TC50-F Datasheet

Contenido

  1. Overview p. 1
    Product introduction and key features of TC50-F
  2. Spec p. 1
    Technical specifications table for TC50-F
  3. p. 1 table, 26 rows

Technical data according to DIN 12876

ParameterValue
Operating temperature range-50...50 °C
Cooling power
at 0°C0,3 kW
at -20°C0,26 kW
at -30°C0,2 kW
Safety classificationI / NFL
Refrigeration machineair-cooled, natural
refrigerant
Refrigerant (ASHRAE, GHS)R-290 (A3, H220)
Global Warming Potential (GWP)0,02
Refrigerant quantity0,06 kg
CO2 equivalent0,0 t
Nominal diameter probe13 mm
Length of probe900 mm
Length flexible connection1150 mm
Overall dimensions WxDxH **260x330x415 mm
Net weight25 kg
sound pressure level +/- 4 dB(A)60 dB(A)
Power supply requirement208-240V 1~/2~ 50/60Hz
max. current3 A
min. Fuse10A
max. Fuse16A
Pressure equipment categoryArt. 4.3 PED
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.S417599

** Please respect space requirements. See operating conditions at www.huber-online.com

Engineering Facts

14 facts · every claim linked to a page
Selection

The unit's cooling capacity is rated at 0.3 kW at 0 °C, 0.26 kW at -20 °C and 0.2 kW at -30 °C, so it suits low-load immersion cooling only.

Cooling capacity at the actual bath temperature determines whether the unit can hold the process setpoint.

Source
“at 0°C 0,3 kW at -20°C 0,26 kW at -30°C 0,2 kW”
Safety

The refrigerant is R-290 (propane, A3, H220), a flammable natural refrigerant, so ignition sources and ventilation must be considered at installation.

Flammable A3 refrigerant drives site safety, ventilation and zoning requirements.

Source
“R-290 (A3, H220)”
Installation

The unit is only rated for ambient temperatures from 5 °C to 40 °C, so the installation room must stay within this band.

Ambient limits define where the air-cooled condenser can be installed and how much heat rejection space is needed.

Source
“min. ambient temperature 5 °C max. ambient temperature 40 °C”
Performance

Cooling capacity figures are valid at 20 °C room temperature and may drop as ambient temperature rises.

Rated capacity must be derated for warm rooms, otherwise the unit will be undersized.

Source
“Output data valid for: Room temperature 20°C. If the ambient temperature rises, the cooling capacity may drop.”
Installation

Voltage tolerance is ±5 % only with a simultaneous frequency tolerance of ±2 %, and the combination -5 % voltage with +2 % frequency is not allowed.

Supply quality must be checked before connection; some voltage/frequency combinations void the rating.

Source
“-5% voltage and +2% frequency -> not allowed!”
Installation

The power supply requirement is 208-240 V, 1~/2~, 50/60 Hz with max. current 3 A and a fuse between 10 A and 16 A.

Electrical supply and protection sizing must match these values for a compliant connection.

Source
“Power supply requirement 208-240V 1~/2~ 50/60Hz max. current 3 A min. Fuse 10A max. Fuse 16A”
Installation

Single/two-phase devices are delivered with a power cable and country-specific plug that must be specified at ordering.

The plug type must be stated on the order, otherwise the unit arrives with the wrong connector.

Source
“Single / two-phase devices (100V to 240V) --> with power cable and country-specific plug (please specify when ordering)”
Cumplimiento

The equipment is EMC class A, group 1, so it is intended for industrial environments and may cause interference in residential areas.

Class A equipment restricts where it can be installed and what EMC measures are needed.

Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Cumplimiento

The unit complies with US-SNAP and applicable EU laws, with the US-SNAP end-use being industrial process refrigeration.

Regulatory end-use classification affects approval and permitted application of the unit.

Source
“This equipment is compliant to US-SNAP and all applicable EU laws. The US-SNAP end-use for this equipment is the industrial process refrigeration.”
Installation

Space requirements around the unit must be respected as stated in the operating conditions.

Clearances are needed for the air-cooled condenser and for service access.

Source
“** Please respect space requirements. See operating conditions at www.huber-online.com”
Selection

The flexible condensing probe is 900 mm long with a 13 mm nominal diameter and a 1150 mm flexible connection line.

Probe and line lengths determine whether the unit reaches the vessel and how it is routed.

Source
“Nominal diameter probe 13 mm Length of probe 900 mm Length flexible connection 1150 mm”
Installation

The unit weighs 25 kg net with overall dimensions 260x330x415 mm, which sets handling and bench/frame requirements.

Weight and footprint dictate lifting, mounting and space planning.

Source
“Overall dimensions WxDxH ** 260x330x415 mm Net weight 25 kg”
Operation

The refrigeration unit runs continuously, so it should not be switched on/off as a temperature control strategy.

Continuous operation affects energy use, noise and expected compressor life.

Source
“The refrigeration unit works continiously.”
Safety

The sound pressure level is 60 dB(A) ±4 dB(A), which must be considered for personnel exposure and room acoustics.

Noise level matters for workplace limits and installation location.

Source
“sound pressure level +/- 4 dB(A) 60 dB(A)”