Unichiller 025w Technical Data

Contenido

  1. Overview p. 1
    Introduces the Unichiller 025w chiller model and its main features
  2. Spec pp. 1-2
    Performance, temperature, cooling capacity and dimensional data per DIN 12876
  3. p. 2 table, 6 rows
  4. p. 1 table, 34 rows

Technical data according to DIN 12876

ParameterValue
min. Fuse10A
max. Fuse16A
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:S575976

Technical data according to DIN 12876

ParameterValue
Operating temperature range-10...40 °C
temperature set point / display5,7" colour Touchscreen
Internal temperature sensorPt100
Temperature stability at -10°C0,5 K
Interface digitalEthernet, USB (Host u. Device), RS232
Safety classificationI / NFL
Cooling power at ambient temperature +20°C
at 15°C2,5 kW
at 0°C2 kW
at -10°C1,2 kW
Refrigeration machinewater-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)R-449A (A1, H280)
Global Warming Potential (GWP)1397
Refrigerant quantity0,4 kg
CO2 equivalent0,6 t
UN-numberUN 2857
Circulation pumpImmersion pump
max. delivery29 l/min
max. delivery pressure1 bar
Delivery at 0,5 bar21 l/min
Pump connectionG3/4 male
Cooling water connectionG1/2 male
Consumption at water 15°C, flow 15°C84 l/h
Consumption at water 15°C, flow 0°C84 l/h
Consumption at water 15°C, flow -10°C72 l/h
min. cooling water differential pressure1 bar
max. cooling water pressure6 bar
min. filling capacity3,8 l
Volume of expansion1,7 l
Overall dimensions WxDxH **420x487x579 mm
Net weight59 kg
sound pressure level +/- 4 dB(A)60 dB(A)
Power supply requirement220-240V 1~ 50Hz
max. current8,5 A

** Overall dimensions WxDxH

Engineering Facts

12 facts · every claim linked to a page
Installation

The chiller is water-cooled and requires a cooling water supply with inlet temperature between 5°C and 20°C and at least 1 bar differential pressure.

Cooling water conditions must be verified at the installation site before commissioning.

Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Installation

If cooling water temperature can fall below the condensation point, the cooling water pipes inside the unit must be insulated.

Prevents condensation and possible damage or electrical risk in low-temperature cooling water installations.

Source
“When cooling water temperatures fall below the condensation point, it is essential to insulate the cooling water pipes in the unit.”
Performance

Cooling power drops and cooling water flow rate may increase as the cooling water temperature rises.

Selection must account for actual site cooling water temperature, not only nominal performance.

Source
“As the cooling water temperature increases, drop in the cooling power should be expected, and also an increased cooling water flow rate possible.”
Installation

The cooling water circuit contains copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only suitable cooling water may be used.

Water quality and treatment must be compatible with these materials to avoid corrosion or fouling.

Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.”
Operation

The plastic pump is not resistant to acetone or polyglycols, so these media must not be used without checking compatibility.

Prevents pump damage when selecting heat transfer fluid for the application.

Source
“The plastic pump is not resistant against acetone and polyglycols (depending on the manufacturer).”
Operation

For freeze protection, water should be mixed with glysantine or ethylene glycol rather than relying on pure water.

Guides heat transfer fluid selection for operation below or near freezing.

Source
“It is recommended that water is mixed with either glysantine or ethylene glycol for freeze protection.”
Cumplimiento

Voltage tolerance is ±5% only with simultaneous frequency tolerance of ±2%, and certain combinations such as -5% voltage with +2% frequency are not allowed.

Site power quality must be checked against these limits before installation.

Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Cumplimiento

The unit is EMC classified as Class A, Group 1 according to EN55011.

Class A equipment may require EMC precautions in residential or sensitive environments.

Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Installation

Single/two-phase devices are supplied with a power cable and country-specific plug, which must be specified when ordering.

Ensures the correct plug type is ordered for the destination country.

Source
“Single / two-phase devices (100V to 240V) --> with power cable and country-specific plug (please specify when ordering)”
Installation

Space requirements around the unit must be respected according to the operating conditions.

Affects layout and clearance planning during installation.

Source
“Please respect space requirements. See operating conditions at www.huber-online.com”
Safety

The refrigerant is R-449A with safety group A1 and GHS hazard H280, and the charge is 0.4 kg corresponding to 0.6 t CO2 equivalent.

Relevant for refrigerant handling, safety classification and environmental reporting.

Source
“R-449A (A1, H280)”
Operation

The unit has a minimum filling capacity of 3.8 l and an expansion volume of 1.7 l.

Determines the minimum heat transfer fluid volume and expansion space needed for safe operation.

Source
“min. filling capacity”