Installation
The chiller requires cooling water supply pressure between 3 and 6 bar, so the building water system must be verified before installation.
Insufficient differential pressure will prevent proper cooling and may trip the unit.
Source
“min. cooling water differential pressure 3 bar”
Installation
Cooling water inlet temperature must be kept between 5°C and 20°C, and pipes must be insulated if condensation occurs.
Out-of-range cooling water temperature reduces cooling capacity and risks condensation damage.
Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Operation
Cooling capacity drops as cooling water temperature rises, and the required cooling water flow rate may increase.
Sizing the cooling water circuit must account for reduced performance at higher water temperatures.
Source
“As the cooling water temperature increases, drop in the cooling power should be expected, and also an increased cooling water flow rate possible.”
Safety
The refrigerant is R-290 (propane, A3 flammable), so the installation area must be assessed for flammable refrigerant handling.
Flammable refrigerant classification affects site safety requirements and ventilation.
Source
“R-290 (A3, H220)”
Safety
The plastic pump is not resistant to acetone or polyglycols, so these fluids must not be used as heat transfer media.
Using incompatible fluids will damage the pump and void reliable operation.
Source
“The plastic pump is not resistant against acetone and polyglycols (depending on the manufacturer).”
Operation
For freeze protection, water should be mixed with glysantine or ethylene glycol rather than relying on pure water.
Prevents freezing and damage when operating below 0°C.
Source
“It is recommended that water is mixed with either glysantine or ethylene glycol for freeze protection.”
Installation
The unit has IP20 protection, so it must be installed in a dry indoor location protected from water ingress.
IP20 is not water-resistant; outdoor or washdown environments are unsuitable.
Source
“Degree of Protection IP20”
Installation
Ambient temperature must stay between 5°C and 40°C for reliable operation.
Exceeding ambient limits reduces cooling performance and may cause alarms.
Source
“max. ambient temperature 40 °C”
Cumplimiento
Voltage and frequency tolerances are ±5% and ±2% respectively, but not simultaneously in opposite directions.
Ensures the power supply quality is adequate and avoids invalid combinations.
Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Installation
The unit is classified as EN55011 Class A Group 1, so it is intended for industrial electromagnetic environments.
Class A equipment may cause interference in residential areas; installation location matters.
Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Selection
The return line can accept up to 80°C, which is higher than the operating range and must be considered for process heat loads.
Defines the maximum allowable return temperature from the process.
Source
“Permissible temperature in return line 80 °C”
Operation
Whisper mode reduces sound pressure to 51 dB(A) but also reduces pump delivery to 14 l/min and 0.2 bar.
Noise-sensitive installations may need whisper mode, but pump performance drops.
Source
“Reduced sound pressure level: 51 dB(A): +/- 4 dB(A)”
Maintenance
The cooling water circuit contains copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible cooling water should be used.
Incompatible water chemistry can corrode or foul the cooling circuit.
Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.”