Unichiller 015-H Technical Data

Contenido

  1. Overview p. 1
    Product name and general description of the chiller unit
  2. Spec pp. 1-2
    Technical data and performance parameters according to DIN 12876
  3. p. 2 table, 3 rows
  4. p. 1 table, 34 rows

Technical data according to DIN 12876

ParameterValue
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:S608927

Technical data according to DIN 12876 — Unichiller 015-H (Order-No.: 3051.0044.01)

ParameterValue
Operating temperature range-20...100 °C
temperature set point / display5,7" colour Touchscreen
Internal temperature sensorPt100
Sensor external connectionPt100
Temperature stability at -10°C0,2 K
Interface digitalEthernet, USB (Host u. Device), RS232
Safety classificationIII / FL
Heating power at 240V2,1 kW
Heating power at 230V2 kW
Heating power at 220V1,8 kW
Cooling power at ambient temperature +20°C
at 15°C1,5 kW
at 0°C1 kW
at -10°C0,7 kW
at -20°C0,3 kW
Refrigeration machineair-cooled, natural refrigerant
Refrigerant (ASHRAE, GHS)R-290 (A3, H220)
Global Warming Potential (GWP)0,02
Refrigerant quantity0,17 kg
CO2 equivalent0,0 t
Circulation pump:Immersion pump
max. delivery29 l/min
max. delivery pressure1 bar
Delivery at 0,5 bar21 l/min
Pump connectionG3/4 male
min. filling capacity3,8 l
Volume of expansion1,7 l
Overall dimensions WxDxH **420x487x579 mm
Net weight55 kg
sound pressure level +/- 4 dB(A)63 dB(A)
Power supply requirement220-240V 1~/2~ 50Hz
max. current14,5 A
Fuse16 A
Degree of ProtectionIP20

** Overall dimensions WxDxH

Engineering Facts

14 facts · every claim linked to a page
Selection

The chiller is rated for an operating temperature range of -20 to 100 °C, so it must be matched to the process temperature window.

Defines whether the unit can cover the required process temperatures before selection.

Source
“Operating temperature range -20...100 °C”
Selection

Cooling capacity falls with ambient temperature, so the required heat load must be checked against the capacity curve at the actual site ambient.

Prevents undersizing when the chiller is installed in a warm room.

Source
“Output data valid for: Room temperature 20°C. If the ambient temperature rises, the cooling capacity may drop.”
Installation

The unit requires a 220-240 V single/two-phase 50 Hz supply with a 16 A fuse and draws up to 14.5 A, so the electrical supply must be sized accordingly.

Determines the electrical circuit, protection and cabling needed at the installation site.

Source
“Power supply requirement 220-240V 1~/2~ 50Hz max. current 14,5 A Fuse 16 A”
Installation

Voltage and frequency tolerances are linked, and a -5% voltage combined with +2% frequency is explicitly not allowed.

Prevents installation on a supply where combined voltage/frequency deviation exceeds the permitted envelope.

Source
“-5% voltage and +2% frequency -> not allowed!”
Installation

The chiller is air-cooled and requires respect of space requirements, so sufficient clearance for air flow must be provided.

Air-cooled units need adequate clearance to avoid reduced cooling performance or overheating.

Source
“** Please respect space requirements. See operating conditions at www.huber-online.com”
Safety

The refrigerant is R-290 (A3, H220), a flammable natural refrigerant, which affects site safety classification and handling requirements.

Flammable refrigerant classification influences installation location, ventilation and safety planning.

Source
“Refrigerant (ASHRAE, GHS) R-290 (A3, H220)”
Operation

The plastic pump is not resistant to acetone or polyglycols, so these fluids must not be used without a more resistant pump option.

Prevents chemical attack and pump failure when selecting the heat transfer fluid.

Source
“Special Case: Acetone and Polyglycol: The plastic pump is not resistant against acetone and polyglycols (depending on the manufacturer).”
Operation

For freeze protection the manufacturer recommends mixing water with glysantine or ethylene glycol rather than using pure water.

Guides correct fluid preparation to avoid freezing and damage at low temperatures.

Source
“It is recommended that water is mixed with either glysantine or ethylene glycol for freeze protection.”
Installation

Single/two-phase devices are supplied with a power cable and country-specific plug, which must be specified at ordering.

Ensures the correct plug type is ordered for the destination country and avoids rework on site.

Source
“Single / two-phase devices (100V to 240V) --> with power cable and country-specific plug (please specify when ordering)”
Cumplimiento

The equipment is classified as EN55011 Class A Group 1, so it is intended for industrial electromagnetic environments rather than residential ones.

Affects EMC planning and suitability for the installation environment.

Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Performance

The circulation pump delivers a maximum of 29 l/min at 1 bar and 21 l/min at 0.5 bar, which sets the achievable flow in the external circuit.

Used to check whether the pump can overcome the pressure drop of the external loop.

Source
“max. delivery 29 l/min max. delivery pressure 1 bar Delivery at 0,5 bar 21 l/min”
Operation

The minimum filling capacity is 3.8 l with 1.7 l expansion volume, so the external loop and fluid charge must be planned within these limits.

Determines the required heat transfer fluid volume and expansion allowance.

Source
“min. filling capacity 3,8 l Volume of expansion 1,7 l”
Installation

The pump connection is G3/4 male, so mating hoses and fittings must match this thread.

Ensures correct hose couplings are available for commissioning.

Source
“Pump connection G3/4 male”
Operation

The unit is rated IP20, so it must be protected from water ingress and is not suitable for wet or washdown environments.

Constrains the installation location with respect to moisture and dust.

Source
“Degree of Protection IP20”