Product name and general introduction to the chiller unit
Specp. 2
Technical data table according to DIN 12876 standard
p. 1table, 39 rows
Unichiller 015 OLÉ — Technical data according to DIN 12876
Parameter
Value
Order-No.
3051.0018.98
Operating temperature range
-20...40 °C
temperature set point / display
digital
Internal temperature sensor
Pt100
Resolution of display
0,1 K
Interface digital
USB (Device), RS232
Interface
Temperature stability at -10°C
0,5 K
Alarm message
optic, acoustic
Safety classification
I / NFL
Cooling power at 15°C
1,5 kW
Cooling power at 0°C
1 kW
Cooling power at -10°C
0,7 kW
Cooling power at -20°C
0,3 kW
Refrigeration machine
air-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)
R-449A (A1, H280)
Global Warming Potential (GWP)
1397
Refrigerant quantity
0,33 kg
CO2 equivalent
0,5 t
UN-number
UN 2857
Circulation pump
Immersion pump
at 0,5 bar
21 l/min
max. delivery
29 l/min
max. delivery pressure
1 bar
Pump connection
G3/4 male
min. filling capacity
3,8 l
expansion tank
1,7 l
Overall dimensions WxDxH **
420x487x579 mm
Net weight
60 kg
sound pressure level +/- 4 dB(A)
63 dB(A)
Power supply requirement
220-240V 1~ 50Hz
max. current
4,5 A
min. Fuse
10A
max. Fuse
16A
Pressure equipment category
Art. 4.3 PED
Degree of Protection
IP20
min. ambient temperature
5 °C
max. ambient temperature
40 °C
from Serial-No.
S593455
** Overall dimensions WxDxH
Engineering Facts
12 facts · every claim linked to a page
Selection
The chiller is rated for an operating temperature range of -20 to 40 °C, so process requirements outside this window cannot be met.
Defines the process temperature envelope for which the unit can be selected.
Source
“Operating temperature range
-20...40 °C”
Performance
Cooling capacity drops with falling temperature, from 1.5 kW at 15 °C to only 0.3 kW at -20 °C, so the load must be checked at the lowest required temperature.
Sizing must be based on the worst-case cooling capacity at the coldest set point.
Source
“at 15°C
1,5 kW
at 0°C
1 kW
at -10°C
0,7 kW
at -20°C
0,3 kW”
Installation
The unit requires a 220-240 V single-phase 50 Hz supply with a minimum 10 A fuse, so the electrical installation must be sized accordingly.
Determines the power circuit and protective device rating at the installation site.
Source
“Power supply requirement
220-240V 1~ 50Hz
max. current
4,5 A
min. Fuse
10A”
Installation
The chiller is air-cooled and needs ambient temperature between 5 and 40 °C, so the room must provide adequate ventilation and climate control.
Ambient conditions affect cooling performance and must be verified before siting the unit.
Source
“min. ambient temperature
5 °C
max. ambient temperature
40 °C”
Performance
Cooling capacity is specified at 20 °C room temperature and may drop if ambient temperature rises, so capacity margins should account for warmer rooms.
Prevents undersizing when the unit is installed in a warm environment.
Source
“Output data valid for: Room temperature 20°C. If the ambient temperature rises, the cooling capacity may drop.”
Cumplimiento
Voltage tolerance of ±5% is only valid together with a simultaneous frequency tolerance of ±2%, and certain combinations are not allowed.
The site power quality must be checked against these combined tolerances to avoid invalid operating conditions.
Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Operation
The plastic pump is not resistant to acetone or polyglycols, so these media must be avoided or a more resistant pump option ordered.
Media compatibility must be confirmed before commissioning to prevent pump damage.
Source
“The plastic pump is not resistant against acetone and polyglycols (depending on the manufacturer).”
Operation
For freeze protection, water should be mixed with glysantine or ethylene glycol rather than relying on pure water.
Guides the heat transfer fluid selection to avoid freezing at low temperatures.
Source
“It is recommended that water is mixed with either glysantine or ethylene glycol for freeze protection.”
Installation
Single- and two-phase devices are delivered with a power cable and country-specific plug, which must be specified at ordering.
The correct plug type must be stated on the order to match the site's power outlet.
Source
“Single / two-phase devices (100V to 240V) --> with power cable and country-specific plug (please specify when ordering)”
Safety
The refrigerant R-449A is classified A1 with GHS hazard H280, so refrigerant handling and ventilation precautions apply.
Safety classification and hazard statement inform site risk assessment and service procedures.
Source
“R-449A (A1, H280)”
Installation
The unit has IP20 protection, so it must be installed in a dry indoor location protected from water and dust ingress.
Restricts the installation environment and prevents placing the unit where moisture or dust is present.
Source
“Degree of Protection
IP20”
Maintenance
The minimum filling capacity is 3.8 l and the expansion tank is 1.7 l, so the external circuit volume must be planned within these limits.
Determines the required heat transfer fluid volume and expansion allowance for the loop.
Source
“min. filling capacity
3,8 l
expansion tank
1,7 l”