Product introduction and key features of KISS 208B
Specp. 2
Technical data according to DIN 12876
p. 1table, 36 rows
KISS 208B Technical data according to DIN 12876
Parameter
Value
Operating temperature range
25...200 °C
with water cooling
20...200 °C
with refrigerator
-30...200 °C
Temperature stability at 70°C
0,05 K
temperature set point / display
digital
Absolute accuracy
setup for calibration
Internal temperature sensor
Pt100
Interface digital
USB (Device), RS232
Interface
Alarm message
optic, acoustic
Safety classification
III / FL
Heating power at 240V
2,1 kW
Heating power at 230V
2 kW
Heating power at 220V
1,8 kW
Heating power at 208V
1,6 kW
Heating power at 200V
1,5 kW
max. delivery
14 l/min
max. delivery pressure
0,25 bar
max. delivery (suction)
10,5 l/min
max. delivery pressure (suction)
0,17 bar
Pump connenction (optional)
M16x1 male
Bath volume
7,5 l
min. filling capacity
4 l
Height of bath opening
205 mm
Width bath opening WxD/ bath depth
230x127/ 150 mm
Overall dimensions WxDxH **
290x350x375 mm
Net weight
10 kg
Power supply requirement
200-240V 1~/2~ 50/60Hz
max. current
10 A
min. Fuse
10A
max. Fuse
16A
Pressure equipment category
N/A
Degree of Protection
IP20
min. ambient temperature
5 °C
max. ambient temperature
40 °C
from Serial-No.
398599
** Technical details and dimensions are subject to change. No liability is accepted for errors or omissions. Illustrations can deviate from the original.
Engineering Facts
14 facts · every claim linked to a page
Selection
The operating temperature range depends on the cooling method, so the required range must be matched to water cooling or a refrigerator.
Determines whether the bath can reach the process temperatures without additional cooling equipment.
Source
“Operating temperature range 25...200 °C with water cooling 20...200 °C with refrigerator -30...200 °C”
Installation
The unit requires a 200-240 V single- or two-phase supply and a branch circuit rated between 10 A and 16 A.
Electrical supply and protection must be sized accordingly before installation.
Source
“Power supply requirement 200-240V 1~/2~ 50/60Hz max. current 10 A min. Fuse 10A max. Fuse 16A”
Installation
An external branch circuit protection according to UL 489 is required for this equipment.
The installer must provide suitable branch circuit protection to meet the compliance requirement.
Source
“External branch circuit protection according to UL 489 required.”
Installation
Voltage and frequency tolerances are limited: a -10% voltage deviation combined with +3% frequency is not allowed.
Supply quality must be checked because certain combined tolerances can damage or misoperate the unit.
Source
“Example: -10% voltage and +3% frequency -> not allowed !”
Operation
The plastic pump is not resistant to acetone or polyglycols, so these media must be avoided or a more resistant plastic must be specified.
Prevents chemical attack and premature pump failure when selecting the heat transfer fluid.
Source
“The plastic pump is not resistant against acetone and polyglycols (depending on the manufacturer).”
Operation
For freeze protection, water should be mixed with glysantine or ethylene glycol rather than using pure water.
Avoids freezing-related damage and ensures reliable operation at low temperatures.
Source
“It is recommended that water is mixed with either glysantine or ethylene glycol for freeze protection.”
Safety
The unit has adjustable overtemperature protection according to DIN 12876 and safety class III/FL.
Confirms the safety concept and the need to set the overtemperature limit correctly during commissioning.
Source
“With adjustable overtemperature protection according to DIN 12876.”
Operation
The autostart function restarts the unit after a power failure, which may be undesirable for some processes.
Engineers must decide whether automatic restart after power loss is acceptable for their application.
Source
“Autostart function for power failure”
Installation
The ambient temperature during operation must remain between 5 °C and 40 °C.
Site conditions must be checked to avoid derating or malfunction outside this range.
Source
“min. ambient temperature 5 °C max. ambient temperature 40 °C”
Installation
The unit has IP20 protection, so it must be installed in a dry location protected from water ingress.
Prevents electrical hazards and failures in wet or dusty environments.
Source
“Degree of Protection IP20”
Selection
The bath has a 7.5 l volume with a minimum filling capacity of 4 l.
Determines the required fluid volume and whether the bath is large enough for the application.
Source
“Bath volume 7,5 l min. filling capacity 4 l”
Performance
The pressure pump delivers up to 14 l/min at 0.25 bar, while the suction pump delivers up to 10.5 l/min at 0.17 bar.
Used to verify that the external circuit flow and pressure requirements are met.