Induztr
Resumen técnico Visor de documentos Modelos relacionados

HB240 Heating Bath Circulator

Contenido

  1. Overview p. 1
    Product introduction and model identification for HB240
  2. Spec p. 2
    Technical data per DIN 12876 including performance parameters
  3. p. 1 table, 21 rows

Technical data according to DIN 12876

ParameterValue
Order-No.:HB240
Operating temperature range60...250 °C
Resolution of display0,01 K
Temperature stability at 70°C0,1 K
temperature set point / display5,7" colour Touchscreen
Absolute accuracysetup for calibration
Internal temperature sensorPt100
External sensorPt100
Interface digitalEthernet, USB (Host u. Device), RS232
Safety classificationII / FL
Heating power24 kW
Circulation pump:MK pump
max. delivery100 l/min
max. delivery pressure3,5 bar
Pump connectionM30x1,5 male
Overall dimensions WxDxH570x700x1171 mm
Degree of ProtectionIP20
Power supply (3 Phase)400V 3~ 50Hz
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:1.0/25

2063.0005.01

Engineering Facts

12 facts · every claim linked to a page
Selection

The unit is designed exclusively for external open operation and has no expansion tank, so the application must provide filling and volume compensation.

Determines whether the unit can be used with a closed loop or requires an external open bath/reservoir with level and expansion volume.

Source
“The machine is developed exclusively for external open operation, and does not possess an expansion tank.”
Installation

The three-phase power supply is 400 V 3~ 50 Hz and the device is delivered with cable but without a plug, so the customer must provide a suitable 3-phase connection and plug.

Affects electrical installation planning, plug selection and connection by a qualified electrician.

Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Installation

Voltage and frequency tolerances are ±10% only if they do not act in opposite directions, so supply quality must be checked before commissioning.

Prevents nuisance faults or damage when the site supply deviates from nominal voltage and frequency simultaneously.

Source
“Voltage + / - 10%, as long as the frequency tolerance does not run in the opposite direction.”
Cumplimiento

The unit is EMC class A group 1, so it may cause radio interference in residential areas and is intended for industrial environments.

Affects site selection and EMC coordination, especially if installed near sensitive electronics or in residential areas.

Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Safety

Leakage current from external temperature control must be considered, so proper earthing and RCD selection are required.

Warns the installer that leakage current from the external application may affect protective device selection and earthing.

Source
“Attention: leakage current externe Temperierung”
Operation

The minimum temperature of 60 °C depends on ambient temperature being below 60 °C and on application heat loss between 0.1 kW and 2.8 kW depending on model.

The user must verify that the external application heat loss is within the allowed range, otherwise 60 °C cannot be reached.

Source
“The heat loss must be between 0.1kW and 2.8kW, depending on the HB model, to reach the minimum temperature of 60°C.”
Installation

The pump connection is M30x1.5 male, so mating hoses or adapters must match this thread.

Ensures correct selection of connection hoses and avoids leaks or incompatible fittings during installation.

Source
“Pump connection M30x1,5 male”
Operation

The ambient temperature must remain between 5 °C and 40 °C for safe operation.

Defines the allowable room conditions and prevents operation outside the specified ambient range.

Source
“min. ambient temperature 5 °C”
Performance

The circulation pump delivers up to 100 l/min at up to 3.5 bar, which must be checked against the pressure drop of the external circuit.

Confirms whether the pump can overcome the hydraulic resistance of the external application.

Source
“max. delivery pressure 3,5 bar”
Selection

The unit has 24 kW heating power and a 60...250 °C operating range, so the electrical supply and heat load must be sized accordingly.

Affects electrical supply sizing and whether the required heating rate and temperature can be achieved.

Source
“Heating power 24 kW”
Safety

The unit has adjustable overtemperature protection according to DIN 12876 and safety classification II/FL, which must be set for the application.

The installer must configure the overtemperature limit to protect the external application and fluid.

Source
“With adjustable overtemperature protection according to DIN 12876.”
Maintenance

The unit uses stainless steel for heating, housing and pump, which affects fluid compatibility and cleaning procedures.

Guides selection of thermofluid and maintenance chemicals to avoid corrosion of wetted stainless-steel parts.

Source
“Heating, housing and pump of stainless steel.”