Introduces the Unichiller 022w chiller and its main features
Specpp. 1-2
Lists performance, temperature, and physical specifications per DIN 12876
p. 2table, 6 rows
p. 1table, 33 rows
Technical data according to DIN 12876
Parameter
Value
max. current
6,5 A
Fuse
16 A
Degree of Protection
IP20
min. ambient temperature
5 °C
max. ambient temperature
40 °C
from Serial-No.:
S580475
Technical data according to DIN 12876
Parameter
Value
Operating temperature range
-10...40 °C
temperature set point / display
5,7" colour Touchscreen
Internal temperature sensor
Pt100
Sensor external connection
Pt100
Temperature stability at -10°C
0,5 K
Interface digital
Ethernet, USB (Host u. Device), RS232
Safety classification
I / NFL
Cooling power at ambient temperature +20°C at 15°C
2,2 kW
at 0°C
1,6 kW
at -10°C
1 kW
Refrigeration machine
water-cooled, natural refrigerant
Refrigerant (ASHRAE, GHS)
R-290 (A3, H220)
Global Warming Potential (GWP)
0,02
Refrigerant quantity
0,17 kg
CO2 equivalent
0,0 t
UN-number
UN 3358
Circulation pump
Immersion pump
max. delivery
29 l/min
max. delivery pressure
1 bar
Delivery at 0,5 bar
21 l/min
Pump connection
G3/4 male
Cooling water connection
G1/2 male
Consumption at water 15°C, flow 15°C
95 l/h
Consumption at water 15°C, flow 0°C
96 l/h
Consumption at water 15°C, flow -10°C
77 l/h
min. cooling water differential pressure
1 bar
max. cooling water pressure
6 bar
min. filling capacity
3,8 l
Volume of expansion
1,7 l
Overall dimensions WxDxH **
420x487x579 mm
Net weight
61 kg
sound pressure level +/- 4 dB(A)
60 dB(A)
Power supply requirement
220-240V 1~ 50Hz
** Overall dimensions WxDxH
Engineering Facts
13 facts · every claim linked to a page
Installation
The chiller requires cooling water inlet temperature between 5°C and 20°C, so the facility water supply must be capable of maintaining that range.
Determines whether the site cooling water supply is suitable and whether additional chilling or mixing is needed.
Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Installation
If cooling water temperature can fall below the condensation point, the cooling water pipes inside the unit must be insulated.
Prevents condensation and corrosion/electrical risk when cold cooling water is used.
Source
“When cooling water temperatures fall below the condensation point, it is essential to insulate the cooling water pipes in the unit.”
Performance
Cooling capacity drops and cooling water flow rate may need to increase as the cooling water temperature rises.
Affects sizing of the cooling water supply and expected process performance at higher facility water temperatures.
Source
“As the cooling water temperature increases, drop in the cooling power should be expected, and also an increased cooling water flow rate possible.”
Installation
The cooling water circuit contains copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so the cooling water must be compatible with these materials.
Guides water quality and treatment requirements to avoid corrosion or deposits.
Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.”
Operation
The plastic pump is not resistant to acetone or polyglycols, so these fluids must not be used as the heat transfer medium.
Prevents chemical attack and pump failure when selecting the heat transfer fluid.
Source
“The plastic pump is not resistant against acetone and polyglycols (depending on the manufacturer).”
Operation
For freeze protection, water should be mixed with glysantine or ethylene glycol rather than relying on pure water.
Ensures safe operation at sub-zero temperatures and avoids freezing damage.
Source
“It is recommended that water is mixed with either glysantine or ethylene glycol for freeze protection.”
Cumplimiento
Voltage tolerance is ±5% only with simultaneous frequency tolerance of ±2%, and certain combinations such as -5% voltage with +2% frequency are not allowed.
Defines acceptable power supply quality and prevents invalid supply combinations during site assessment.
Source
“-5% voltage and +2% frequency -> not allowed!”
Cumplimiento
The unit is classified as EN55011 Class A, Group 1 for electromagnetic disturbance, which has implications for the installation environment.
Class A equipment may require additional EMC measures in residential or sensitive environments.
Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Installation
Single/two-phase devices are supplied with a power cable and country-specific plug, which must be specified at ordering.
Ensures the correct plug type is delivered and avoids delays or rework at installation.
Source
“Single / two-phase devices (100V to 240V) --> with power cable and country-specific plug (please specify when ordering)”
Selection
The unit is compliant with US-SNAP and applicable EU laws, with US-SNAP end-use defined as industrial process refrigeration.
Confirms regulatory suitability for industrial process refrigeration applications in the US and EU.
Source
“This equipment is compliant to US-SNAP and all applicable EU laws. The US-SNAP end-use for this equipment is the industrial process refrigeration.”
Installation
Space requirements around the unit must be respected according to the operating conditions published online.
Prevents installation in insufficient space that could affect performance or service access.
Source
“Please respect space requirements. See operating conditions at www.huber-online.com”
Operation
The Pilot ONE controller supports USB and network connections, enabling remote control and data transmission with Spy software.
Useful for integration into monitoring or data logging systems.
Source
“The USB port allows connection of the system to a PC or notebook. Together with the Spy software, requirements such as remote control or data transmission are easily achieved in a cost-effective manner.”
Selection
The unit uses R-290 refrigerant with GWP 0.02 and a charge of 0.17 kg, which is relevant for environmental and safety assessments.
R-290 is flammable (A3), so safety measures and refrigerant regulations must be considered.