Unichiller 025w-H Recirculating Chiller Datasheet

Contenido

  1. Overview p. 1
    Introduces the Unichiller 025w-H chiller and its main features
  2. Spec pp. 1-2
    Lists performance, temperature, and physical specifications per DIN 12876
  3. p. 2 table, 8 rows
  4. p. 1 table, 33 rows

Technical data according to DIN 12876

ParameterValue
sound pressure level +/- 4 dB(A)60 dB(A)
Power supply requirement220-240V 1~ 50Hz
max. current15 A
Fuse16 A
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:1.1/24

Unichiller 025w-H — Technical data according to DIN 12876

ParameterValue
Operating temperature range-10...100 °C
temperature set point / display5,7" colour Touchscreen
Internal temperature sensorPt100
Sensor external connectionPt100
Temperature stability at -10°C0,5 K
Interface digitalEthernet, USB (Host u. Device), RS232
Safety classificationIII / FL
Heating power at 240V2,1 kW
Heating power at 230V2 kW
Heating power at 220V1,8 kW
Cooling power at ambient temperature +20°C at 15°C2,5 kW
at 0°C2 kW
at -10°C1,2 kW
Refrigeration machinewater-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)R-449A (A1, H280)
Global Warming Potential (GWP)1397
Refrigerant quantity0,4 kg
CO2 equivalent0,6 t
Circulation pumpImmersion pump
max. delivery29 l/min
max. delivery pressure1 bar
Delivery at 0,5 bar21 l/min
Pump connectionG3/4 male
Cooling water connectionG1/2 male
Consumption at water 15°C, flow 15°C84 l/h
Consumption at water 15°C, flow 0°C84 l/h
Consumption at water 15°C, flow -10°C72 l/h
min. cooling water differential pressure1 bar
max. cooling water pressure6 bar
min. filling capacity3,8 l
Volume of expansion1,7 l
Overall dimensions WxDxH **420x487x579 mm
Net weight63 kg

** см. сноску в оригинале (в тексте страницы пояснение к ** отсутствует)

Engineering Facts

12 facts · every claim linked to a page
Installation

The chiller is water-cooled and requires a cooling water supply with a minimum differential pressure of 1 bar and a maximum pressure of 6 bar.

Determines the required cooling water supply pressure and the need for a pressure-reducing valve if the site pressure exceeds 6 bar.

Source
“min. cooling water differential pressure 1 bar max. cooling water pressure 6 bar”
Installation

The unit is designed for cooling water intake temperatures between 5°C and 20°C, and below the condensation point the cooling water pipes inside the unit must be insulated.

Defines the acceptable cooling water temperature window and the mandatory insulation measure to prevent condensation inside the unit.

Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C. When cooling water temperatures fall below the condensation point, it is essential to insulate the cooling water pipes in the unit.”
Performance

Cooling power drops as cooling water temperature rises, and the required cooling water flow rate may increase accordingly.

Allows the engineer to size the cooling water supply and anticipate reduced cooling capacity at higher water temperatures.

Source
“As the cooling water temperature increases, drop in the cooling power should be expected, and also an increased cooling water flow rate possible.”
Installation

The cooling water circuit contains copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only suitable cooling water may be used.

Guides water quality and material compatibility checks to avoid corrosion or fouling of the cooling water circuit.

Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM. Please use suitable cooling water.”
Operation

The plastic pump is not resistant to acetone or polyglycols, so these media must not be used; water mixed with glysantine or ethylene glycol is recommended for freeze protection.

Prevents chemical attack on the pump and specifies an acceptable antifreeze strategy for low-temperature operation.

Source
“Special Case: Acetone and Polyglycol: The plastic pump is not resistant against acetone and polyglycols (depending on the manufacturer). It is recommended that water is mixed with either glysantine or ethylene glycol for freeze protection.”
Cumplimiento

Voltage and frequency tolerances are ±5% and ±2% respectively, but the combination of -5% voltage with +2% frequency is not allowed.

Defines the acceptable power supply quality and a prohibited tolerance combination that could damage the unit.

Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2% Example -5% voltage and +2% frequency -> not allowed!”
Installation

Single- and two-phase devices are supplied with a power cable and country-specific plug, which must be specified at ordering.

Ensures the correct mains plug is ordered for the installation site and avoids delays at commissioning.

Source
“Single / two-phase devices (100V to 240V) --> with power cable and country-specific plug (please specify when ordering)”
Safety

The refrigerant R-449A is classified A1 with GHS hazard H280, and the unit contains 0.4 kg corresponding to 0.6 t CO2 equivalent.

Informs refrigerant safety classification, charge quantity and CO2-equivalent reporting for site safety and environmental compliance.

Source
“R-449A (A1, H280) Global Warming Potential (GWP) 1397 Refrigerant quantity 0,4 kg CO2 equivalent 0,6 t”
Installation

The unit has protection degree IP20 and must be operated within an ambient temperature range of 5°C to 40°C.

Sets the environmental limits and ingress protection for selecting a suitable installation location.

Source
“Degree of Protection IP20 min. ambient temperature 5 °C max. ambient temperature 40 °C”
Selection

The maximum current is 15 A and the fuse rating is 16 A on a 220-240 V single-phase 50 Hz supply.

Used to size the electrical supply circuit and protective device for the chiller.

Source
“Power supply requirement 220-240V 1~ 50Hz max. current 15 A Fuse 16 A”
Operation

The minimum filling capacity is 3.8 l and the expansion volume is 1.7 l, which must be respected when filling the bath.

Ensures sufficient heat transfer fluid volume and expansion space to avoid overflow or pump cavitation.

Source
“min. filling capacity 3,8 l Volume of expansion 1,7 l”
Maintenance

The unit is covered by a 4-year warranty, but registration is required.

Reminds the engineer to register the unit to secure the extended warranty coverage.

Source
“4-year warranty - registration required.”