Product introduction and general description of the Petite Fleur-eo unit
Specp. 2
Technical data and performance parameters according to DIN 12876
p. 1table, 28 rows
Technical data according to DIN 12876
Parameter
Petite Fleur-eo
Order-No.
1030.0055.01
Operating temperature range
-40...200 °C
Temperature stability at -10°C
0,01 K
temperature set point / display
5,7" colour Touchscreen
Resolution of display
0,01 K
Internal temperature sensor
Pt100
Sensor external connection
Pt100
Interface digital
Ethernet, USB (Host u. Device), RS232
digital input
ECS ONE
digital output
POKO ONE
Alarm message
optic, acoustic, relay
Safety classification
III / FL
Heating power convertible
1,6 - 2 kW
Cooling power with Thermooil at 200°C
0,48 kW
Cooling power with Thermooil at 150°C
0,48 kW
Cooling power with Thermooil at 100°C
0,48 kW
Cooling power with Thermooil at 50°C
0,48 kW
Cooling power with Thermooil at 20°C
0,48 kW
Cooling power with Ethanol at 0°C
0,45 kW
Cooling power with Ethanol at -20°C
0,27 kW
Cooling power with Ethanol at -30°C
0,16 kW
Cooling power with Ethanol at -40°C
0,04 kW
Refrigeration machine
air-cooled, natural refrigerant
Refrigerant (ASHRAE, GHS)
R-290 (A3, H220)
Global Warming Potential (GWP)
0,02
Refrigerant quantity
0,08 kg
CO2-Äquvivalent
0,0 t
UN-number
UN 3358
Engineering Facts
14 facts · every claim linked to a page
Installation
The unit is designed only for external open operation and has no expansion tank, so the external application must accommodate fluid filling and temperature-dependent volume changes.
Determines whether an external reservoir/expansion volume must be engineered into the loop before installation.
Source
“The machine is developed exclusively for external open operation, and does not possess an expansion tank.”
Safety
The refrigerant is R-290 (A3, H220), a flammable natural refrigerant, so installation must respect flammable-refrigerant zoning and ignition-source control.
Affects site selection, ventilation, and electrical/ignition-source precautions during installation.
Source
“R-290 (A3, H220)”
Performance
Cooling capacity is only 0.04 kW at -40 °C with ethanol, so the -40 °C lower limit is practical only for low heat loads.
Prevents over-sizing the cooling duty expectation at the extreme low end of the operating range.
Source
“at -40°C
0,04 kW”
Performance
Cooling capacity is rated at 20 °C room temperature and may drop if ambient rises, so selection must account for the actual site ambient.
Cooling performance must be derated for warm rooms; otherwise the chosen unit may be undersized.
Source
“Output data valid for: Room temperature 20°C. If the ambient temperature rises, the cooling capacity may drop.”
Installation
The permissible ambient temperature range is 5 to 40 °C, so the installation location must stay within this window.
Defines the acceptable room conditions for reliable operation and warranty compliance.
Source
“min. ambient temperature
5 °C
max. ambient temperature
40 °C”
Installation
The pump connection is M16x1 male, so mating fittings and hoses must match this thread.
Ensures correct plumbing adapters are ordered and avoids leaks or mismatched connections.
Source
“Pump connection
M16x1 male”
Installation
The minimum filling capacity is 2 l, so the external loop plus application must provide at least this fluid volume.
Prevents commissioning with insufficient thermal fluid, which would impair circulation and temperature control.
Source
“min. filling capacity
2 l”
Performance
The pump delivers up to 25 l/min at 0.9 bar, so external circuit pressure drop must be checked against this capability.
Confirms whether the external application's flow resistance is compatible with the circulator.
The maximum permissible kinematic viscosity is 50 mm²/s, so the selected thermofluid must not exceed this at the lowest operating temperature.
Prevents pump overload and poor circulation when choosing a fluid for low-temperature operation.
Source
“max. permissible kin. viscosity
50 mm²/s”
Cumplimiento
Voltage tolerance is ±5% only with simultaneous frequency tolerance of ±2%, and certain combinations are not allowed, so supply quality must be verified.
Avoids invalid supply conditions that could damage the unit or void compliance.
Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Installation
Single/two-phase devices are supplied with a country-specific plug that must be specified at order, so the correct power connection must be declared.
Prevents delivery with the wrong plug and avoids rework at commissioning.
Source
“with power cable and country-specific plug (please specify when ordering)”
Cumplimiento
The equipment is EMC Class A Group 1, so it is intended for industrial environments and may cause interference in residential areas.
Affects site selection and EMC planning when installing near sensitive equipment.
Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Safety
The unit has adjustable overtemperature protection according to DIN 12876, so the protection setpoint must be configured for the application.
Requires commissioning step to set safe overtemperature limit for the fluid and application.
Source
“With adjustable overtemperature protection according to DIN 12876.”
Operation
The controller supports selectable internal/process temperature control mode, so the sensor strategy must be chosen for the application.
Determines whether the external Pt100 is used for process control and affects control tuning.
Source
“selectable temperature control mode (Internal/Process)”