Product introduction and key features of KISS 118A
Specp. 2
Technical data according to DIN 12876
p. 1table, 36 rows
Technical data according to DIN 12876 — KISS 118A (Order-No.: 2053.0003.98)
Parameter
Value
Operating temperature range
25...100 °C
with water cooling
20...100 °C
with refrigerator
15...100 °C
Temperature stability at 70°C
0,05 K
temperature set point / display
digital
Absolute accuracy
setup for calibration
Internal temperature sensor
Pt100
Interface digital
USB (Device), RS232
Interface
Alarm message
optic, acoustic
Safety classification
III / FL
Heating power at 240V
2,1 kW
Heating power at 230V
2 kW
Heating power at 220V
1,8 kW
Heating power at 208V
1,6 kW
Heating power at 200V
1,5 kW
max. delivery
14 l/min
max. delivery pressure
0,25 bar
max. delivery (suction)
10,5 l/min
max. delivery pressure (suction)
0,17 bar
Pump connenction (optional)
M16x1 male
Bath volume
18 l
min. filling capacity
11 l
Height of bath opening
165 mm
Width bath opening WxD/ bath depth
275x321/ 150 mm
Overall dimensions WxDxH **
333x520x335 mm
Net weight
8 kg
Power supply requirement
200-240V 1~/2~ 50/60Hz
max. current
10 A
min. Fuse
10A
max. Fuse
16A
Pressure equipment category
N/A
Degree of Protection
IP20
min. ambient temperature
5 °C
max. ambient temperature
40 °C
from Serial-No.
393787
** Overall dimensions WxDxH; значения приведены дословно, включая опечатку «Pump connenction».
Engineering Facts
14 facts · every claim linked to a page
Selection
The usable temperature range depends on the cooling method: 25–100 °C without cooling, 20–100 °C with water cooling, and 15–100 °C with a refrigerator.
Determines whether the bath can reach the required low temperature with the available cooling utility.
Source
“Operating temperature range 25...100 °C with water cooling 20...100 °C with refrigerator 15...100 °C”
Performance
Temperature stability at 70 °C is 0.05 K, which sets the achievable control accuracy for sensitive processes.
Needed to judge whether the circulator meets process tolerance requirements.
Source
“Temperature stability at 70°C 0,05 K”
Installation
The unit requires a 200–240 V single- or two-phase supply at 50/60 Hz with a maximum current of 10 A.
Defines the electrical supply and breaker sizing for installation.
Source
“Power supply requirement 200-240V 1~/2~ 50/60Hz max. current 10 A”
Safety
External branch circuit protection according to UL 489 is required.
Mandatory for compliant electrical installation in UL jurisdictions.
Source
“External branch circuit protection according to UL 489 required.”
Installation
Voltage and frequency tolerances are ±10% only if they do not run in opposite directions; for example, -10% voltage with +3% frequency is not allowed.
Prevents installation on a supply whose combined voltage/frequency deviation exceeds the allowed envelope.
Source
“Voltage + / - 10%, as long as the frequency tolerance does not run in the opposite direction.”
Operation
The plastic pump is not resistant to acetone or polyglycols, so these media must be avoided or a more resistant plastic option specified.
Material compatibility must be checked before using the bath with these fluids.
Source
“The plastic pump is not resistant against acetone and polyglycols (depending on the manufacturer).”
Operation
For freeze protection, water should be mixed with glysantine or ethylene glycol rather than relying on pure water.
Affects fluid selection and low-temperature operation to avoid freezing damage.
Source
“It is recommended that water is mixed with either glysantine or ethylene glycol for freeze protection.”
Installation
Single- and two-phase devices are delivered with a power cable and country-specific plug, which must be specified at ordering.
Ensures the correct plug is supplied for the installation site.
Source
“Single / two-phase devices (100V to 240V) --> with power cable and country-specific plug (please specify when ordering)”
Installation
Three-phase devices below 63 A are supplied with cable but without a plug.
The installer must provide and fit the plug for three-phase units.
Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Installation
Three-phase devices above 63 A are supplied without cable and without plug.
Site wiring must be planned and provided by the installer for high-current units.
Source
“Three-phase devices with current consumption greater than 63A --> without cable, without plug”
Selection
An optional Pt100 sensor connection can display (not control) process temperature and is only available factory fitted.
Must be ordered with the unit if external temperature monitoring is needed; it cannot be retrofitted in the field.
Source
“Option: Pt100 sensor connection #10688 to display (not control) e.g. of the process temperature (only available factory fitted, additional charge).”
Safety
The unit has adjustable overtemperature protection according to DIN 12876 and safety class III/FL.
Confirms the safety concept required for unattended or flammable-liquid operation.
Source
“With adjustable overtemperature protection according to DIN 12876.”
Operation
The autostart function restarts the unit after a power failure, which may be undesirable for some processes.
Engineers must decide whether automatic restart after power loss is acceptable for their process.
Source
“Autostart function for power failure”
Installation
The unit is rated IP20 and must be operated between 5 °C and 40 °C ambient temperature.
Sets environmental limits and ingress protection for the installation location.
Source
“Degree of Protection IP20 min. ambient temperature 5 °C max. ambient temperature 40 °C”