CS 200 Circulating Bath — Technical Datasheet

Contenido

  1. Overview p. 1
    Product introduction and general description of CS 200 bath
  2. Spec p. 2
    Technical data according to DIN 12876 standard
  3. p. 1 table, 33 rows

CS 200 by Van der Heijden — Technical data according to DIN 12876

ParameterValue
Order-No.VDH1101105
Operating temperature range-20...15 °C
temperature set point / displaycolour LED Touchscreen
Internal temperature sensorPt100
Temperature stability at -10°C0,5 K
Interface digitalRS232
Safety classificationI / NFL
Cooling power at ambient temperature 20°C
at 15°C28 kW
at 10°C20 kW
at 0°C18 kW
at -10°C15 kW
at -20°C9,5 kW
Refrigeration machineair-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)R-449A (A1, H280)
Refrigerant quantity12 kg
CO2 equivalent16,76 t
Circulation pump:
max. delivery105 l/min
max. delivery pressure4,7 bar
Pump connectionG1 male
min. filling capacity225 l
max. filling capacity245 l
Overall dimensions WxDxH **1400x1000x1800 mm
Net weight510 kg
sound pressure level +/- 4 dB(A)78 dB(A)
Power supply (3 Phase)400V 3~ 50Hz
max. current (3 Phase)23 A
Fuse (3 phase)3x32A
Degree of ProtectionIP52
min. ambient temperature5 °C
max. ambient temperature32 °C
from Serial-No.1.0/24

** Overall dimensions WxDxH

Engineering Facts

13 facts · every claim linked to a page
Selection

The chiller's cooling capacity is rated at 20°C ambient and will drop if ambient temperature rises, so selection must account for actual site ambient conditions.

Cooling capacity must be derated for higher ambient temperatures to avoid undersizing.

Source
“Output data valid for: Room temperature 20°C. If the ambient temperature rises, the cooling capacity may drop.”
Installation

The unit requires a 3-phase 400V 50Hz supply with max current 23 A and 3x32A fuses, so the electrical installation must be sized accordingly.

Electrical supply and protection must match the unit's requirements for safe installation.

Source
“Power supply (3 Phase) 400V 3~ 50Hz”
Installation

Voltage and frequency tolerances are interdependent: -5% voltage with +2% frequency is not allowed, but -5% voltage with -2% frequency is allowed.

Site power quality must be checked against these combined tolerances to avoid invalid supply conditions.

Source
“-5% voltage and +2% frequency -> not allowed!”
Installation

Three-phase devices drawing less than 63A are supplied with cable but without a plug, so a suitable plug must be provided on site.

The installer must plan for a compatible plug and connection point.

Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Installation

The unit has IP52 protection and must be installed in ambient temperatures between 5°C and 32°C.

Installation location must respect the ambient temperature range and IP rating.

Source
“min. ambient temperature 5 °C”
Operation

The recommended thermofluid is a 50:50 water-monoethylene glycol mixture, which must be used for proper operation.

Using the correct fluid ensures heat transfer performance and protects the system.

Source
“Recommended thermofluid: Water - Monoethylene Glycol 50:50”
Operation

The flow rate can be adjusted via a manual bypass valve on the backside of the chiller, allowing balancing of the circuit.

Operators need to know how to adjust flow to match process requirements.

Source
“The flow rate can be adjusted via the manual bypass valve on the backside of the chiller.”
Safety

The refrigerant R-449A is classified A1 with GHS hazard H280, and the system contains 12 kg (16.76 t CO2 equivalent), which affects handling and leak response.

Refrigerant safety classification and charge size determine safety measures and environmental reporting.

Source
“R-449A (A1, H280)”
Safety

The unit has safety classification I/NFL according to DIN 12876, indicating suitability for flammable liquids only under restricted conditions.

This classification limits the type of fluids and applications the chiller can safely handle.

Source
“Safety classification I / NFL”
Performance

The circulation pump delivers up to 105 l/min at 4.7 bar, which must be sufficient for the process loop pressure drop.

Pump pressure and flow determine whether the chiller can overcome the external circuit resistance.

Source
“max. delivery pressure 4,7 bar”
Performance

The internal tank holds between 225 l and 245 l, so the system volume and expansion must be planned within this range.

Filling capacity affects fluid inventory, expansion volume, and system design.

Source
“min. filling capacity 225 l”
Cumplimiento

The equipment emits Class A, Group 1 disturbances per EN55011, which may require additional EMC measures in residential or sensitive environments.

EMC classification determines installation environment restrictions and potential filtering needs.

Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Installation

Space requirements must be respected as per operating conditions, so clearance around the 1400x1000x1800 mm unit must be planned.

Adequate clearance is needed for airflow, maintenance access, and safe operation.

Source
“Please respect space requirements. See operating conditions at www.huber-online.com”