The unit is rated for an operating temperature range of -40 to 200 °C, so the application must stay within these limits.
Defines whether the circulator can cover the required process temperatures.
Source
“Operating temperature range -40...200 °C”
| Parameter | Value |
|---|---|
| max. delivery pressure | 0,9 bar |
| Pump connection | M24x1,5 male |
| max. permissible kin. viscosity | 50 mm²/s |
| min. filling capacity | 1,5 l |
| Filling capacity expansion tank | 3 l |
| Overall dimensions WxDxH ** | 295x530x570 mm |
| Net weight | 53 kg |
| sound pressure level +/- 4 dB(A) | 61 dB(A) |
| Power supply requirement | 208-240V 1~/2~ 50/60Hz |
| max. current | 11,5 A |
| Fuse | 16 A |
| Pressure equipment category | Art. 4.3 PED |
| Degree of Protection | IP20 |
| min. ambient temperature | 5 °C |
| max. ambient temperature | 40 °C |
| from Serial-No.: | S393363 |
** Please respect space requirements. See operating conditions at www.huber-online.com
| Characteristic | Value |
|---|---|
| Operating temperature range | -40...200 °C |
| Temperature stability at -10°C | 0,01 K |
| temperature set point / display | 5,7" colour Touchscreen |
| Resolution of display | 0,01 K |
| Internal temperature sensor | Pt100 |
| Sensor external connection | Pt100 |
| Interface digital | Ethernet, USB (Host u. Device), RS232 |
| digital input | ECS ONE |
| digital output | POKO ONE |
| Alarm message | optic, acoustic, relay |
| Safety classification | III / FL |
| Heating power at 240V | 2 kW |
| Heating power at 230V | 1,8 kW |
| Heating power at 220V | 1,6 kW |
| Heating power at 208V | 1,5 kW |
| Heating power convertible | 1,5 - 2 kW |
| Cooling power with Thermooil at 200°C | 0,6 kW |
| Cooling power with Thermooil at 150°C | 0,6 kW |
| Cooling power with Thermooil at 100°C | 0,6 kW |
| Cooling power with Thermooil at 50°C | 0,6 kW |
| Cooling power with Thermooil at 20°C | 0,6 kW |
| Cooling power with Ethanol at 0°C | 0,6 kW |
| Cooling power with Ethanol at -20°C | 0,35 kW |
| Cooling power with Ethanol at -30°C | 0,2 kW |
| Cooling power with Ethanol at -40°C | 0,04 kW |
| Refrigeration machine | air-cooled, natural refrigerant |
| Refrigerant (ASHRAE, GHS) | R-290 (A3, H220) |
| Global Warming Potential (GWP) | 0,02 |
| Refrigerant quantity | 0,099 kg |
| CO2-Äquvivalent | 0,0 t |
| UN-number | UN 3358 |
| Circulation pump: max. delivery | 47 l/min |
The unit is rated for an operating temperature range of -40 to 200 °C, so the application must stay within these limits.
Defines whether the circulator can cover the required process temperatures.
“Operating temperature range -40...200 °C”
Cooling capacity drops sharply at low temperatures with ethanol, from 0.6 kW at 0 °C to only 0.04 kW at -40 °C.
Low-temperature heat loads may exceed available cooling capacity near the lower limit.
“at -40°C 0,04 kW”
The refrigerant is R-290 (A3, H220), a flammable natural refrigerant, which affects installation and safety planning.
Flammable refrigerant classification requires appropriate site safety measures.
“R-290 (A3, H220)”
The unit is air-cooled, so sufficient ventilation and respect for space requirements are needed at the installation site.
Air-cooled condensers need adequate airflow and clearances to maintain cooling performance.
“Refrigeration machine air-cooled, natural refrigerant”
The pump connection is M24x1.5 male, so hoses or adapters must match this thread.
Prevents mismatch with existing pipework or flexible tempering hoses.
“Pump connection M24x1,5 male”
The maximum permissible kinematic viscosity of the heat transfer fluid is 50 mm²/s, limiting fluid selection at low temperatures.
Ensures the pump can circulate the fluid properly without excessive load.
“max. permissible kin. viscosity 50 mm²/s”
The minimum filling capacity is 1.5 l and the expansion tank holds 3 l, which sets the required fluid charge.
Avoids underfilling or overflow during thermal expansion.
“min. filling capacity 1,5 l”
The unit requires a 208-240 V single-phase or two-phase supply with a maximum current of 11.5 A and a 16 A fuse.
Determines electrical supply and protection sizing for the installation.
“max. current 11,5 A”
The degree of protection is IP20, so the unit must be installed in a dry, protected environment.
IP20 is not water-protected and is unsuitable for wet or washdown areas.
“Degree of Protection IP20”
The permissible ambient temperature range is 5 to 40 °C, and cooling capacity may drop if ambient temperature rises.
High ambient temperature reduces cooling performance and may limit operation.
“max. ambient temperature 40 °C”
Voltage tolerance is ±5% only with a simultaneous frequency tolerance of ±2%, and certain combinations are not allowed.
Ensures the electrical supply stays within the allowed tolerance envelope.
“-5% voltage and +2% frequency -> not allowed!”
Single- and two-phase devices are delivered with a power cable and country-specific plug, which must be specified when ordering.
Prevents delivery of the wrong plug type for the installation country.
“with power cable and country-specific plug (please specify when ordering)”
The equipment is classified as EN55011 Class A, Group 1 for electromagnetic disturbance.
Class A equipment may require EMC measures in residential or sensitive environments.
“Classification (disturbance) to EN55011: Class A, Group 1”