate can be adjusted via the manual bypass valve on the backside of the chiller.”
Installation
The unit produces about 70 dB(A) sound pressure, so acoustic treatment or separation from occupied areas may be needed.
Affects placement and noise control in the plant room.
Source
“sound pressure level +/- 4 dB(A) 70 dB(A)”
Cumplimiento
The equipment is compliant with US-SNAP for industrial process refrigeration, so its end use must match this application.
Confirms the permitted application for regulatory compliance.
Source
“The US-SNAP end-use for this equipment is the industrial process refrigeration.”