The unit covers a working temperature range from -45 to 100 °C, so it must be matched to the process temperature window.
Defines whether the cooler can reach the required process temperature at all.
Source
“Operating temperature range -45...100 °C”
| Parameter | Value |
|---|---|
| Order-No. | 3003.0043.00 |
| Model | TC45 |
| Operating temperature range | -45...100 °C |
| Cooling power at 0°C | 0,24 kW |
| Cooling power at -20°C | 0,18 kW |
| Cooling power at -30°C | 0,1 kW |
| Cooling power at -40°C | 0,05 kW |
| Safety classification | I / NFL |
| Refrigeration machine | air-cooled, natural |
| refrigerant | |
| Refrigerant (ASHRAE, GHS) | R-290 (A3, H220) |
| Global Warming Potential (GWP) | 0,02 |
| Refrigerant quantity | 0,041 kg |
| CO2 equivalent | 0,0 t |
| UN-number | UN 3358 |
| Nominal diameter probe | 48 mm |
| Length of probe | 105 mm |
| Length flexible connection | 1050 mm |
| Overall dimensions WxDxH ** | 190x295x360 mm |
| Net weight | 16 kg |
| Power supply requirement | 208-240V 1~/2~ 50/60Hz |
| max. current | 1,4 A |
| min. Fuse | 10A |
| max. Fuse | 16A |
| Pressure equipment category | Art. 4.3 PED |
| min. ambient temperature | 5 °C |
| max. ambient temperature | 40 °C |
| from Serial-No. | S394777 |
** Please respect space requirements. See operating conditions at www.huber-online.com
The unit covers a working temperature range from -45 to 100 °C, so it must be matched to the process temperature window.
Defines whether the cooler can reach the required process temperature at all.
“Operating temperature range -45...100 °C”
Cooling capacity drops sharply at low temperature (0.24 kW at 0 °C down to 0.05 kW at -40 °C), so the load must be checked at the coldest setpoint.
Sizing must use the capacity at the actual operating temperature, not the nominal figure.
“at 0°C 0,24 kW at -20°C 0,18 kW at -30°C 0,1 kW at -40°C 0,05 kW”
Rated output data are valid only at 20 °C room temperature and fall as ambient rises.
Installation room ventilation and ambient conditions directly affect achievable cooling.
“Output data valid for: Room temperature 20°C. If the ambient temperature rises, the cooling capacity may drop.”
The unit may only be operated between 5 °C and 40 °C ambient temperature.
Site conditions must be verified before installation to avoid derating or shutdown.
“min. ambient temperature 5 °C max. ambient temperature 40 °C”
The refrigerant is R-290 (propane, A3 flammable, H220), so ignition sources and ventilation in the installation area must be controlled.
Flammable refrigerant classification drives room ventilation and ignition-source restrictions.
“Refrigerant (ASHRAE, GHS) R-290 (A3, H220)”
The charge is only 0.041 kg R-290 with negligible CO2 equivalent, which keeps refrigerant-related obligations small.
Small charge simplifies leak-check and F-gas type record-keeping requirements.
“Refrigerant quantity 0,041 kg CO2 equivalent 0,0 t”
Voltage tolerance is ±5% only with a simultaneous frequency tolerance of ±2%, and the combination -5% voltage with +2% frequency is not allowed.
Supply quality at the site must be checked; some voltage/frequency combinations are prohibited.
“-5% voltage and +2% frequency -> not allowed!”
Single/two-phase devices are delivered with a power cable and country-specific plug that must be specified at ordering.
The correct plug must be ordered up front or the unit cannot be connected on site.
“Single / two-phase devices (100V to 240V) --> with power cable and country-specific plug (please specify when ordering)”
The device is EMC Class A Group 1, i.e. industrial environment equipment that may cause interference in residential areas.
Determines suitability for industrial vs. residential/lab-sensitive EMC environments.
“Classification (disturbance) to EN55011: Class A, Group 1”
Space requirements around the unit must be respected per the operating conditions.
Air-cooled condenser needs clearance; ignoring it reduces capacity and can trip the unit.
“** Please respect space requirements. See operating conditions at www.huber-online.com”
The probe is 48 mm nominal diameter and 105 mm long with a 1050 mm flexible connection, which constrains how it fits into the vessel.
Geometry must be checked against the vessel opening and routing before mounting.
“Nominal diameter probe 48 mm Length of probe 105 mm Length flexible connection 1050 mm”
The refrigeration unit runs continuously rather than cycling on/off.
Continuous operation affects energy consumption, noise and expected wear.
“The refrigeration unit works continiously.”
The unit is an immersion cooler with air-cooled refrigerating unit and stainless steel condensing probe.
Confirms the cooling principle and wetted material compatibility for the application.
“Immersion Cooler with air-cooled refrigerating unit. Housing of stainless steel with 2 handles, condensing probe of stainless steel”