Unichiller 020T Technical Data

Contenido

  1. Overview p. 1
    Introduces the Unichiller 020T chiller unit
  2. Spec pp. 1-2
    Performance, temperature, and dimensional data per DIN 12876
  3. p. 2 table, 2 rows
  4. p. 1 table, 34 rows

Technical data according to DIN 12876

ParameterValue
max. ambient temperature40 °C
from Serial-No.:S514125

Technical data according to DIN 12876 — Unichiller 020T (Order-No.: 3024.0057.01)

ParameterValue
Operating temperature range-20...40 °C
temperature set point / display5,7" colour Touchscreen
Internal temperature sensorPt100
Sensor external connectionPt100
Interface digitalEthernet, USB (Host u. Device), RS232
Temperature stability at -10°C0,5 K
Safety classificationI / NFL
Cooling power at 15°C2 kW
Cooling power at 0°C2 kW
Cooling power at -10°C1,5 kW
Cooling power at -20°C0,8 kW
Refrigeration machineair-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)R-449A (A1, H280)
Global Warming Potential (GWP)1397
Refrigerant quantity1,2 kg
CO2 equivalent1,7 t
Circulation pump:B
max. delivery25 l/min
max. delivery pressure3 bar
Delivery at 0,5 bar20 l/min
Delivery at 1,0 bar17 l/min
Delivery at 2,0 bar10 l/min
Delivery at 3,0 bar5 l/min
Pump connectionG3/4 male
min. filling capacity2,5 l
Volume of expansion2 l
Overall dimensions WxDxH **450x510x1230 mm
Net weight143 kg
Power supply requirement230V 1~ 50Hz
max. current13 A
Fuse16 A
Pressure equipment categoryArt. 4.3 PED
Degree of ProtectionIP20
min. ambient temperature5 °C

** Overall dimensions WxDxH

Engineering Facts

14 facts · every claim linked to a page
Selection

The chiller is rated for an operating temperature range of -20 to 40 °C, so the process must stay within these limits.

Defines the application window and excludes processes requiring lower or higher temperatures.

Source
“Operating temperature range -20...40 °C”
Performance

Cooling capacity drops from 2 kW at 0 °C to 0.8 kW at -20 °C, so low-temperature duty needs a larger unit or reduced heat load.

Sizing must account for the reduced capacity at the coldest operating point.

Source
“at -20°C 0,8 kW”
Installation

The unit requires a 230 V single-phase 50 Hz supply with a 16 A fuse and draws up to 13 A.

Determines the electrical circuit, breaker rating and cabling for installation.

Source
“Power supply requirement 230V 1~ 50Hz max. current 13 A Fuse 16 A”
Installation

The chiller is air-cooled and needs adequate space and ambient conditions, with ambient temperature between 5 and 40 °C.

Air-cooled units reject heat to the room, so ventilation and ambient limits must be respected.

Source
“min. ambient temperature 5 °C”
Performance

Cooling capacity is quoted for 20 °C room temperature and may fall if ambient temperature rises.

Prevents overestimating capacity in warm plant rooms.

Source
“Output data valid for: Room temperature 20°C. If the ambient temperature rises, the cooling capacity may drop.”
Cumplimiento

Voltage and frequency tolerances are ±5% and ±2% respectively, but they must not be combined in opposite directions.

Ensures the supply quality is acceptable and avoids invalid tolerance combinations.

Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Safety

The refrigerant is R-449A with GWP 1397 and a charge of 1.2 kg, giving 1.7 t CO2 equivalent.

Affects refrigerant handling, leak reporting and environmental compliance.

Source
“Refrigerant (ASHRAE, GHS) R-449A (A1, H280) Global Warming Potential (GWP) 1397 Refrigerant quantity 1,2 kg CO2 equivalent 1,7 t”
Installation

The pump connection is G3/4 male, so hoses and fittings must match this thread.

Prevents mismatch of process connections during installation.

Source
“Pump connection G3/4 male”
Performance

The circulation pump delivers 25 l/min maximum and 5 l/min at 3 bar, so system pressure drop must be checked against the pump curve.

Ensures the pump can overcome the external circuit resistance.

Source
“max. delivery 25 l/min max. delivery pressure 3 bar Delivery at 3,0 bar 5 l/min”
Installation

The unit has IP20 protection, so it must be installed in a dry indoor location protected from water and dust ingress.

Restricts installation environment and prevents moisture or dust damage.

Source
“Degree of Protection IP20”
Operation

The unit is intended for externally closed applications, so an open process circuit is not suitable.

Defines the hydraulic configuration and prevents incorrect use with open tanks.

Source
“For externally closed applications.”
Maintenance

The minimum filling capacity is 2.5 l and expansion volume is 2 l, which sets the heat-transfer fluid charge and expansion allowance.

Needed to prepare the correct fluid volume and expansion vessel capacity.

Source
“min. filling capacity 2,5 l Volume of expansion 2 l”
Cumplimiento

The equipment is classified as Class A, Group 1 for electromagnetic disturbance to EN55011.

Class A equipment may require additional EMC measures in residential or sensitive environments.

Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Installation

Single-phase devices are supplied with a power cable and country-specific plug, which must be specified at order.

Avoids delays and ensures the correct plug for the installation site.

Source
“Single / two-phase devices (100V to 240V) --> with power cable and country-specific plug (please specify when ordering)”