Contenido

  1. Overview p. 1
    Product introduction and main features of HTS 3
  2. Spec p. 2
    Technical data according to DIN 12876
  3. p. 1 table, 30 rows

HTS 3 — Technical data according to DIN 12876

ParameterValue
Operating temperature range ( secondary side )(3)...(95) °C
Temperature stability0,1 K
temperature set point / display5,7" colour Touchscreen
Internal temperature sensorPt100
Sensor external connectionPt100
Interface digitalEthernet, USB (Host u. Device), RS232
Alarm messageoptic, acoustic, relay
Safety classificationI / NFL
Cooling power with Water at 20°C3 kW
Circulation pumpImmersion pump
max. delivery33 l/min
max. delivery pressure0,7 bar
Pump connectionM16x1 male
max. permissible kin. viscosity50 mm²/s
Cooling water connectionG1/2 male
Consumption at water 10°C, flow 20°C450 l/h
min. cooling water differential pressure2 bar
max. cooling water pressure6 bar
min. filling capacity3,5 l
Volume of expansion2 l
Overall dimensions WxDxH **280x491x414 mm
Net weight21 kg
Power supply requirement200-240V 1~/2~ 50/60Hz
max. current1 A
min. Fuse10A
max. Fuse16A
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:S398904

** Overall dimensions WxDxH

Engineering Facts

12 facts · every claim linked to a page
Installation

The unit requires cooling water with a minimum differential pressure of 2 bar and must not exceed 6 bar, so the site water supply must be verified before installation.

Ensures the cooling water circuit can deliver adequate flow without overpressurizing the heat exchanger.

Source
“min. cooling water differential pressure 2 bar”
Installation

Cooling water inlet temperature must be between 5°C and 20°C, and pipes must be insulated if condensation conditions occur.

Prevents condensation, corrosion, and performance loss; defines site cooling water requirements.

Source
“This temperature control unit has been designed to operate with cooling water intake temperature between 5°C and 20°C.”
Selection

The plastic pump is not compatible with acetone or polyglycols, so alternative fluids or a more resistant pump option must be selected.

Avoids chemical attack and premature pump failure when choosing the heat transfer fluid.

Source
“The plastic pump is not resistant against acetone and polyglycols (depending on the manufacturer).”
Installation

The electrical supply must be 200–240 V single/two-phase 50/60 Hz with a fuse rated between 10 A and 16 A.

Defines the site power and protection requirements for safe connection.

Source
“Power supply requirement 200-240V 1~/2~ 50/60Hz”
Cumplimiento

Voltage and frequency tolerances must not oppose each other; for example, -10% voltage with +3% frequency is not allowed.

Prevents motor overheating and ensures compliant operation per EN60034-1.

Source
“-10% voltage and +3% frequency -> not allowed !”
Installation

The unit is rated IP20 and must be installed in an ambient temperature between 5°C and 40°C.

Defines the permissible operating environment and protection against ingress.

Source
“min. ambient temperature 5 °C”
Operation

The cooling water circuit contains copper, stainless steel 1.4401, MS, PA, PPE, PTFE, and EPDM, so only compatible cooling water may be used.

Guides water quality selection to avoid corrosion and material degradation.

Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.”
Performance

The circulation pump delivers up to 33 l/min at 0.7 bar, which must be checked against the external circuit pressure drop.

Ensures the pump can overcome the resistance of the external application loop.

Source
“max. delivery 33 l/min”
Installation

The cooling water connection is G1/2 male and the pump connection is M16x1 male, so matching adapters and hoses are required.

Prevents delays and leaks by specifying correct fittings before installation.

Source
“Cooling water connection G1/2 male”
Selection

The maximum permissible kinematic viscosity of the heat transfer fluid is 50 mm²/s.

Ensures the pump can circulate the fluid and maintain cooling performance.

Source
“max. permissible kin. viscosity 50 mm²/s”
Operation

The unit is classified as Class A, Group 1 for electromagnetic disturbance per EN55011.

Informs EMC planning and potential restrictions in residential environments.

Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Installation

Three-phase devices with current consumption less than 63 A are supplied with cable but without a plug.

Requires the installer to provide a suitable plug and connection.

Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”