Introduces the Unichiller 020Tw chiller unit and its main features
Specpp. 1-2
Lists performance, temperature, and dimensional data per DIN 12876
p. 2table, 10 rows
p. 1table, 34 rows
Technical data according to DIN 12876
Parameter
Value
Power supply requirement
230V 1~ 50Hz
max. current
11 A
Fuse
16 A
min. Fuse
10A
max. Fuse
16A
Pressure equipment category
Art. 4.3 PED
Degree of Protection
IP20
min. ambient temperature
5 °C
max. ambient temperature
40 °C
from Serial-No.:
S456129
Technical data according to DIN 12876
Parameter
Value
Operating temperature range
-20...40 °C
temperature set point / display
5,7" colour Touchscreen
Internal temperature sensor
Pt100
Sensor external connection
Pt100
Temperature stability at -10°C
0,5 K
Interface digital
Ethernet, USB (Host u. Device), RS232
Safety classification
I / NFL
Cooling power at 15°C
2 kW
Cooling power at 0°C
2 kW
Cooling power at -10°C
1,5 kW
Cooling power at -20°C
0,8 kW
Refrigeration machine
water-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)
R-449A (A1, H280)
Global Warming Potential (GWP)
1397
Refrigerant quantity
0,4 kg
CO2 equivalent
0,6 t
Circulation pump
B
max. delivery
25 l/min
max. delivery pressure
3 bar
Delivery at 0,33 bar
22 l/min
Delivery at 0,5 bar
20 l/min
Delivery at 1,0 bar
17 l/min
Delivery at 2,0 bar
10 l/min
Delivery at 3,0 bar
5 l/min
Pump connection
G3/4 male
Cooling water connection
G1/2 male
Consumption at water 15°C, flow 0°C
90 l/h
min. cooling water differential pressure
1 bar
max. cooling water pressure
6 bar
min. filling capacity
2,5 l
Volume of expansion
4 l
Overall dimensions WxDxH **
400x440x1230 mm
Net weight
133 kg
sound pressure level +/- 4 dB(A)
60 dB(A)
** Overall dimensions WxDxH
Engineering Facts
12 facts · every claim linked to a page
Installation
The chiller is water-cooled and requires a cooling water supply with a minimum differential pressure of 1 bar and a maximum pressure of 6 bar.
To ensure the cooling water circuit can provide adequate flow and pressure for heat rejection.
Source
“min. cooling water differential pressure 1 bar”
Installation
The unit is designed for cooling water intake temperatures between 5°C and 20°C, and pipes must be insulated if condensation occurs.
Operating outside this range may reduce cooling capacity or cause condensation and damage.
Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Performance
Cooling power drops as cooling water temperature rises, and the required cooling water flow rate may increase.
To correctly size the cooling water supply and predict performance under actual site conditions.
Source
“As the cooling water temperature increases, drop in the cooling power should be expected, and also an increased cooling water flow rate possible.”
Installation
The cooling water circuit materials include copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible cooling water should be used.
To avoid corrosion or damage by selecting suitable water quality and treatment.
Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.”
Cumplimiento
Voltage tolerance is ±5% only with a simultaneous frequency tolerance of ±2%, and certain combinations are not allowed.
To verify site power quality meets the allowed supply tolerances before installation.
Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Installation
The power cable configuration depends on the device type: single/two-phase devices come with a country-specific plug, while three-phase devices below 63A come with cable but no plug.
To order the correct power connection and plan for plug installation on site.
Source
“Single / two-phase devices (100V to 240V) --> with power cable and country-specific plug (please specify when ordering)”
Safety
The refrigerant is R-449A with GHS classification A1, H280, and a global warming potential of 1397.
To assess refrigerant safety, handling requirements, and environmental impact.
Source
“R-449A (A1, H280)”
Operation
The minimum filling capacity is 2.5 liters and the expansion volume is 4 liters.
To ensure the external circuit has sufficient heat transfer fluid volume for proper operation.
Source
“min. filling capacity 2,5 l”
Installation
The unit has a protection degree of IP20 and must be operated within an ambient temperature range of 5°C to 40°C.
To select a suitable indoor location free from water ingress and within ambient limits.
Source
“Degree of Protection IP20”
Cumplimiento
The equipment emits electromagnetic disturbance according to EN55011 Class A, Group 1.
To evaluate electromagnetic compatibility with nearby sensitive equipment in the installation environment.
Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Maintenance
The refrigerant quantity is 0.4 kg, corresponding to 0.6 t CO2 equivalent.
To plan refrigerant handling, leak checks, and environmental reporting obligations.
Source
“Refrigerant quantity 0,4 kg”
Selection
The circulation pump delivers a maximum of 25 l/min and a maximum delivery pressure of 3 bar.
To verify the pump can overcome the pressure drop of the external circuit at the required flow.