Selection
The unit covers a working temperature range from -25 to 200 °C, so it suits both sub-ambient and high-temperature bath duties.
Defines whether the bath can handle your process setpoints without a second unit.
Source
“Operating temperature range -25...200 °C”
Performance
Cooling capacity falls from 0.2 kW at 20 °C to 0.05 kW at -20 °C, so low-temperature heat loads must be sized against the reduced capacity.
Prevents undersizing the refrigeration when operating near the cold end of the range.
Source
“Cooling power at 20°C 0,2 kW at 0°C 0,15 kW at -10°C 0,1 kW at -20°C 0,05 kW”
Safety
The refrigerant is R-290 (A3, H220), a flammable natural refrigerant, which affects room ventilation and ignition-source planning.
Flammable refrigerant classification drives installation location and safety review.
Source
“Refrigerant (ASHRAE, GHS) R-290 (A3, H220)”
Installation
The refrigerating unit is air-cooled, so adequate clearance and cool intake air are required at the installation site.
Air-cooled condensers lose capacity in hot or confined spaces; space requirements must be respected.
Source
“Refrigeration machine air-cooled, natural refrigerant”
Installation
The electrical supply must be 208-240 V single or two-phase at 50/60 Hz with a fuse between 10 A and 16 A.
Determines the site power feed and protective device rating before installation.
Source
“Power supply requirement 208-240V 1~/2~ 50/60Hz”
Installation
Voltage and frequency tolerances are ±5% and ±2% respectively, but the combination of -5% voltage with +2% frequency is not allowed.
Avoids operating the unit on a supply outside the permitted combined tolerance envelope.
Source
“-5% voltage and +2% frequency -> not allowed!”
Operation
The unit is rated IP20 and must be operated between 5 °C and 40 °C ambient, so it is for indoor dry locations only.
Prevents installation in wet or hot environments where protection and cooling would be inadequate.
Source
“Degree of Protection IP20 min. ambient temperature 5 °C max. ambient temperature 40 °C”
Operation
Cooling capacity figures are valid at 20 °C room temperature and may drop if ambient temperature rises.
Requires derating of published cooling capacity for warm plant rooms.
Source
“Output data valid for: Room temperature 20°C. If the ambient temperature rises, the cooling capacity may drop.”
Selection
The plastic pump is not resistant to acetone or polyglycols, so these media require an alternative wetted-material option.
Prevents chemical attack on the pump when choosing the heat transfer fluid.
Source
“The plastic pump is not resistant against acetone and polyglycols (depending on the manufacturer).”
Operation
For freeze protection the manufacturer recommends water mixed with glysantine or ethylene glycol rather than pure water.
Guides correct heat transfer fluid preparation for sub-zero operation.
Source
“It is recommended that water is mixed with either glysantine or ethylene glycol for freeze protection.”
Installation
Single and two-phase devices are delivered with a power cable and country-specific plug, which must be specified at ordering.
Ensures the correct mains plug is supplied for the installation country.
Source
“Single / two-phase devices (100V to 240V) --> with power cable and country-specific plug (please specify when ordering)”
Cumplimiento
The equipment is classified Class A, Group 1 for electromagnetic disturbance to EN55011.
Class A equipment may require EMC precautions in residential or sensitive environments.
Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Safety
The unit has adjustable overtemperature protection and safety class III/FL according to DIN 12876.
Confirms the protective cut-out required for unattended flammable-liquid operation.
Source
“With adjustable overtemperature protection according to DIN 12876.”
Maintenance
Wetted parts are stainless steel or plastics, which limits compatible fluids and cleaning agents.
Informs fluid selection and cleaning procedures to avoid corrosion or attack.
Source
“Wetted parts made from stainless steel or plastics.”