CS 25-H Thermostatic Bath/Circulator Datasheet

Contenido

  1. Overview p. 1
    Product introduction and general description of CS 25-H
  2. Spec p. 2
    Specifications according to DIN 12876 standard
  3. p. 1 table, 35 rows

Technical data according to DIN 12876

CharacteristicValue
Order-No.VDH4100342
CS 25-H by Van der Heijden
Operating temperature range-20...85 °C
temperature set point / displaycolour LED Touchscreen
Internal temperature sensorPt100
Temperature stability at -10°C0,5 K
Interface digitalRS232
Safety classificationI / NFL
Heating power1 kW
Cooling power at ambient temperature 20°C
at 15°C3,5 kW
at 10°C2,5 kW
at 0°C1,6 kW
at -10°C0,98 kW
at -20°C0,7 kW
Refrigeration machineair-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)R-449A (A1, H280)
Refrigerant quantity2 kg
CO2 equivalent2,79 t
Circulation pump:
max. delivery40 l/min
max. delivery pressure4,5 bar
Pump connectionG1/2 male
min. filling capacity45 l
max. filling capacity51 l
Overall dimensions WxDxH **590x620x1210 mm
Net weight140 kg
sound pressure level +/- 4 dB(A)64 dB(A)
Power supply requirement230V 1~ 50Hz
max. current15,5 A
Fuse16 A
Degree of ProtectionIP52
min. ambient temperature5 °C
max. ambient temperature32 °C
from Serial-No.:1.0/24

** Technical details and dimensions are subject to change. No liability is accepted for errors or omissions. Illustrations can deviate from the original.

Engineering Facts

12 facts · every claim linked to a page
Selection

The chiller's cooling capacity is rated at 20°C ambient and drops as ambient temperature rises, so sizing must account for the actual room temperature.

Prevents undersizing when the unit is installed in a warm room.

Source
“Output data valid for: Room temperature 20°C. If the ambient temperature rises, the cooling capacity may drop.”
Installation

The unit requires a 230 V single-phase 50 Hz supply with a 16 A fuse and draws up to 15.5 A, so the electrical circuit must be dedicated and adequately rated.

Determines cable, breaker and supply sizing during installation.

Source
“Power supply requirement 230V 1~ 50Hz max. current 15,5 A Fuse 16 A”
Installation

Voltage and frequency tolerances are coupled: ±5% voltage is allowed only with ±2% frequency, and certain combinations are explicitly not allowed.

Avoids installing on a supply whose combined voltage/frequency deviation exceeds EN60034-1 limits.

Source
“-5% voltage and +2% frequency -> not allowed!”
Installation

The chiller is IP52 and must be operated only between 5°C and 32°C ambient, which constrains the installation location.

Ensures the site environment meets protection and ambient limits.

Source
“min. ambient temperature 5 °C max. ambient temperature 32 °C”
Installation

The refrigerant is R-449A (A1, H280) with 2 kg charge and 2.79 t CO2 equivalent, which affects safety classification and leak/CO2 reporting obligations.

Needed for refrigerant safety assessment and CO2-equivalent documentation.

Source
“R-449A (A1, H280)”
Installation

The pump connection is G1/2 male and the tank holds 45–51 L, so hoses and fluid volume must be planned accordingly.

Determines hose fittings and heat-transfer fluid quantity for commissioning.

Source
“Pump connection G1/2 male”
Operation

Flow rate is adjusted via a manual bypass valve on the back of the chiller, so rear access is needed for flow balancing.

Requires rear clearance for commissioning and flow adjustment.

Source
“The flow rate can be adjusted via the manual bypass valve on the backside of the chiller.”
Operation

The feed pump and cooling unit can be operated separately, allowing pump-only circulation without refrigeration.

Enables independent pump operation for testing or process needs.

Source
“Separate operating option for the feed pump and the cooling unit with convenient adjustment of the desired setpoint.”
Operation

The recommended thermofluid is a 50:50 water–monoethylene glycol mixture, which sets the fluid choice for the operating range.

Guides correct heat-transfer fluid selection to avoid freezing or corrosion.

Source
“Recommended thermofluid: Water - Monoethylene Glycol 50:50”
Cumplimiento

The unit is EMC Class A Group 1, so it may cause radio interference in residential environments and is intended for non-domestic use.

Affects site suitability and EMC compliance in residential areas.

Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Installation

Single/two-phase devices are delivered with a country-specific plug that must be specified at order, while three-phase units under 63 A come without a plug.

Ensures the correct power connection is ordered and available at installation.

Source
“Single / two-phase devices (100V to 240V) --> with power cable and country-specific plug (please specify when ordering)”
Installation

Space requirements around the unit must be respected per the operating conditions, so clearance must be planned before siting.

Prevents airflow restriction and ensures service access.

Source
“Please respect space requirements. See operating conditions at www.huber-online.com”