The chiller requires a three-phase 460 V 60 Hz supply with a 3x10 A fuse, so the site must provide a matching branch circuit.
Confirms the electrical supply and protection needed before installation.
Source
“Power supply (3 Phase) 460V 3~ 60Hz”
| Parameter | Value |
|---|---|
| Degree of Protection | IP20 |
| min. ambient temperature | 5 °C |
| max. ambient temperature | 40 °C |
| from Serial-No.: | 1.0/20 |
| Parameter | Value |
|---|---|
| Operating temperature range | -10...40 °C |
| temperature set point / display | 5,7" colour Touchscreen |
| Internal temperature sensor | Pt100 |
| Sensor external connection | Pt100 |
| Interface digital | Ethernet, USB (Host u. Device), RS232 |
| Temperature stability at -10°C | 0,5 K |
| Safety classification | I / NFL |
| Cooling power at 40°C | 3 kW |
| Cooling power at 15°C | 3 kW |
| Cooling power at 0°C | 3 kW |
| Cooling power at -10°C | 2 kW |
| Refrigeration machine | air-cooled, CFC- and HCFC-free |
| Refrigerant (ASHRAE, GHS) | R-449A (A1, H280) |
| Global Warming Potential (GWP) | 1397 |
| Circulation pump | B |
| max. delivery | 31 l/min |
| max. delivery pressure | 4 bar |
| Delivery at 0,5 bar | 27 l/min |
| Delivery at 1,0 bar | 22 l/min |
| Delivery at 1,5 bar | 19 l/min |
| Delivery at 2,0 bar | 16 l/min |
| Delivery at 2,5 bar | 12 l/min |
| Delivery at 3,0 bar | 8 l/min |
| Delivery at 3,5 bar | 5 l/min |
| Pump connection | G3/4 male |
| min. filling capacity | 3,5 l |
| Volume of expansion | 3,5 l |
| Overall dimensions WxDxH ** | 500x552x1451 mm |
| Net weight | 164 kg |
| sound pressure level +/- 4 dB(A) | 69 dB(A) |
| Power supply (3 Phase) | 460V 3~ 60Hz |
| max. current | 4,5 A |
| Fuse (3 phase) | 3x10A |
| Pressure equipment category | Art. 4.3 PED |
** Overall dimensions WxDxH
The chiller requires a three-phase 460 V 60 Hz supply with a 3x10 A fuse, so the site must provide a matching branch circuit.
Confirms the electrical supply and protection needed before installation.
“Power supply (3 Phase) 460V 3~ 60Hz”
External branch circuit protection according to UL 489 is mandatory, so a suitable UL 489 breaker must be included in the installation.
Ensures code-compliant electrical protection is planned and installed.
“External branch circuit protection according to UL 489 required.”
Three-phase devices drawing less than 63 A are supplied with cable but without a plug, so the installer must provide the final connection.
Avoids surprises at commissioning regarding the power connection hardware.
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Voltage and frequency tolerances are ±5% and ±2% respectively, but they must not be combined in opposite directions, so the supply quality must be verified.
Prevents invalid supply conditions that could damage the chiller or void compliance.
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Cooling capacity is rated at 20 °C room temperature and may drop at higher ambient temperatures, so capacity must be derated for warm rooms.
Affects selection margin and expected performance in non-20 °C environments.
“Output data valid for: Room temperature 20°C. If the ambient temperature rises, the cooling capacity may drop.”
The unit may only be operated between 5 °C and 40 °C ambient temperature, so the installation location must stay within this range.
Defines the allowable room conditions for reliable operation.
“min. ambient temperature 5 °C”
The chiller uses refrigerant R-449A with GHS classification H280 and GWP 1397, so refrigerant handling and leak response must follow H280 precautions.
Informs safety planning, labeling, and refrigerant management obligations.
“R-449A (A1, H280)”
The pressure equipment falls under Article 4.3 PED, so the unit is categorized accordingly and must be integrated with compliant system design.
Clarifies the pressure equipment regulatory category for the installation.
“Art. 4.3 PED”
The equipment is Class A, Group 1 for electromagnetic disturbance, so it may require EMC measures in residential or sensitive environments.
Guides EMC planning and suitability for the intended electromagnetic environment.
“Classification (disturbance) to EN55011: Class A, Group 1”
The unit has IP20 protection, so it must be installed in a dry indoor location protected from water and dust ingress.
Restricts installation to suitable indoor environments and informs enclosure planning.
“Degree of Protection IP20”
The pump connection is G3/4 male, so mating hoses and fittings must match this thread size.
Ensures correct hydraulic connections are ordered and available at installation.
“Pump connection G3/4 male”
The minimum filling capacity is 3.5 l and expansion volume is 3.5 l, so the external circuit volume and expansion allowance must be checked.
Prevents underfilling or inadequate expansion capacity in the external loop.
“min. filling capacity 3,5 l”
The pump delivers 31 l/min maximum and 4 bar maximum pressure, with flow dropping as pressure rises, so system pressure drop must be matched to the pump curve.
Supports hydraulic selection and verification of adequate flow in the external circuit.
“max. delivery 31 l/min”
The unit has a digital level indicator and an atmospheric open expansion tank, so fluid level must be monitored and maintained.
Indicates the need for routine fluid level checks as part of maintenance.
“With digital level indicator.”