KISS K15 Immersion Circulator

Contenido

  1. Overview p. 1
    Product introduction and main features of KISS K15
  2. Spec pp. 1-2
    Technical data and performance specifications per DIN 12876
  3. p. 2 table, 6 rows
  4. p. 1 table, 36 rows

Technical data according to DIN 12876

ParameterValue
max. Fuse16A
Pressure equipment categoryArt. 4.3 PED
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:S393793

Technical data according to DIN 12876

ParameterValue
Operating temperature range-20...200 °C
Temperature stability at 70°C0,05 K
temperature set point / displaydigital
Absolute accuracysetup for calibration
Internal temperature sensorPt100
Interface digitalUSB (Device), RS232
InterfaceAlarm message
Alarm messageoptic, acoustic
Safety classificationIII / FL
Heating power at 240V2,1 kW
Heating power at 230V2 kW
Heating power at 220V1,8 kW
Cooling power at 20°C0,25 kW
Cooling power at 0°C0,2 kW
Cooling power at -10°C0,12 kW
Cooling power at -20°C0,05 kW
Refrigeration machineair-cooled, natural refrigerant
Refrigerant (ASHRAE, GHS)R-290 (A3, H220)
Global Warming Potential (GWP)0,02
Refrigerant quantity0,041 kg
CO2 equivalent0,0 t
UN-numberUN 3358
Pressure / Suction pump max. delivery14 l/min
Pressure / Suction pump max. delivery pressure0,25 bar
Pressure / Suction pump max. delivery (suction)10,5 l/min
Pressure / Suction pump max. delivery pressure (suction)0,17 bar
Bath volume15 l
Width bath opening WxD290x152 mm
Bath depth200 mm
Height of bath opening265 mm
Overall dimensions WxDxH **350x560x430 mm
Net weight28 kg
Power supply requirement220-240V 1~/2~ 50/60Hz
max. current immersion thermostat10 A
max. current refrigerated bath1,5 A
min. Fuse10A

** Overall dimensions WxDxH

Engineering Facts

14 facts · every claim linked to a page
Selection

The unit is rated for a working temperature range of -20 to 200 °C, so it can cover both sub-ambient and high-temperature bath duties.

Defines whether the bath can handle the required process temperatures before selection.

Source
“Operating temperature range -20...200 °C”
Performance

Cooling capacity falls from 0.25 kW at 20 °C to only 0.05 kW at -20 °C, so low-temperature heat loads must be sized against the reduced capacity.

Prevents oversizing the cooling load at the cold end of the operating range.

Source
“at -20°C 0,05 kW”
Safety

The refrigerant is R-290 (A3, H220), a flammable natural refrigerant, which affects room ventilation and ignition-source restrictions.

Flammable refrigerant classification drives installation location and safety assessment.

Source
“R-290 (A3, H220)”
Installation

The power supply requirement is 220-240 V single or two-phase at 50/60 Hz, so the site must provide the correct mains configuration.

Mismatch with site power will prevent commissioning or require a transformer.

Source
“220-240V 1~/2~ 50/60Hz”
Installation

The minimum fuse rating is 10 A and the maximum current of the refrigerated bath is 1.5 A, so the branch circuit must be protected at no less than 10 A.

Ensures the electrical supply and protection device are correctly sized.

Source
“min. Fuse 10A”
Installation

The unit is rated IP20, so it must be installed in a dry indoor location protected from water ingress.

IP20 limits where the bath can be placed and rules out washdown or wet environments.

Source
“Degree of Protection IP20”
Operation

The allowable ambient temperature range is 5 to 40 °C, and cooling capacity may drop if ambient temperature rises.

High ambient temperature reduces cooling performance and must be considered in room layout.

Source
“max. ambient temperature 40 °C”
Operation

The plastic pump is not resistant to acetone or polyglycols, so these fluids must be avoided or a more resistant plastic pump specified.

Prevents chemical attack and premature pump failure from incompatible bath fluids.

Source
“The plastic pump is not resistant against acetone and polyglycols”
Operation

For freeze protection the manufacturer recommends mixing water with glysantine or ethylene glycol rather than using pure water.

Avoids freezing and damage when operating near or below 0 °C.

Source
“It is recommended that water is mixed with either glysantine or ethylene glycol for freeze protection.”
Installation

Voltage tolerance is ±5% only with a simultaneous frequency tolerance of ±2%, and the combination of -5% voltage with +2% frequency is not allowed.

Defines the acceptable mains quality for reliable operation and warranty compliance.

Source
“-5% voltage and +2% frequency -> not allowed!”
Cumplimiento

The equipment is classified Class A, Group 1 for electromagnetic disturbance to EN55011, so it is intended for industrial rather than residential environments.

Class A equipment may cause interference in residential areas and affects EMC planning.

Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Installation

Single and two-phase devices are supplied with a power cable and country-specific plug, which must be specified at ordering.

The correct plug type must be stated on the order to match the site socket.

Source
“Single / two-phase devices (100V to 240V) --> with power cable and country-specific plug (please specify when ordering)”
Maintenance

The optional Pt100 sensor connection is only available factory fitted, so it must be decided at the time of purchase.

Retrofitting is not possible, so the option must be included in the initial order.

Source
“only available factory fitted, additional charge”
Safety

The unit has adjustable overtemperature protection according to DIN 12876 and safety class III/FL, providing fail-safe protection.

Confirms the safety concept required for unattended or high-temperature operation.

Source
“With adjustable overtemperature protection according to DIN 12876.”