Contenido

  1. Overview p. 1
    Product introduction and general description of the CS 100-H unit
  2. Spec p. 2
    Technical data according to DIN 12876 with performance parameters
  3. p. 1 table, 35 rows

Technical data according to DIN 12876

ParameterValue
Operating temperature range-20...85 °C
temperature set point / displaycolour LED Touchscreen
Internal temperature sensorPt100
Temperature stability at -10°C1,5 K
Interface digitalRS232
Safety classificationI / NFL
Heating power5,5 kW
Cooling power at ambient temperature 20°C
at 15°C18 kW
at 10°C10 kW
at 0°C9,5 kW
at -10°C7,5 kW
at -20°C4,5 kW
Refrigeration machineair-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)R-449A (A1, H280)
Refrigerant quantity4,5 kg
CO2 equivalent6,29 t
Circulation pump: max. delivery83 l/min
max. delivery pressure3,7 bar
Pump connectionG3/4 male
min. filling capacity160 l
max. filling capacity175 l
Overall dimensions WxDxH **800x850x1665 mm
Net weight340 kg
sound pressure level +/- 4 dB(A)70 dB(A)
Power supply (3 Phase)208V 3~ 60Hz
max. current (3 Phase)35 A
Fuse (3 phase)3x35A
power supply convertible (3 phase)460V 3~ 60Hz
max. current convertible (3 phase)16 A
fuse convertible (3 phase)3x20 A
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature32 °C
from Serial-No.:1.0/20

** Technical details and dimensions are subject to change. No liability is accepted for errors or omissions. Illustrations can deviate from the original.

Engineering Facts

12 facts · every claim linked to a page
Installation

The chiller requires a three-phase 208 V 60 Hz supply with a 35 A maximum current and 3x35 A fusing, or a convertible 460 V 60 Hz supply with 16 A and 3x20 A fusing.

Determines site electrical infrastructure, breaker sizing, and whether a voltage conversion is needed before installation.

Source
“Power supply (3 Phase) 208V 3~ 60Hz”
Installation

Three-phase devices drawing less than 63 A are delivered with a cable but without a plug, so the customer must provide and wire the plug.

Affects electrical installation planning and the need to procure a suitable plug and have a qualified electrician terminate it.

Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Installation

The unit must be operated within an ambient temperature range of 5 °C to 32 °C.

Room conditions must be verified; operation outside this range risks reduced cooling capacity or equipment protection trips.

Source
“min. ambient temperature 5 °C”
Performance

Cooling capacity is rated at 20 °C room temperature and drops if ambient temperature rises.

Sizing must account for actual plant ambient conditions, otherwise the chiller may be undersized in warm rooms.

Source
“Output data valid for: Room temperature 20°C. If the ambient temperature rises, the cooling capacity may drop.”
Installation

The chiller needs 800 x 850 x 1665 mm of space and weighs 340 kg net, so floor loading and clearance must be planned.

Determines rigging, floor loading, and access clearance for installation and maintenance.

Source
“Overall dimensions WxDxH ** 800x850x1665 mm”
Operation

The circulation pump delivers up to 83 l/min at 3.7 bar and connects via G3/4 male threads, with flow adjustable by a manual bypass valve on the rear.

Informs piping design, hose/connection selection, and how to balance flow to the process load.

Source
“The flow rate can be adjusted via the manual bypass valve on the backside of the chiller.”
Operation

The tank holds between 160 l and 175 l of heat transfer fluid, and the recommended fluid is a 50:50 water–monoethylene glycol mixture.

Determines fluid volume to procure and freeze protection for operation down to -20 °C.

Source
“Recommended thermofluid: Water - Monoethylene Glycol 50:50”
Safety

The refrigerant is R-449A (A1, H280) with 4.5 kg charge and 6.29 t CO2 equivalent, which carries a pressurized-gas hazard classification.

Affects refrigerant handling, leak detection, and F-gas/CO2-equivalent reporting obligations.

Source
“Refrigerant (ASHRAE, GHS) R-449A (A1, H280)”
Cumplimiento

The unit is EMC Class A, Group 1, so it is intended for industrial environments and may cause interference in residential areas.

Determines whether the installation location is appropriate and what EMC mitigation may be required.

Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Installation

Voltage tolerance is ±5% only with a simultaneous frequency tolerance of ±2%, and certain combinations (e.g., -5% voltage with +2% frequency) are not allowed.

Grid quality at the site must be checked; out-of-tolerance supply can damage the compressor and electronics.

Source
“-5% voltage and +2% frequency -> not allowed!”
Safety

The unit has IP20 protection, so it must be installed in a dry indoor location protected from water ingress.

Prevents specifying the chiller for washdown or outdoor areas where moisture could enter the enclosure.

Source
“Degree of Protection IP20”
Operation

The chiller produces a sound pressure level of 70 dB(A) ±4 dB(A), which must be considered for operator exposure and room acoustics.

Affects placement relative to workstations and possible need for hearing protection or acoustic treatment.

Source
“sound pressure level +/- 4 dB(A) 70 dB(A)”