Technical data table according to DIN 12876 standard
p. 2table, 11 rows
Technical data according to DIN 12876
Parameter
Value
Overall dimensions WxDxH **
400x440x1230 mm
Net weight
141 kg
sound pressure level +/- 4 dB(A)
70 dB(A)
Power supply (3 Phase)
400V 3~ 50Hz
max. current (3 Phase)
9 A
Fuse (3 phase)
3x10A
Pressure equipment category
Art. 4.3 PED
Degree of Protection
IP20
min. ambient temperature
5 °C
max. ambient temperature
40 °C
from Serial-No.:
S407974
** Please respect space requirements. See operating conditions at www.huber-online.com
Engineering Facts
12 facts · every claim linked to a page
Installation
The chiller is water-cooled and requires a cooling water supply with a minimum differential pressure of 1 bar and a maximum pressure of 6 bar.
To ensure the cooling water circuit can provide adequate flow and is not overpressurized, which is critical for proper installation and safe operation.
Source
“min. cooling water differential pressure
1 bar
max. cooling water pressure
6 bar”
Installation
The unit must be operated with cooling water intake temperature between 5°C and 20°C; if below the condensation point, cooling water pipes must be insulated.
To prevent condensation and ensure reliable operation, the installer must verify the cooling water temperature and provide insulation if necessary.
Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C. When cooling water temperatures fall below the condensation point, it is essential to insulate the cooling water pipes in the unit.”
Selection
The cooling power drops as cooling water temperature increases, and the required cooling water flow rate may increase.
To properly size the cooling water supply and ensure the chiller meets the process cooling demand under actual site conditions.
Source
“As the cooling water temperature increases, drop in the cooling power should be expected, and also an increased cooling water flow rate possible.”
Installation
The cooling water circuit materials include copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only suitable cooling water must be used.
To avoid corrosion or damage to the cooling circuit, the water quality must be compatible with these materials.
Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM. Please use suitable cooling water.”
Cumplimiento
The equipment is classified as Class A, Group 1 for electromagnetic disturbance according to EN55011.
To ensure electromagnetic compatibility in the installation environment and to meet regulatory requirements.
Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Installation
Three-phase devices with current consumption less than 63A are supplied with a cable but without a plug.
The installer must provide a suitable plug or hardwire the connection, as the unit does not include one.
Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Operation
The unit has a maximum ambient temperature of 40°C and a minimum of 5°C.
To ensure the chiller operates within its specified environmental limits and maintains performance and reliability.
Source
“min. ambient temperature
5 °C
max. ambient temperature
40 °C”
Safety
The refrigerant is R-449A with safety group A1 and GHS hazard H280, and the global warming potential is 1397.
To assess refrigerant safety, environmental impact, and compliance with regulations regarding refrigerant use and handling.
Source
“R-449A (A1, H280)
Global Warming Potential (GWP)
1397”
Installation
The power supply is 400V 3~ 50Hz with a maximum current of 9A and a 3x10A fuse.
To ensure the electrical supply and protection devices are correctly sized for safe installation.
Source
“Power supply (3 Phase)
400V 3~ 50Hz
max. current (3 Phase)
9 A
Fuse (3 phase)
3x10A”
Operation
The unit has a sound pressure level of 70 dB(A) ±4 dB(A).
To consider noise levels in the installation environment and ensure compliance with workplace noise regulations.
Source
“sound pressure level +/- 4 dB(A)
70 dB(A)”
Installation
The unit has a degree of protection IP20, so it must be installed in a dry, protected environment.
To prevent ingress of dust and water, which could damage the equipment and pose a safety hazard.
Source
“Degree of Protection
IP20”
Selection
The chiller has a minimum filling capacity of 2.5 liters and an expansion volume of 4 liters.
To ensure the external cooling circuit has sufficient volume and to account for thermal expansion of the heat transfer fluid.
Source
“min. filling capacity
2,5 l
Volume of expansion
4 l”