Product introduction and main features of the Minichiller 1200w
Specp. 2
Technical data according to DIN 12876 for the chiller
p. 2table, 15 rows
p. 1table, 24 rows
Technical data according to DIN 12876
Parameter
Value
min. cooling water differential pressure
3 bar
max. cooling water pressure
6 bar
min. filling capacity
2,8 l
Volume of expansion
2,2 l
Overall dimensions WxDxH **
280x490x424 mm
Net weight
38 kg
Power supply requirement
208-240V 1~/2~ 50/60Hz
max. current
5 A
min. Fuse
10A
max. Fuse
16A
Pressure equipment category
Art. 4.3 PED
Degree of Protection
IP20
min. ambient temperature
5 °C
max. ambient temperature
40 °C
from Serial-No.:
S571408
** Please respect space requirements. See operating conditions at www.huber-online.com
Minichiller 1200w — Technical data according to DIN 12876
Parameter
Value
Order-No.
3078.0007.01
Operating temperature range
-20...40 (80)*** °C
temperature set point / display
5,7" colour Touchscreen
Internal temperature sensor
Pt100
Sensor external connection
Pt100
Temperature stability at -10°C
0,5 K
Interface digital
Ethernet, USB (Host u. Device), RS232
Safety classification
I / NFL
Cooling power at ambient temperature +20°C at 15°C
1,2 kW
Cooling power at ambient temperature +20°C at 0°C
0,9 kW
Cooling power at ambient temperature +20°C at -10°C
0,7 kW
Cooling power at ambient temperature +20°C at -20°C
0,35 kW
Refrigeration machine
water-cooled, natural refrigerant
Refrigerant (ASHRAE, GHS)
R-290 (A3, H220)
Global Warming Potential (GWP)
0,02
Refrigerant quantity
0,07 kg
CO2 equivalent
0,0 t
Circulation pump
Pressure- and suction pump
max. delivery
24 l/min
max. delivery pressure
0,7 bar
max. delivery (suction)
18 l/min
max. delivery pressure (suction)
0,4 bar
Pump connection
M16x1 male
Cooling water connection
G1/2 male
*** — сноска в исходном тексте не раскрыта. Дополнительно на странице указаны данные для whisper mode: switch for whisper mode; Reduced sound pressure level: 51 dB(A): +/- 4 dB(A); Pump data for whisper mode: Delivery rate: 14 l/min; Delivery pressure: 0.2 bar; Delivery rate (suction): 11 l/min; Delivery pressure (suction): 0.18 bar.
Engineering Facts
12 facts · every claim linked to a page
Installation
The chiller requires cooling water supply pressure between 3 bar minimum differential and 6 bar maximum, so the facility water system must be sized accordingly.
Insufficient or excessive cooling water pressure will prevent proper operation or damage the unit.
Source
“min. cooling water differential pressure 3 bar max. cooling water pressure 6 bar”
Installation
Cooling water inlet temperature must be kept between 5°C and 20°C, and pipes must be insulated if condensation occurs.
Out-of-range cooling water temperature reduces cooling capacity and causes condensation damage.
Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C. When cooling water temperatures fall below the condensation point, it is essential to insulate the cooling water pipes in the unit.”
Selection
The unit uses flammable R-290 refrigerant (A3, H220), which affects safety classification and installation requirements.
Flammable refrigerant requires appropriate ventilation and safety measures at the installation site.
Source
“Refrigerant (ASHRAE, GHS) R-290 (A3, H220)”
Selection
The plastic pump is not resistant to acetone or polyglycols, so these media must be avoided or a more resistant pump ordered.
Using incompatible media will damage the pump and void reliable operation.
Source
“The plastic pump is not resistant against acetone and polyglycols (depending on the manufacturer).”
Operation
For freeze protection, water should be mixed with glysantine or ethylene glycol rather than relying on pure water.
Prevents freezing and damage in the cooling circuit at low temperatures.
Source
“It is recommended that water is mixed with either glysantine or ethylene glycol for freeze protection.”
Installation
Voltage tolerance is ±5% only with simultaneous frequency tolerance of ±2%, and certain combinations are not allowed.
Ensures the unit is not operated outside its electrical design envelope, avoiding malfunction.
Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Installation
The unit requires a minimum fuse of 10A and maximum of 16A on a 208-240V single/two-phase supply.
Correct fuse sizing protects the equipment and ensures compliance with electrical safety.
Source
“min. Fuse 10A max. Fuse 16A”
Operation
Whisper mode reduces sound pressure to 51 dB(A) but also reduces pump delivery to 14 l/min and 0.2 bar.
Engineers must account for reduced cooling performance when using quiet operation.